Left Button
: if is a functional button and can be configured in the menu 6. Light-, Zero-, Reset- or Min- func-
tion can be assigned to the button. The configured function is active in the display mode. Hold
the button during approx. 2 seconds to activate the function. In the operating mode you move in
the menu system backward
"<<"
or you reduce the set value.
-
Right Button
: if is a functional button and can be configured in the menu 6. Light-, Zero-, Reset- or Max- func-
tion can be assigned to the button. . Hold the button during approx. 2 seconds to activate the
function. In the operating mode you move in the menu system forward ">>" or you raise the set
value.
-
Menu-Button
: by pressure of this button "menu" you reach in the operating mode; moreover, she serves for
the choice of the single menu points "Edit" or in the confirmation the opposed worth "Next".
To the configuration of the single menu points the desired menu point is to be put with the help of the left button "<<"
or right button ">>". Confirm this afterwards with the menu button "Edit" the menu point is marked and the configura-
tion can begin.
To store an opposed value the menu button "Next" must be also pressed. To leave the menu press the menu button
so often to the mark of the single menu point and choose with the left button
"<<"
or right button
">>"
the menu 1
and press the menu button once more "Exit". The operating mode will also leave after approx. 1 min automatically.
Changes become first after activity of the menu button "Next" and after abandonment of the menu point effec-
tive.
With the abandonment of the whole menu system the opposed parameters are checked once again in de-
pendence to each other and concerning the characteristics of the device. With the configuration of the unity a
conversion of the measuring area occurs in the new unity only after abandonment of the menu system. Accord-
ing to pressure area not all units can be also used if necessary.
If the device came in contact with hazardous
substances, certain precautions have to be
complied with for purification!
9. Disposal
The device must be disposed according to the
European Directives 2002/96/EG and 2003/108/EG
(on waste electrical and electronic equipment)
Waste of electrical and electronic equipment may
not be disposed by domestic refuse!
WARNING! Depending on the measuring medium, deposit
on the device may cause danger for the user and the
environment. Comply with adequate precautions for
purification and dispose of it properly.
10. Warranty conditions
The warranty conditions are subject to the legal warranty period
of 24 months from the date of delivery. In case of improper use,
modifications of or damages to the device, we do not accept
warranty claims. Damaged diaphragms will also not be accepted.
Furthermore, defects due to normal wear are not subject to
warranty services
11. Declaration of conformity / CE
The delivered device fulfils all legal requirements. The applied
directives, harmonised standards and documents are listed in the
EC declaration of conformity
Additionally, the operational safety is confirmed by the CE sign
on the manufacturing label.
warranty services.
12. Maintenance
In principle, this device is maintenance-free. If desired, the
housing of the device can be cleaned when switched of using a
damp cloth and non-aggressive cleaning solutions.
With certain media, however, the diaphragm may be polluted or
coated with deposit. It is recommended to define corresponding
service intervals for control. After placing the device out of
service correctly, the diaphragm can usually be cleaned carefully
with a non-aggressive cleaning solution and a soft brush or
sponge. If the diaphragm is calcified, it is recommended to send
the device to PCE Instruments for decalcification. Please read
therefore the chapter “Repair” below.
!
A false cleaning of the device can cause an irreparable
damage on the diaphragm. Therefore never use pointed ob-
jects or pressured air for cleaning the diaphragm.
13. Service / Repair
We recommend that the instrument is regularly recalibrated by
PCE Instruments, with time intervals of approx. 12 months.
Before you return, the device has to be cleaned carefully and
packed shatter-proofed. You have to enclose a notice of return
with detailed defect description when sending the device. Contact
PCE Instruments to ask for this form. If your device came in
contact
with
harmful
substances,
a
declaration
of
decontamination is additionally required. Should you dispatch a
device without a declaration of decontamination and if there are
any doubts in our service department regarding the used
medium, repair will not be started until an acceptable declaration
is sent.
14. Menu list
Switch on
The device can be switched on with every button (left button, menu button, right button).
Menu 1/9
on / off
By push the menu button the operating mode will be selected / left / ("Menu" / "Exit").
Menu 2/9
date / time / info
Setting of topical time and the topical date. Software version will be shown.
Menu 3/9
Units
Setting of the pressure unit
adjustable units: [bar], [mbar], [PSI], [inHg], [cmHg], [mmHg], [hPa], [kPa], [MPa], [mmH2O], [mH2O], [kg/cm
2
]
or [user] (the user-defined unit can be programmed only by means of the software BD|LOG), a conversation of
all pressure related parameters is carried out automatically
Setting of decimal places
adjustable decimal place: standard [Std], [+1], [+2]
Setting of the temperature unit
the unity is not adjustable. Factory setting [°C]
Menu 4/9
Min / Max Values
Language:
Setting of user languages German [DE] or English [EN]
Measuring rate: here it is put how often the measurement is carried out and is indicated.
Possible settings are: a measurement per second [1/sec]
or two measurements per second [2/sec]
Auto Off Time:
Setting of the automatic switch off in minutes. The automatic switch off is able in steps from
[1 min], [2 min], [3 min], [4 min] or [5 min]; 30 seconds before switching off the timer is
activated and indicated in the display.
By the option [Off] the device can be deactivated. After deactivation, the precision digital
gauge is in the continuous operation.
Menu 5/9
Configuration 1
Language:
Setting of user languages German [DE] or English [EN]
Measuring rate: here it is put how often the measurement is carried out and is indicated.
Possible settings are: a measurement per second [1/sec]
or two measurements per second [2/sec]
Auto Off Time:
Setting of the automatic switch off in minutes. The automatic switch off is able in steps from
[1 min], [2 min], [3 min], [4 min] or [5 min]; 30 seconds before switching off the timer is
activated and indicated in the display.
By the option [Off] the device can be deactivated. After deactivation, the precision digital
gauge is in the continuous operation.
Menu 6/9
Configuration 2
Button settings: left button / right button
Left Button:
function configurations: [Min], [Light], [Zero], [Reset]
Right Button:
function configurations: [Max], [Light], [Zero], [Reset]
Function descriptions:
- [Min] / [Max]
minimum / maximum pressure value will be showed
- [Light]
the display backlight is activated
- [Zero]
the zero point is automatically put, in the display appears
- [Reset]
the opposed zero is put back, goes out
Damping:
the damping can be put in 1 second steps between [1 sec] and [10 sec] or be
deactivated by option [Off]
Menu 7/9
Backlight
Backlight settings of display
On Time:
the backlight duration can be put in 1 second steps between [1 sec] and [10 sec] or be
deactivated by option [Off]
Brightness:
the brightness can be put into 5% of steps between [0%] and [100%]
Menu 8/9
Data Logger
Data Logger configuration
Data Logger:
the following settings are possible: linearly [Linear] (value admission to the counter level 8500
is reached), cyclically ( [Loop] (after the value is reached in 8500, the data logger automatically
begins the values once more to grasp and, besides, headlines the old values) or [Off]
(in the display appears if the data logger is activated and goes out if the data logger is off)
Interval:
Intervals to the memory of the measuring values (pressure / temperature) second [1-99 sec],
minute [1-99 min], hour [1-99 hour] or day [1-99 days]
Time:
Measuring value admission: in which time the measuring value admission should occur (is
only efficiently for the setting "daily").
Counter:
Number of the grasped measuring values is indicated. Maximum value: 8500
(Put back counter reading: menu point with "Edit" select → button "Next" four times operate → button ">>" operate. There appears the
question "Reset?" → once more operate the button ">>". It seems "Sure?" additional confirmation whether the value should be reset →
repeated confirming with the button ">>" reset the grasped measuring values. Display announcement "Counter: 0/8500")
Menu 9/9
Sensor Data
[SN:]
the standard number is indicated
[Upper]
Measuring area end
[Lower]
Measuring area beginning
[Date]
Production date
Values are put by factory and are not adjustable.
Error
Display "No Sensor": display and pressure sensor module are separated.