5. DATALOGGER
5-1 Preparation before execute datalogger function
The dose meter measured data first be stored into the meter internal
built EEPROM, can be stored total of 16,000 sets of data, when stored
> 16000 sets , datalogger will be stop and display will show FULL.
Remark :
How to send the storage of information, you have two choices.
1) If select data send to SD card , please use AC adapter . refer to
Chapter 5-2 d.--a,b,c. page 12,13.
2) If select the logger data send to RS232 , please press the
power button > 2 sec. refer to Chapter 5-2 d.---c. page 13.
a. Time setting
If the meter is used at first time, it should to adjust the
clock time exactly, please refer chapter 7-1 ( page 17 ).
b. Decimal format setting
The numerical data structure of SD card is
default used the " . " as the decimal, for
example "20.6" "1000.53" . But in certain
countries ( Europe ...) is used the " , " as the
decimal point, for example " 20, 6 "
"1000,53". Under such situation, it should
change the Decimal character at first, details
of setting the Decimal point, refer to Chapter
7-5, page 19.
5-2 Auto Datalogger ( Set sampling time
≧
≧
≧
≧
1 second )
a. Start the datalogger
Pressing and holding the " ENTER Button "( 3-7, Fig. 1 ) once ,
display is shown " LOGGER " , the " LOGGER " will flashing
at the same time the measuring data along the time information
will be saved into the memory circuit(EEPROM).
and the top right will record items.
11