PCB Piezotronics M209C12 Installation And Operating Manual Download Page 11

 

 

 

 

 

ICP

 FORCE SENSOR OPERATION MANUAL 

 

 6 

Figure 12 - Cable Strain Relief 

 

 

 

 

FORCE RING INSTALLATION 

The sensor is mounted using the supplied mounting stud and 

pilot bushing.  The supplied beryllium copper (BeCu) stud is 

elastic so it allows force transmission to the sensor while 

holding the sensor in place.  The pilot bushing centers the sensor 

about the mounting stud.  After installing the mounting stud in 

the lower of the two surfaces, the pilot bushing is threaded over 

the mounting stud.  The sensor is then placed over the stud and 

pilot bushing combination.  The pilot bushing should fit loosely 

inside of the sensor inner diameter, holding it in place.  Properly 

machined holes for the mounting stud will ensure proper 

vertical orientation of the sensor.  The upper surface should be 

installed and tightened onto the mounting stud.  Refer to the 

installation drawing for additional mounting details.   

 

When installing the sensor as an integrated member, it is 

recommended that the supplied antifriction washers be used to 

eliminate the possibility of damage to the sensing surface of the 

sensor.  This type of damage may occur when imperfections in 

the mounting surface grind against the sensor surface while the 

mounting surfaces are being twisted during installation.  Refer 

to the installation drawing for additional mounting details. 

 

 

PRELOAD REQUIREMENTS FOR FORCE RING AND 

3-COMPONENT FORCE SENSORS  

 

PCB ICP

 Force Rings (Models 201B01 through 207C) and 3-

Component Force Sensors (Models 260A01 through A03) are 

generally installed between two parts of a test structure, as 

shown in 

Figure 4

.  During installation, the sensor should be 

pre-loaded to the amount specified on the specification sheet 

using the supplied elastic beryllium-copper stud.   Preloading in 

this arrangement ensures that the sensor will perform as 

calibrated and have good output linearity at the sensor

s lower 

operating range.  Use of a mounting stud of stiffer material or 

no stud will alter PCB calibrated sensitivity. 

 

The stud or bolt holds the structure together and applies preload 

to the force ring as shown in 

Figure 13

.  In the typical link 

installation, shown on the left in 

Figure 13

, part of the force 

between the two structures is shunted through the mounting 

stud.  The amount of force shunted is dependent on the stud 

material, but may be up to 5% of the total force for the 

beryllium-copper stud supplied with the sensor and up to 50% 

for steel studs.  This typical installation setup is used by PCB 

during standard calibrations of force ring models 201B01 

through 207C.    PCB’s standard calibration, when using the 

BeCu stud, takes this shunted force into account with the final 

calibration value. 

 

A non-typical installation is shown on the right side of 

Figure 

13

.  In this installation, the stud or bolt used to apply the preload 

does not shunt part of the applied force.  The plate on top of the 

sensor has a clearance hole that the stud or bolt passes through.  

In this installation, the stud or bolt is not directly connected to 

the top plate by its threads, as it is in the typical installation, so 

it does not shunt any force. This method of installation 

accommodates mounting a group of sensors under a common 

plate. 

 

NOTE: If any of the following conditions apply to the 

preloading of the force ring in the actual application, the 

sensitivity and linearity performance of the sensor will not 

match the standard PCB calibration values. 

 

1.

 

Use of a stud or bolt other than the supplied beryllium-

copper stud. 

2.

 

Use of no stud or bolt. 

3.

 

Use of an amount of preload other than the 

recommended amount. 

4.

 

Use of installation that is different from PCB setup 

during calibration. 

5.

 

On  rare occasions, a ring sensor may be installed 

WITHOUT a mounting stud (as installed in precise 

fixture).  In these cases, the sensor sensitivity will be 

HIGHER than that shown on the PCB calibration 

certificate. 

 

In these cases, please contact a PCB applications engineer at 

800-828-8840 of the intended installation to discuss your 

special calibration requirements.

  

PCB can calibrate the 

sensor without a stud OR ask that a customer supplied 

mounting stud be provided for use in PCB’s calibration 

process. 

 

 

 

Figure 13 - Force Ring Sensor Installations 

 

PCB in-house calibration procedure requires the installation of 

a force ring with BeCu stud, in the typical installation setup 

above.  This sensor  is then placed  in series with a NIST 

Typical Link Installation 

Non-Typical Installation 

Used for Force Limited 

Vibration Testing 

Summary of Contents for M209C12

Page 1: ...2 2 lb comp 1 lb tension 2200 Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 F...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...tallation outline drawing and specification sheet at the back of this manual for details and dimensions of the particular sensor model number s purchased The following pages give a brief description o...

Page 7: ...ring mounting and preload requirements Figure 4 outlines some possible mounting configurations to which the ring series may be installed Figure 4 Series 201 to 207 ICP Ring Force Sensor Possible Inst...

Page 8: ...simultaneously measuring force in three 3 orthogonal directions X Y and Z Internally three sets of quartz crystals are cut oriented and preloaded so the maximum output from the crystal cut for the spe...

Page 9: ...eturning the sensor to PCB for recalibration Figure 8 SERIES 3 COMPONENT TRIAX FORCE SENSORS PENETRATION Penetration style sensors are similar to the axial models but are specifically designed for com...

Page 10: ...t facing or surface grinding Surface flatness should be held to within 0 001 TIR over the entire mating surface When mounting sensors between two plates care should be taken to assure mounting surface...

Page 11: ...he left in Figure 13 part of the force between the two structures is shunted through the mounting stud The amount of force shunted is dependent on the stud material but may be up to 5 of the total for...

Page 12: ...itor the output from the Z axis connector when preloading 3 component type sensors These sensors require a preload in the Z axis that is 10 times their shear range Some models require this to be accom...

Page 13: ...hundreds of feet meters depending on the frequency of interest It is necessary to supply the sensor with a 2 to 20 mA constant current at 20 to 30 VDC through a current regulating diode or equivalent...

Page 14: ...ant in the system For this reason the readout device as well as the signal conditioner must be DC coupled Figure 16 Characteristic Discharge Time Constant Curve The discharge time constant represents...

Page 15: ...urned to the factory for servicing in the event of serious malfunction Observe the following precautions in using the sensor A Do not exceed the maximum load levels for the force sensor see specificat...

Page 16: ...ss Steel Sealing Hermetic Hermetic Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Electrical Connection Position Side Side Mounting Thread 10 32 Female No Metric Equivalent All specificati...

Page 17: ......

Reviews: