Decimal Point and Brightness Selection
Digital Span Selection for Analog Range Output
STEP A Enter the Decimal Point and Brightness Mode Through the Sub
Menu [CAL]{oFF]
2) Press the
and the
buttons at the same time.
Display toggles between [cAL] and [oFF].
3) Press the
button. Display shows the previous [dp] selection.
STEP E Set the Decimal Point
1)
Using the
and
buttons, adjust the display to the desired decimal
point setting. (Factory Default is X.XXX)
2)
Press the
button. Display toggles between [Br] and the previous
[Br] setting.
STEP F Set the Display Brightness
1)
Using the
and
buttons, adjust the display to the desired
brightness setting (4 is the brightest setting).
2)
Press the
button. Display brightness changes to new setting and
display toggles between [Anhi] and the previous [Anhi] setting.
STEP G Setting the Digital Span Point for Analog High Output
1)
Using the
and
buttons, adjust the display to the desired digital
value which sets the point at which the selected analog high output
range will occur. (Factory Default is 1000)
2)
Press the
button. Display toggles between [AnLo] and the
previous [AnLo] setting.
STEP H Setting the Digital Span Point for Analog Low Output
1)
Using the
and
buttons, adjust the display to the desired digital
value which sets the point at which the selected analog low output
range will occur. (Factory Default is 0)
2)
Press the
button. The display exits the calibration mode and
returns to the operational display.
Note:
Any two digital scale points from
–
1999 to 9999 can be selected. The digital
scale points for analog high and analog low can be reversed for a 20mA to 4mA
output. The span of the digital scale can be as small as two counts however small
spans cause the 16 bit D to A to increment in stair case steps.