background image

 

Drawing Number: 21096 
Revision: NR

 

 

SERIES 134A20 TOURMALINE PRESSURE BAR GENERAL OPERATION MANUAL

 

1.0 

INTRODUCTION

 

 
Series  134A20  Tourmaline  Pressure  Bar  with  In-
Line  Amplifier  consists  of  a  Series  134A 
Tourmaline Pressure Bar connected to a three-inch, 
low-noise  cable  that  connects  into  Series  402  In-
Line  Source  Follower  Voltage  Amplifier.    See 
Figure 1.  
 

 

 
The  Tourmaline  Pressure  Bars  is  a  high-pressure, 
fast  rise  time,  suppressed-resonance  blast  sensor 
ideal for single-shot, high-frequency measurements 
of incident or reflected shock wave pressures found 
in  studies  of  plasma  physics  and  hyper-sonics. 
Ranges up to 20,000 psi are available. 
 
The Series 134A20 is typically powered by a PCB 
Series  480  ICP

®

  Power  Conditioner  (constant 

current)  or  equivalent.  Power  and  signal  is 
conducted  over  a  single-conductor  coaxial  cable 
with  the  shield  serving  as  a  signal  return.  Output 
from  the  power  conditioner  is  usually  fed  into  a 
high-speed  readout  instrument,  such  as  a  digital 
oscilloscope. 
 

2.0 

DESCRIPTION 

 
The  Series  134A20  Tourmaline  Pressure  Bar  with 
In-Line  Amplifier  consists  of  three  components 
attached  together  and  calibrated  as  a  system.  The 
Series  134A  Tourmaline  Pressure  Bar  senses  and 
transduces  a  pressure/shock  wave  into  a  high-
impedance, electrostatic charge. This charge is sent 
through  the  three-inch,  low-noise  cable  into  the 
Series  402A  In-Line  Amplifier,  where  it  is 
converted  into  a  low-impedance  voltage  with  less 
than 100 ohms output impedance.  See Figure 2. 

The  circuit  in  the  in-line  amplifier  consists  of  a 
MOSFET  input  IC  with  a  very  high  trans-
conductance, resulting in a voltage gain that is very 
close to unity. The source terminal "follows" the  
gate  voltage  instantaneously,  without  a  shift  in 
phase (i.e., the amplifier is non-inverting). 
 

 

 

2.1 

SYSTEM OUTPUT 

 
The  system  output  of  Series  134A20  is  a  low-
impedance  output  governed  by  the  following 
equation: 
 

3

C

2

C

1

C

Q

V

 

 
where:    



V = voltage sensitivity of system (V/psi) 



Q = charge output of Tourmaline Bar (pC/psi), 
Model 134A 



C1  =  capacitance  of  Tourmaline  Bar  (pF), 
Series 134A 



C2  =  capacitance  of  three-inch  cable  (pF), 
Model 003A000-03 



C3  =  capacitance  of  input  of  in-line  amplifier 
(pF), Series 402A 

 
A  system  calibration  is  supplied  with  Series 
134A20 that shows the system output in mV/psi. 
 

3.0 

INSTALLATION 

 
See  the  Installation  Drawing  included  in  the  back 
of  this  manual  for  a  description  of  the  Tourmaline 
Bar  and  its  installation. 

Install  in  a  gas  medium 

where  the  measurement  is  to  be  taken.  The  sensor 
can not be normally used in a liquid medium.

 

Summary of Contents for M134A27

Page 1: ...25 mV psi 0 2 microsecond rise Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 F...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...es 134A Tourmaline Pressure Bar senses and transduces a pressure shock wave into a high impedance electrostatic charge This charge is sent through the three inch low noise cable into the Series 402A I...

Page 7: ...r conditioner or in the power conditioner Allow the system to stabilize for a minute A signal drift may occur when connecting the power conditioner to the readout instrument This should stabilize as c...

Page 8: ...IER The Series 402A In Line Amplifier is used to convert the high impedance output of the tourmaline pressure bar into a low impedance output The in line amplifier may also be used as a unity gain amp...

Page 9: ......

Page 10: ......

Reviews: