14
- for peak level (L
Pk
)
WARNING! The extents of measurement range for peak level at A and Z
frequency weightings are determined for continuous 1000Hz sine signal;
measurement error extended by expanded measurement uncertainty is less than 1.1
dB.
Self-generated noise
- measured with electrical input facility installed:
° at A frequency weighting
13dB
° at C frequency weighting
19dB
° at Z frequency weighting
25dB
- measured with WK-21 microphone installed:
° at A frequency weighting
19dB
° at C frequency weighting
24dB
° at Z frequency weighting
27dB
1.5.2 1/1-octave analyzer mode
Frequency weightings
A, C, Z
AC electrical output (sine, 1000 Hz)
- weighted frequency response
A, Z
- FSD output voltage for level range
14V
p-p
- maximum output voltage 22V
p-p
- output impedance
1k
Ω
- load impedance
10k
Ω
Time weightings
SLOW, FAST
Extents of measurement range at specified level range and frequency weighting, for
time-weighted sound level L and time-average sound level L
eq
(analyzer):