Weight
6.2 oz. (175g) including battery
Dimensions
164 x 65 x 32mm (6.5 x 2.6 x 1.3”) (HXWXD)
Jaw opening
18mm (0.7”)
Standards
IEC61010 Category II 600V,
z
Range Specifications
Function
Range
Resolution
Accuracy (of reading)
DC Current
200A
100mA
± (2.8% + 8 digits)
AC Current (50/60Hz)
200A
100mA
±(3.0% + 8 digits)
Frequency
(Auto-Ranging)
40-51Hz
0.01Hz
±(1.2% + 5 digits)
Input Sensitivity: 5A AC RMS min.
51-510Hz 0.1Hz
0.51-1kHz 0.001kHz
Non-Contact AC Voltage
100VAC to 600VAC 50/60Hz
Meter Description
1, Current sense jaw
2. Flashlight button
3. Rotary function switch
4. Data HOLD and Backlight key
5. LCD display
6. Clamp trigger
7. One-touch DCA ZERO key
8. Non-contact AC voltage indicator light
9.
Non-contact AC voltage detector probe tip
Safety Information
!
Caution! Refer to the explanation in this Manual
Double Insulation
This meter has been designed to be safe in use, but the operator must use caution in its operation. The
rules listed below should be carefully followed for safe operation.
1.
NEVER
apply voltage or current to the meter that exceeds the specified maximum:
2.
USE EXTREME CAUTION
when working with voltages greater than 25VAC rms. These voltages are
considerd a shock hazard.
3.
NEVER
operate the meter unless the back cover and the battery/fuse door are in place and fastened
securely
Input Limits
CC