PCB Piezotronics M108A02 Installation And Operating Manual Download Page 9

 

Drawing Number: 21113 
Revision: A 

 

OPERATION MANUAL FOR 

ICP

®

 HIGH PRESSURE SENSORS 

Series 108 & 109 

3.7 

CIRCUIT CONNECTIONS

 

 

The  figures  below  show  two  typical  circuit 
connections normally used to connect the 108 & 109 
Series to power units.   

 

The  first  figure  is  the  connection  scheme  used  for 
most  applications.    In  this  set  up,  the  signal  is  AC-
coupled from the bias voltage meaning the voltage at 
the "scope" terminal will be at a zero volt bias level. 

 

 

Normal Connection AC-Coupled Output

 

 

In  this  case,  the  system  discharge  time  constant  will 
be determined by the input resistance and capacitance 
of  the  power  unit.    Most  PCB  power  units  have 
approximately a 10 second discharge time constant. 
 

Alternate Connection to take Advantage of Sensor 
Discharge TC

 

 

The  second  figure  illustrates  an  optional  connection 
scheme  which  can  be  used  during  calibration  to  take 
full advantage of the sensor’s discharge time constant. 

 

 

 

With  this  arrangement,  the  readout  is  direct-coupled 
to  the  sensor  and  the  output  signal  will  be  added  to 
the DC bias voltage (11 VDC nominal). 

 

The  discharge  time  constant  will  then  be  determined 
only by the sensor and should be in the order of 2000 
seconds;  this  is  sufficient  for  most  types  of  static 
calibration methods. (See Calibration Section 4.0) 

    

 

Typical Connection

 

 

The  figure  above  shows  the  108  &  109  Series 
connected to a PCB Model 484B06 power unit.  The 
484B06  is  a  power  unit  that  can  be  AC  or  DC-
coupled. 

 

In  the  DC-couple  mode,  a  level-shifting  circuit 
removes  the  bias  level  and  provides  a  zero  adjust 
feature  at  the  front  panel.    The  system’s  discharge 
time  constant  is  determined  by  the  sensor  in  this 
mode. 

 

An  AC-coupled  mode  is  provided  for  normal  AC-
coupled operation. 
 

4.0 

CALIBRATION 

 

PCB  108  &  109  Series  sensors  are  supplied  with  a 
calibration 

certification 

from 

the 

factory.  

Recalibration services are provided at the factory for 
a nominal fee. 

 

Static calibration methods may be employed using the 
set-up shown in the Alternate Connection figure or by 
use of the Model 484B06 Power Unit as shown in the 
figure above – use the 484B06 in DC mode. 

 

Following  thermal  stabilization  of  the  sensor,  use  a 
high  pressure  pump  with  dial  reference  sensor  or  a 
dead  weight  tester  to  apply  pressure  in  desired 
increments to full scale.  Release pressure after taking 
the  reading  and  before  proceeding to the next higher 
pressure level. 

 

With  a  2000  second  discharge  time  constant,  1%  of 
the  signal  will  be  lost  in  20  seconds,  so  it  is 
imperative  that  the  pressure  setting  and  recording  of 
output be accomplished quickly. 

 

A  calibration  graph  can  be  plotted  using  output 
voltage vs. input pressure to determine sensitivity and 
linearity.

 

108 & 109 

SERIES 

002 SERIES 

      ICP

® 

POWER SUPPLY 

READOUT 

108 & 109 

SERIES 

ICP

® 

POWER SUPPLY 

 

002 SERIES 

READOUT 

ICP

®

 

POWER SUPPLY 

 

SERIES 002 

READOUT 

108 & 109 

SERIES 

Summary of Contents for M108A02

Page 1: ...10k psi 0 5 mV psi long life Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...f diaphragm give these sensors a high natural frequency and linearity The microelectronic amplifier converts the high impedance voltage from the quartz package into a low impedance high level output s...

Page 7: ...nicks and other imperfections which might cause leaks at high pressures 3 3 RECESSED MOUNT The recessed installation is shown in the Recessed Mount figure This type of mounting protects the sensor dia...

Page 8: ...e of recessed installation In severe pyrotechnic environments sensor life may be severely limited with flush installation 3 5 FLASH TEMPERATURE EFFECTS The ceramic coating on the diaphragm of these se...

Page 9: ...shows the 108 109 Series connected to a PCB Model 484B06 power unit The 484B06 is a power unit that can be AC or DC coupled In the DC couple mode a level shifting circuit removes the bias level and p...

Page 10: ...instrument several minutes to warm up and thermally stabilize When output from the power unit is connected to readout equipment a drift in the voltage signal will be noticed this is normal This volta...

Page 11: ......

Page 12: ......

Reviews: