Drawing Number: 21210 
Revision: NR

 

 

 

 

AN IMPROVED TECHNIQUE FOR  

UTILIZATION OF CONFORMAL BALLISTICS  

SENSOR CALIBRATION DATA

 

 1 

 
The  PCB  Series  117B  Conformal  Ballistics  Sensor 
(patented)  measures  chamber  pressure  in  a  novel 
manner. 
 
The  sensitive  surface  of  the  sensor,  when  properly 
installed  in  the  gun  chamber,  conforms

 

or  blends  in 

precisely with the inside surface of the chamber actually 
forming  a  portion  of  the  wall  adjacent  to  the  cartridge 
case. 
 
When  the  round  is  fired,  the  pressure  builds  up  rapidly 
inside  the  case,  obturating  the  case  outward  against  the 
chamber.    The  force,  transmitted  through  the  thin 
cartridge case wall stresses the sensor crystals producing 
an output charge analogous to chamber pressure. 
 
An  exclusive  calibration  technique  has  been  developed 
in conjunction with the conformal sensor, consisting of a 
simulated gun chamber which positions the sensor at the 
same  location  as  in  the  test  barrel  and  provides  for 
hydraulic pressurizations of an actual cartridge case. 
 
Using  this  system  and  a  reference  standard  pressure 
source,  a  point  by  point  calibration  is  performed  on 
samples  of  the  actual  lot  of  ammunition  under  test  to 
take  into  account  transmissibility  characteristics  of  the 
cartridge case for each lot of ammunition. 
 
Because a certain amount of pressure is required to fully 
obturate  the  cartridge  case  in  the  test  chamber,  a rather 
severe  non-linearity  is  exhibited  over  the  first  several 
thousand  psi  of  input  pressure,  i.e.  until  the  cartridge 
case is pressed firmly against the inside chamber walls. 
 
Figure  1  illustrates  a  typical  calibration  result  with  .22 
caliber rimfire ammunition. 

 

 

Figure 1:  Typical calibration graph 

Conformal sensor with .22 caliber rimfire. 

 
As  shown  in  figure 

1, 

the  best  straight  line that can  be 

drawn  through  the  majority  of  the  data  points  in  the 
linear  portion  of  the  curve,  does  not  pass  through  the 
origin. 

 

Because  of  this,  the  widely  accepted  zero  based  best 
straight line method of linearity determination cannot be 
applied  to  the  conformal  sensor.    Also,  the  actual 
sensitivity  of  this  sensor  may  only  be  specified  at  one 
input  pressure  level  since  the  sensitivity  is  not  a 
constant, but varies with input pressure level. 
 
This  paper  outlines  a  simple  alternative  approach  to 
utilization of conformal sensor data. 
 
The  most  universally  accepted  method  of  defining 
conventional  sensor  nonlinearity  is  the  full  scale  error 
band  method  used  in  conjunction  with  the  zero  based 
best straight line. 

Summary of Contents for M117B30

Page 1: ...RESSURE SENSOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...gement allows for tolerance in the location of the guide pin hole in an axial direction on the test barrel or calibration adaptor The precise depth adjustment is obtained by the use of the correct thi...

Page 7: ...ct a thinner spacer and remount Once the proper thickness is found for perfect flushness the sensor may be removed and reinstalled using this same spacer and the proper depth will be achieved each tim...

Page 8: ...to allow easy insertion and extraction the case must be expanded slightly by the internal pressure before force can be transmitted to the sensor See the enclosed guide An Improved Technique for Utili...

Page 9: ...Drawing Number 21086 Revision NR OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B 4...

Page 10: ...Drawing Number 21086 Revision NR OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B 5...

Page 11: ...munition under test to take into account transmissibility characteristics of the cartridge case for each lot of ammunition Because a certain amount of pressure is required to fully obturate the cartri...

Page 12: ...r two percent of full scale is specified for quartz sensors An acceptable instrument is then defined as one whose data points all fall within this error band The sensitivity of an instrument so calibr...

Page 13: ...nce between the cartridge case and the chamber wall Obturation of the cartridge occurs in this region However when obturation is complete the sensor output is then linear with pressure from this point...

Page 14: ...e data from 20 consecutive calibration runs performed on a conformal sensor yielded a slope of 04 pC psi and an intercept of 7 500 psi determined by constructing the best fitting straight line as desc...

Page 15: ...can be further refined to include a variable voltage source as a bias in series with the output of the charge amplifier This voltage bias can be set to simulate the offset pressure and then the true p...

Page 16: ...cale pressure to find full scale picocoulombs 40 000 psi x 250 pC psi 10 000 pC Set the charge calibrator for 10 000 pC and apply this signal to the input of the amplifier Now adjust the gain of the o...