Operation 

 

5.1 

Measurement 

 

Edition with internal sensor: 

After  making  the  settings,  you  can  start  a 
measurement.  To  do  so,  you  first  need  to 
attach the suitable adaptor. If required, you can 
also  use  the  extension  rod.  Do  not  apply  too 
much force when attaching the adaptor and / 
or  the  extension  rod  to  avoid  damage  to  the 
sensor. 

Edition with external sensor: 

Connect the sensor to the force gauge. 
Then attach the suitable adaptor to the sensor 
without  applying  too  much  force  as  this  can 
damage the sensor. 

 

Switch on the device by pressing the On / Off key. You are now in the measurement window. Start 
by checking the battery level in the upper right corner. If the battery level is low, charge the device 
using the mains adaptor that comes with the device. For charging, it does not matter if the device is 
switched on or off. It is possible to make measurements during charging. When the device  is fully 
charged, a notification will appear on the display. You should then disconnect the device from the  
power supply immediately. 

 

You  can  now  set  the  parameters.  Press  “OK”  when  you  are  in  measurement  mode  to  enter  the 
settings window. Set the unit, the force area, zero tracking, the sampling speed, fracture alarm, the 
upper and lower limit alarm, the Peak Hold function, the capture trigger as well as the capture length 
(duration). Press the Back key to return to measurement mode. 

 

To start your measurement, attach the device to a suitable test stand. Press the Zero key and the  
Arrow Right key. You can now take a direct measurement or capture a curve. If you make a direct 
measurement,  the  force  will  be  measured  in  real  time,  as  well  as  the  peak  values  and  further 
parameters.  These  will  not  be  saved.  The  previously  measured  values  will  be  lost  when  a  new 

measurement is taken. If you want to capture a curve, enter capture mode by pressing the “Arrow 
Left” key when you are in measurement mode. The measurement will start automatically as soon 
as the trigger condition is met. The measurement ends if you press “OK“ or when the set capture 
duration has been reached. The measurement  will return peak values, measurement curves and 
further parameters. 

 

These will be saved. Only one curve with the associated values can be saved. You can retrieve the 
saved data by pressing the “Query“ button. The curve will be deleted when the force gauge is re- 
started or when a new measurement is taken. A maximum of 100 reports can be saved if no curve 
is saved. These can also be retrieved by pressing the 

“Query“ button. 

 

Press the Back key to get  back to the measurement  window. To switch off the device, press the 
On/Off key. Remove the sensor if you have a device with an  external sensor and clean the force 
gauge. It is recommended to store the device in its original carrying case. 

E

n

g

li

s

h

 

Summary of Contents for CAL-PCE-DFG N

Page 1: ...is taliano espa ol portugu s nederlands t rk polski can be downloaded here www pce instruments com Son de i iklik last change 3 March 2021 V 1 5 i Dinamometre Digital Force Gauge PCE DFG N Serisi Kull...

Page 2: ...2 Teslimat eri i 3 2 3 Opsiyonel Aksesuarlar 4 3 Sistem A klamas 5 3 1 Cihaz 5 3 2 Aray z 5 3 3 Ekran 6 3 4 Fonksiyon tu lar 7 4 Kullan ma Haz rl k 7 4 1 G kayna 7 4 2 Ayarlar 8 5 Kullan m 12 5 1 l m...

Page 3: ...contents 17 2 3 Optional accessories 18 3 System description 19 3 1 Device 19 3 2 Interfaces 19 3 3 Display 20 3 4 Function keys 21 4 Getting started 21 4 1 Power supply 21 4 2 Settings 22 5 Operatio...

Page 4: ...cihaz n l lebilir maksimum de eri uygulanan darbe y k n n iki kat kadar olmal d r Darbe testleri ger ekle tirirken yaralanmalar n n ne ge mek i in maske ve koruyucu eldiven giyilmelidir Test stand b...

Page 5: ...G N 10 PCE DFG N 20 PCE DFG N 200 PCE DFG N 500 l m Aral 0 5 N 0 10 N 0 20 N 0 200 N 0 500 N z n rl k 0 001 N 0 002 N 0 005 N 0 01 N 0 05 N Boyut 200 x 97 x 42 mm A rl k Yakla k 540g Harici l m h crel...

Page 6: ...ar 10 50 C 5 95 n o yo u mas z Koruma S n f IP 54 2 2 Teslimat eri i Dahili h creli versiyon Harici h creli versiyon 1 x Dinamometre PCE DFG N 5 x l m Ucu Adapt rleri Sivri Kafa Adapt r entikli Kafa A...

Page 7: ...N set 2 par a PCE C A 2500N EB G zl civata M12 2 500 N set 2 par a PCE C A 5000N EB G zl civata M12 5 000 N set 2 par a PCE C A 10000N EB G zl civata M12 10 000 N set 2 par a PCE C A 60KN EB G zl civ...

Page 8: ...i h creli versiyon 1 l m mili 6 Sivri kafa adapt r 2 Ekran 7 Kanca adapt r 3 Klavye 8 entikli kafa adapt r 4 Uzatma ubu u 9 D z kafa adapt r 5 Keski kafa adapt r 10 Uzatma ubu u i in adapt r 3 2 Aray...

Page 9: ...rici giri k anahtar 2 k alt s n r 3 k st s n r 4 GND topraklama 1 Haf za kullan m 7 Pil seviyesi g stergesi 2 l m e risi 8 Ayarlanm rnekleme h z 3 Kuvvet de eri 9 Alt alarm de eri 4 st alarm de eri 10...

Page 10: ...naylama Sol E ri kayd na ba lama E ri kayd na ba lama Yan p s nen say y bir basamak sola ta ma Sa Peak de erini silme Peak de erini silme Yan p s nen say y bir basamak sa a ta ma 4 Kullan ma Haz rl k...

Page 11: ...r S f r Noktas Zero Tracking S f r noktas i in a a daki de erler ayarlanabilir Kapal 0 1 N 0 2 N 0 3 N 0 4 N 0 5 N S f r noktas stabilizasyonundan nce ayarlanan de erin alt ndaki de erler otomatik ola...

Page 12: ...girmek Kullan c kaydedilen kalibrasyon verilerini girer Kalibrasyon herhangi bir cihaz ya da a rl k olmadan yap l r 2 Standart kalibrasyon Cihaz kalibrasyon stand ya da kalibrasyon a rl arac l yla kal...

Page 13: ...m engellenir Tu Komut Yukar tu una bast n zda ya da k komutu al nd nda bir kez k ger ekle ir Cihaz bilgisayara ba land nda bilgisayardaki programlar k fonksiyonunu otomatik olarak devre d b rak r De i...

Page 14: ...ri bilgisayara aktar m s ras nda kaybolur l m sonu lar cihaz n i inde kal r G r nt leme A s Display Angle Burada g r nt leme a s n ayarlayabilirsiniz 0 ya da 180 i se ebilirsiniz Otomatik Arka Ayd nla...

Page 15: ...iyi ve kay t s resini ayarlay n l m moduna geri d nmek i in Geri tu una bas n l me ba lamak i in cihaz uygun test stand na yerle tirin S f r ve sa ok tu una bas n Do rudan l m ger ekle tirebilir ya da...

Page 16: ...m H1 mm M mm A rl k 1 kN 50 8 76 2 20 8 15 3 M10 x 1 5 540 g 2 5 5 kN 50 8 76 2 20 8 13 6 M12 x 1 75 540 g 10 kN 50 8 76 2 27 2 12 1 M12 x 1 75 680 g 20 50 kN 76 2 101 6 27 2 20 M20 x 1 5 1350 g 100 k...

Page 17: ...bilgilendirmek zorunday z Kullan lm piller evsel at k olarak at lmamal d r T keticiler pilleri t ketici i in cretsiz olan uygun bir ticari veya belediye toplama noktas na vermekle y k ml d r Kullan lm...

Page 18: ...TION For impact tests the maximum measurable value of the force gauge should be twice as high as the applied impact load When doing impact tests wear a mask and protective gloves to avoid injuries Do...

Page 19: ...PCE DFG N 10 PCE DFG N 20 PCE DFG N 200 PCE DFG N 500 Range 0 5 N 0 10 N 0 20 N 0 200 N 0 500 N Resolution 0 001 N 0 002 N 0 005 N 0 01 N 0 05 N Accuracy Dimensions 200 x 97 x 42 mm Weight approx 540...

Page 20: ...Output port 12 V 50 mA Operating conditions 10 50 C 5 95 RH non condensing Protection class IP 54 2 2 Delivery contents Version with internal cell 1 x force gauge PCE DFG N 5 x test head adaptors poin...

Page 21: ...set 2 pieces PCE C A 5000N EB eyebolt M12 5 000 N set 2 pieces PCE C A 10000N EB eyebolt M12 10 000 N set 2 pieces PCE C A 60KN EB eyebolt M20 60 000 N 6 t set 2 pieces PCE C A 120KN EB eyebolt M30 12...

Page 22: ...metric cell 1 Measuring shaft 6 Pointed head adaptor 2 Display 7 Hook adaptor 3 Keypad 8 Notched head adaptor 4 Extension rod 9 Flat head adaptor 5 Chisel head adaptor 10 Adaptor for extension rod 3 2...

Page 23: ...t 2 Output lower limit 3 Output upper limit 4 GND 3 3 Display In measurement mode 1 memory usage 7 battery level indicator 2 measurement curve 8 set sampling rate 3 force value 9 lower alam value 4 up...

Page 24: ...e capturin g Move flashing number left by one digit Right Delete value peak Delete peak value Move flashing number right by one digit 4 Getting started 4 1 Power supply The PCE DFG N is equipped with...

Page 25: ...set Off 0 1 N 0 2 N 0 3 N 0 4 N 0 5 N Values below the value set are automatically excluded before the zero point stabilises After stabilisation of the reading the sampling rate will be 1 x per second...

Page 26: ...d calibration The force gauge is calibrated by means of the calibration stand or calibration weight User Gravity Here you can set the gravity at the place of use The value can be between 9 700 and 9 9...

Page 27: ...ata transfer The following parameters can be set Prohibit The real time data transfer of the serial interface is prohibited Key Order A single output will take place when you press the Up key or when...

Page 28: ...00 bps 1600 Hz 230400 bps Due to the limited speed of serial interfaces some data get lost when transferred to a PC if the sampling rate is higher than 800 Hz The readings will however not get lost in...

Page 29: ...ure trigger as well as the capture length duration Press the Back key to return to measurement mode To start your measurement attach the device to a suitable test stand Press the Zero key and the Arro...

Page 30: ...nents 7 Appendix load cell 1 100 kN L mm H mm W1 mm H1 mm M mm 1 kN 50 8 76 2 20 8 15 3 M10 x 1 5 2 5 5 kN 20 8 76 2 20 8 13 6 M12 x 1 75 10 kN 50 8 76 2 27 2 12 1 M12 x 1 75 20 50 kN 76 2 101 6 27 2...

Page 31: ...lutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devices b...

Page 32: ...Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece stanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 05...

Page 33: ...ts 30 PCE Instruments Specifications are subject to change without notice User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be downloaded here www pce in...

Reviews: