6.6.3 

Sampling time / log interval 

 

Navigate to the advanced settings as described above. 

 

Press the "SET" key repeatedly until "SP-t" appears on the display. 

 

Select the desired log 

interval with the "▼" or "▲ / Time" keys and confirm the entry 

with the "LOG / Enter" key. The following can be selected: 
5, 10, 30, 60, 120, 300, 600 seconds and auto. 

 

Press the "SET" key repeatedly until you return to measuring mode or wait for 5 
seconds; then the meter will return to measuring mode automatically. 

 

Attention!  

"auto" means that each time the measured values change (>±10 digits), the values are saved. 
If the setting is 1 second, individual data records may be lost. 

6.6.4 

Key / control sound  

 

Navigate to the advanced settings as described above. 
Press the "SET" key repeatedly until "bEEP" appears on the display. 

 

Use the 

"▼" or "▲ / Time" keys to select yes or no. 

 

Confirm the selection with the "LOG / Enter" key. 

 

Press the "SET" key repeatedly until you return to measuring mode or wait for 5 
seconds; the meter will return to measuring mode automatically. 

6.6.5 

Decimal character  

 

Navigate to the advanced settings as described above. 
Press the "SET" key repeatedly until "dEC" appears on the display. 

 

Use the 

"▼" or "▲ / Time" keys to select "Euro" or "USA".  

"Euro" corresponds to the decimal comma and "USA" corresponds to the decimal 
point. In Europe, the "comma" is predominantly used as decimal character. 

 

Confirm the selection with the "LOG / Enter" key. 

 

Press the "SET" key repeatedly until you return to measuring mode or wait for 5 
seconds; the meter will return to measuring mode automatically. 

6.6.6 

RS232 output  

 

Navigate to the advanced settings as described above. 
Press the "SET" key repeatedly until "rS232" appears on the display. 

 

Use the 

"▼" or "▲ / Time" keys to select yes or no. 

 

Confirm the selection with the "LOG / Enter" key. 

 

Press the "SET" key repeatedly until you return to measuring mode or wait for 5 
seconds; the meter will return to measuring mode automatically. 

 

 

Summary of Contents for 4250348719621

Page 1: ...ssure Data Logger User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last chan...

Page 2: ...scope 2 5 Device description 3 6 Operation 4 6 1 Measurement preparation 4 6 2 Display information 5 6 3 Measuring logging 5 6 4 Backlight 5 6 5 Data transfer to the PC 6 6 6 Advanced settings 7 7 Bat...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...s of measurement bar psi kg cm mmHg inHg mH2O inH2O ATP kPa Input impedance 4 20 mA 10 Storage rate in seconds 5 10 30 60 120 300 600 Auto Memory SD card up to 16 GB Display LCD with illumination Refr...

Page 5: ...nput Mounting hole Opening Reset key Stand RS232 output Battery compartment SD card slot Battery compartment screw Display Measuring input channel 1 LOG Enter key Measuring input channel 2 Set key Mea...

Page 6: ...re that the decimal character is set correctly The default decimal character is a dot In Europe however the comma is customary If the decimal character is not set correctly in your country this can le...

Page 7: ...s are displayed 6 3 1 Starting the log function To start the logger press and hold the LOG Enter key for 2 seconds Scan quickly appears in the upper part of the display as a confirmation Datalogger ap...

Page 8: ...n SD card reader connected to the PC Start the spreadsheet programme on your PC open the file on the SD card and read out the data 6 5 1 SD card structure The following structure is automatically crea...

Page 9: ...selection with the LOG Enter key If you select yes you must confirm the security prompt again with the LOG Enter key Press the SET key repeatedly until you return to measuring mode or wait for 5 seco...

Page 10: ...ion with the LOG Enter key Press the SET key repeatedly until you return to measuring mode or wait for 5 seconds the meter will return to measuring mode automatically 6 6 5 Decimal character Navigate...

Page 11: ...lly Attention With the measurement range setting of 200 bar with the unit of measurement mmHg and with the measurement range setting of 400 bar with the unit of measurement mmHg and inH2O the measured...

Page 12: ...numbers stand for the following parameters D15 Start word D14 4 D13 When upper display data is sent 1 is sent When middle display data is sent 2 is sent When lower display data is sent 3 is sent D12...

Page 13: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Page 14: ...tanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 0583 975 114 Tel 0212 471 11 47 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 Faks 0212 705 53 93 info pce instruments co uk...

Reviews: