Calibration 

Before first use and after not using the hardness tester for a period of time, it must be calibrated 

by means of the Leeb test block. 

Go to calibration mode via menu item 7. 

x/5 

000 

Calibration 

 

1/5 

772          HL 

Calibration 

The  device  shows  values  in  HLDL.  In  order  to  calculate  your  HLD  value,  please  refer  to  the 

conversion chart in 10.5. 

Battery replacement 

 

Unscrew the cover on the back side of the display unit. 

 

Remove the battery. 

 

Insert the new battery. Observe correct polarity. 

 

Screw the cover back on. 

 

Switch on the device to find out if it works properly. 

10  Additional information 

10.1 Specifications of impact device 

Type of impact device 

D (PCE-2500N), DL (PCE-2600N) 

Impact energy 

Impact weight 

11 mJ 

5,5 g 

Ball hardness 

Ball diameter 

Ball material 

1600 HV 

3 mm 

Tungsten carbide 

Diameter of impact device 

Length of impact device 

Weight of impact device 

20 mm 

147 mm 

50 g 

10.2 Sample requirements 

Max. sample hardness 

940 HV 

Max. sample roughness Ra 

1.6 μm 

Min. sample weight 

Measurement without stabilisation 

Stabilisation  

Coupling necessary  

 

>5 kg 

2 ~ 5 kg 

0.05 ~ 2 kg 

Max. sample thickness 

Measurement without coupling 

Coupling necessary 

 

>5 mm 

≤5 mm 

Min. depth of hardened layer 

0.8 mm 

 

 

Measure 5 measuring points of the Leeb hardness block. 

The impact direction must be vertically down. 

 

After the measurement, the hardness tester will show the 

average value. 

Press  the  arrow  keys  to  set  the  average  value  to  the 

standard value of the test block. (You can set the value 

up  to  15  HL  higher  or  lower).  Confirm  with  the 

Menu/Confirm  key.  “Calibration  complete“  will  be 

displayed. 

Summary of Contents for PCE-2500N

Page 1: ...CE 2600N Durometer User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last cha...

Page 2: ...reparation of sample surface 8 7 2 Stabilisation of the sample 9 7 3 Prior to first measurement 9 7 4 Power on 9 7 5 Charging 9 7 6 Measurement 10 7 7 Readings displayed 10 8 Calibration 12 9 Battery...

Page 3: ...agnetic fields oil etc The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The...

Page 4: ...ed by 1000 The harder a material surface the higher the rebound velocity will be The Leeb hardness value represents the direct relation of a group of materials e g steel aluminium to its hardness prop...

Page 5: ...asure the hardness of a material with the conventional static test method the pressure you apply will affect the reading e g 720 HLD 720 HLC Since different impact devices correlate with different con...

Page 6: ...PCE Instruments 4 5 Device description PCE 2500N 1 Display 2 Keypad 3 Loading tube 4 D type impact body within guide tube 5 Micro USB charging socket 6 Support ring 1 2 3 4 5 6...

Page 7: ...PCE Instruments 5 PCE 2600N 1 Display 2 Loading tube 3 Keypad 4 DL type impact body within guide tube 5 Micro USB charging socket 1 2 3 4 5...

Page 8: ...he device it will show the main screen HL Hardness scale AVE Average value Impact direction Difference between PCE 2500N and PCE 2600N PCE 2500N has a D probe for even surfaces PCE 2600N has a DL prob...

Page 9: ...ess steel grey cast iron spheroidal graphite iron cast aluminium copper zinc alloy brass copper tin alloy copper bronze Accuracy HLD 0 5 800 HLD Repeatability HLD 0 8 800 HLD Resolution 128 x 64 OLED...

Page 10: ...the end of this manual When preparing the sample surface bear in mind that heat cold etc affect the hardness of the surface to be tested An excessively rough test surface can cause measurement errors...

Page 11: ...thickness are as required This causes incorrect readings In this case the test surface of the sample should be reinforced or supported The magnetic field of the sample should not exceed 30 Gauss 7 3 P...

Page 12: ...s of impact Distance from centre of point of impact to edge of sample 3 mm 5 mm At the end of the measurement the results can be seen on the display The hardness value is the average value of 5 measur...

Page 13: ...to Mean Times 5 1 10 poss 10 mat see specifications HL HV HB HRC HS HRB HRA 0000 9999 possible Menu Confirm to go to next digit 1 View File 2 Delete File 3 Select File No 4 Return Confirm DEL Select Y...

Page 14: ...600N Impact energy Impact weight 11 mJ 5 5 g Ball hardness Ball diameter Ball material 1600 HV 3 mm Tungsten carbide Diameter of impact device Length of impact device Weight of impact device 20 mm 147...

Page 15: ...0 4 Conversion chart materials Material Hardness scale Value Steel and cast steel HRC 17 9 68 5 HRB 59 6 99 6 HRA 59 1 85 8 HB 127 651 HV 83 976 HS 32 2 99 5 Steel HB 143 650 Alloyed tool steel HRC 20...

Page 16: ......

Page 17: ...726 HLD 518 521 524 527 530 532 534 537 540 544 546 548 551 554 557 560 563 566 569 572 574 HLDL 728 730 732 734 736 738 740 742 744 746 748 750 752 754 756 758 760 762 764 766 768 HLD 578 581 584 587...

Page 18: ......

Page 19: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Page 20: ...583 975 114 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english w...

Reviews: