background image

 
 

 

6

MIN./MAX. FUNCTION 

 

1.  Press and hold 

 button until (MIN) appears on display. (See Fig. G) Display is 

now showing minimum humidity and temperature readings in memory. 

2.  Press and hold 

 button again until (MAX) appears on display. (See Fig.H). 

Display is now showing maximum humidity and temperature readings in memory. 

3.  To return to current temperature and humidity readings press and hold MN/MX or 

 button until Min or Max disappear from display. 

4.  Press 

RESET

 or 

 button more than two seconds to clear current reading from 

memory. 

 

WARNING:

 While checking MIN/MAX value for EX, T2, 

 and   , don’t replace probe 

otherwise you will get error code displayed. 

If you don’t plug the probe into socket before turning the meter on, the error code will also 

appear when you check MIN/MAX value. 

 

Note: 

Remenber Low Batteries will tend to give inaccurate readings, so make sure you have 

enough power. 

 

 

 

 &   

These functions are only for PCE-310. 

 

 is calculated by subtracting the T2 (Ex.temp.) from T1 (Air temp.) 
 is calculated by subtracting the DP (dewpoint) from T2 (Ex. Temp.) 

(1) Press 

 button until   appears on display. (See Fig.J) 

(2) Press 

 button until   appears on display. (See Fig.K) 

Summary of Contents for PCE-310

Page 1: ...1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire South...

Page 2: ...nable cap to protect sensor External Temp probe Triple LCD digital display T1 T2 T2 DP function Model 8716 Data Hold to capture readings Pocket size easy to fit in pocket Low battery indication Fast r...

Page 3: ...3 WARNING External probe port IMPORTANT BE SURE TO OPEN THE SENSOR PROTECTION CAP BEFORE STARTING THE MEASUREMENT IN ORDER TO GET ACCURATE VALUE FOR ALL MODELS CONTROLS AND INDICATORS...

Page 4: ...n itself off after 10 minutes To override Auto Power Off function press ON OFF HOLD while the meter is off When n appears See Fig A release HOLD button Meter is now in Non Sleep Mode For PCE 310 the a...

Page 5: ...ON Press button until DP appears on display Select to display dew point T1 air temp or delta T T1 T2 in any mode while the meter is on See Fig E WET BULB FUNCTION Press until Wbappear on the LCD User...

Page 6: ...than two seconds to clear current reading from memory WARNING While checking MIN MAX value for EX T2 and don t replace probe otherwise you will get error code displayed If you don t plug the probe int...

Page 7: ...s displayed as that means the values are invalid in the calibration process After 30 min the flashing will stop to indicate the procedure is finished 3 Move the sensor to 75 salt bottle and press more...

Page 8: ...auto power off function Er 1 Circuit error in RH measurement channel return the meter to the store for repairing Er 2 Circuit error in Internal temperature measurement channel return the meter to the...

Page 9: ...son When requiring a RA Return Authorization please include data regarding the defective reason the meters are returned along with good packing to prevent any damage in shipment and insured against po...

Reviews: