9.3.2
Calibration with one calibration standard
The calibration with only one calibration standard is similar to the calibration with two standards. The
difference is that you do not calibrate the “TEST-H” part. Press the “MENU” key after “TEST-L” calibration
and enter the nominal value. Finish the calibration by pressing “MENU” again.
9.3.3
Batch-Mode
Within the calibration settings, you can also enter the tolerance thresholds. If you have entered the upper
and lowest value, the instrument shows if the measured value is within that area every time you measure.
If the value is within that area the instrument shows “PASS”. If the value is not within that area it will
display “NG”.
To set the tolerances, use the arrow down key after you have set the “TEST-L” and “TEST-H” values. You
will now get to the points “UP” and “DOWN”. Up stands for the upper threshold, DOWN the lower value.
To set
the values is made equal to setting “TEST-L and “TEST-H”.