Calibration mode
Never perform a calibration without calibration weights.
1. Turn off the scale.
2. Remove all weights from the scale, so that the scale is unloaded.
3. Keep the
key pressed and turn on the scale.
4. Now release the
button and the display should show
“SPEN” now.
5. Press the
button again briefly. The display will now show
“SET” and then “00000”.
6. Now set the weight on same value as your calibration weight, by choosing values with the
button and jumping to next digit left to the prior by pressing the
button.
7. Press again briefly the
button, to save value.
8. The scale will then automatically switch to the normal weighing mode.
Settings in internal menu
1. Turn on the scale
2. While the scale plays a countdown from 9 to 0 on its display, briefly press
.
3. Now the display will indicate
“bEE 0”.
4. With
you can switch the acoustic signal for stabilized values on (
“1”) or off (“0”).
5. Confirm with
.
6. Now the display will indicate
“CHt 0”. With
you can now chose the unit.
7. Press
again to confirm.
8. Now the display will indicate
“Aut 0“. Here you can either turn off the background lights (“0”) or
switch to Automatic mode (
“1”).
Disposal
Batteries may not be put in the household waste. They have to be given in the set up collection points.
Please contact PCE Instruments if you have any questions concerning our products, service etc.