18  Maintenance 

 

18.1 

Replacing the Batteries 

 

(1) Remove back cover screw to remove cover. Verify polarity and install new LR03(AAA 

size) alkaline batteries. 

(2) Fit cover properly and tighten screw. 

 

 
 
 
 

 

 

18.2 

Cleaning 

 

In  order  to  ensure  the  accuracy  of  the  thermometer  for  a  longer  period  of  time  you  should 
calibrate it once a year. 

Clean the device and the window of the display with a clean, lint-free, antistatic and dry cleaning 
cloth. 

 

 
 
 
 
 

 

The  unit’s  back  cover  is  fitted  with  rubber  rings.  After  replacing  the  battery, 
check  that  the  rubber  rings  are  properly  seated  before  reinstalling  the  back 
cover. Improper seating of the rubber rings will compromise the unit’s water-
resistant structure, and possibly result in damage to the equipment. 

Do no use cleaning agents that contain carbon or benzenes, alcohol or anything similar to 
clean the product since these substances damage the surface of the measuring instrument. 
Moreover, these fumes are hazardous to health and explosive. Do not use tools with sharp 
edges, screwdrivers, metal brushes or anything similar to clean the device. 

Summary of Contents for PCE-T 318

Page 1: ...ometer PCE T 318 User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last chang...

Page 2: ...AVG Operation 6 11 Auto Power Off 6 12 Low Battery Condition 6 13 Temperature Measurement 7 13 1 Correct Measurement Method 7 13 2 Connector Configuration 7 14 Temperature Meter Calibration Setup 8 15...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...and Humidity 0 C 50 C 32 F 122 F 0 80 RH Storage Temperature and Humidity 10 C 60 C 14 F 140 F 0 80 RH Altitude up to 2000 meters Accuracy 148 F 572 F 0 1 reading 0 8 F 100 C 300 C 0 1 reading 0 4 C...

Page 5: ...xx Platinum Type Indication 4 1 1 1 1 1 1 HOLD This indicates that the display data is being hold MAX The Maximum value is now being displayed MIN The Minimum value is now being displayed AVG The Aver...

Page 6: ...nnector 2 LCD display 3 Power ON OFF button 4 HOLD button 5 MAX MIN Average control button 6 Back light button 7 Relative readout button 8 C F control button 5 Operation Instructions 5 1 Power Up Pres...

Page 7: ...button When the hold data is no longer needed one may release the data hold operation by pressing HOLD button again When the meter is under Data Hold operation the REL and C F button are disabled 8 Ba...

Page 8: ...g is the present temperature One may press MAX MIN to circulate the display mode among these options When the meter is under MAX MIN operation REL and C F are disabled To exit the MAX MIN mode one may...

Page 9: ...he temperature sensor is located at the end of the metal sheath of the sheath type TEMPERATURE PROBE To accurately test internal temperature insert the probe into the item you want to measure to a dis...

Page 10: ...er Calibration Setup Below table diagram shows each button function when user enter into calibration mode Note Turn the Power OFF before attempting following SETUP Setup mode is cancelled during below...

Page 11: ...ion Press and hold Power Light buttons for 2 seconds to enter setup mode Press Hold button to enter Pt selection mode Press Max button to select between Pt probe type Pt100 Pt500 Pt1000 is circulate P...

Page 12: ...ibration Press and hold Power Light buttons for 2 seconds to enter setup mode Press REL button to enter calibration mode Enter Pt selection first Press C F button to confirm Pt selection Press REL but...

Page 13: ...irm present calibration value Press C F button to confirm selection 17 Recall Default factory setting value Press and hold Power Light buttons for 2 seconds to enter setup mode Press REL button to ent...

Page 14: ...nstruments 12 Press C F button to confirm Pt selection Press Light button to enter recall mode Press C F button to confirm revert back to Default factory setting value Press C F button to confirm sele...

Page 15: ...oth The unit s back cover is fitted with rubber rings After replacing the battery check that the rubber rings are properly seated before reinstalling the back cover Improper seating of the rubber ring...

Page 16: ...t 100 4 wires Accuracy IEC751 class A 0 15 C 0 002t t measurement temperature Measurement Range 100 to 400 C Temperature Sensor Dimensions Approx 3 2mm 0 125 Temperature Sensor Length Approx 120 mm 4...

Page 17: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Page 18: ...583 975 114 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english w...

Reviews: