LOW-BATTERY 

 

Battery symbol will appear on display at right top corner when the power is low. Failure to 

replace batteries will have an effect on the accuracy of the readings. 

 

 

 

1.  Open the battery cover on the rear side. 

2.  Remove the expired batteries. 

3.  Insert new 2 pcs AAA batteries and make sure the batteries are in correct polarity. 

Then, put on the cover. 

 

CALIBRATION 

 

1.  Turn the meter off and plug the sensor into 33% salt bottle. Press “ON/OFF + C/F” 

keys more than one sec. to enter calibration mode. 

2.  “32.8%” will flash on LCD and DP, WB is displayed as “---.-“ that means the values are 

invalid in the calibration process. After 30 min, the flashing will stop to indicate the 

procedure is finished.  

3.  Move the sensor to 75% salt bottle and press 

 more than one sec. to enter 

75.3% calibration. “75.3%” will flash on LCD. 

4.  Flashing stops after 30 mins, up to now the whole calibration is completed and the 

calibration data has been saved in memory. 

 

Note: 

a.  You can exit calibration procedure without saving by pressing “ON/OFF” key before 

step4. At step 4, pressing “ON/OFF” key to exit calibration. 

b.  Auto power off is disabled in calibration mode. 

c.  To get high accuracy, calibration should be operated at 23 degree C. 

d.  If the reading is out of 75.3% ±0.5 % at step 4, it means the calibration is failed.  

Summary of Contents for PCE-310

Page 1: ...1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire South...

Page 2: ...nable cap to protect sensor External Temp probe Triple LCD digital display T1 T2 T2 DP function Model 8716 Data Hold to capture readings Pocket size easy to fit in pocket Low battery indication Fast r...

Page 3: ...3 WARNING External probe port IMPORTANT BE SURE TO OPEN THE SENSOR PROTECTION CAP BEFORE STARTING THE MEASUREMENT IN ORDER TO GET ACCURATE VALUE FOR ALL MODELS CONTROLS AND INDICATORS...

Page 4: ...n itself off after 10 minutes To override Auto Power Off function press ON OFF HOLD while the meter is off When n appears See Fig A release HOLD button Meter is now in Non Sleep Mode For PCE 310 the a...

Page 5: ...ON Press button until DP appears on display Select to display dew point T1 air temp or delta T T1 T2 in any mode while the meter is on See Fig E WET BULB FUNCTION Press until Wbappear on the LCD User...

Page 6: ...than two seconds to clear current reading from memory WARNING While checking MIN MAX value for EX T2 and don t replace probe otherwise you will get error code displayed If you don t plug the probe int...

Page 7: ...s displayed as that means the values are invalid in the calibration process After 30 min the flashing will stop to indicate the procedure is finished 3 Move the sensor to 75 salt bottle and press more...

Page 8: ...auto power off function Er 1 Circuit error in RH measurement channel return the meter to the store for repairing Er 2 Circuit error in Internal temperature measurement channel return the meter to the...

Page 9: ...son When requiring a RA Return Authorization please include data regarding the defective reason the meters are returned along with good packing to prevent any damage in shipment and insured against po...

Reviews: