Click “White Calibration”, it will prompt you to put away the white board. Put the measuring
aperture on white board. Make sure the measuring aperture is close to the white board. Click
“OK” or press “Testing” key to perform white calibration. Please pay attention that the serial
number of white board must correspond to the instrument.
Click “Black Calibration”, it will prompt you to put away black cavity. Put the measuring
aperture into
black calibration cavity. Click “OK” or press “Testing” key to perform black
calibration.
Then, white calibration and black calibration are completed.
4.3 Measurement
4.3.1 Measurement Locating
The instrument can visually locate through the cross of measuring aperture. It can be done
by observing the alignment between the cross and the sample, while moving the measuring
aperture to adjust the position.
4.3.2 Standard Measurement
Please enter “Standard Measurement” interface to perform measurements as shown in
Figure 6.
Put the measuring aperture on the standard sample, press “Testing” key. The colorimeter will
display color data of the standard sample.
Figure 6 Standard Measurement