Manual
12
5.3.4 Tolerances
To set up or adjust tolerances, press the Menu button
, select “Tolerance“ and press Enter
. Now
you get to the tolerance screen.
Here you can set up tolerances. Select an option by using the arrow keys (
and
) and press Enter
to confirm.
Now you get to an entry screen. By using the arrow keys (
and
), you can adjust the selected digit.
Press Enter
to go to the next digit. After confirming the last digit by pressing Enter
, the tolerance is
set up.
Note: The threshold value refers to the „ΔE*“ value in the sample measurement display. It specifies the
maximum acceptable value of the deviation. If the measured value exceeds the threshold value, you can
see a “Fail” indication in the display. Otherwise, the display will show “Pass”.
5.3.5 Delete recorded data
To delete the recorded data, press the Menu button
, select “Delete“ and press Enter
. Now, you get
to the deletion screen.
You have two different options here. You can either delete all sample measurem
ents (“Delete ALL
Samples”) or all records (“Delete ALL Records”).
Select the desired option by using the arrow keys (
and
) and press Enter
. A confirmation screen
appears. Press Enter
to confirm or Back
to cancel.
Note: If you delete all records, all sample measurements will be deleted as well.
5.3.6 Select the light source
To select the light source, press the Menu button
, select “Light“ and press Enter
. Now you get to
the light source settings.
Here you can choose between the following options: D65, D50, A, C, F2(CWF), F6, F7(DLF), F8,
F10(TPL5), F11(TL84), F12(TL83/U30).
Use the arrow keys (
and
) to select a light source and press Enter
to confirm.