Drawing Number: 21167 
Revision: NR 

 

OPERATION MANUAL FOR

 

ICP

 WATER-COOLED PRESSURE SENSORS 

MODELS 124A21, A22, A24 

 

 2 

The Models 124A21, A22 and A24 sensors use 
low  output  impedance  probes  (ICP

)  and  will 

not  require  the  extreme  insulation  resistance 
precaution  commonly  associated  with  charge 
mode sensors. 
 
Connect  the  sensor  cable  to  the  appropriate 
ICP

  power  unit  and  check  the  front  panel 

monitor 

meter 

for 

continuity 

of 

cable 

connections and internal amplifier. 
Consult  Guide  G-0001B,  "General  Guide  to 
ICP

 Instrumentation" for installation and usage 

hints

 

for ICP

 instruments

.

 

 

4.0 

OPERATION

 

 
Consult  Guide  G-0001B  for  a  complete 
description  of  the  low-impedance  concept  in 
instrumentation. 
 
All PCB power units designed for ICP

 sensors 

use  contain  built-in  fault  monitor  meters  to  aid 
in locating circuit faults (shorts or opens) and to 
identify  normal  operation  of  the  miniature 
impedance  converting  amplifier  built  into  the 
sensor. 
 
Do  not  attempt  to  apply  voltage  directly  to  the 
input  pin  of  the  sensor  from  any  other  source.

  

The  input  current  must  be  limited  to  20  mA 
maximum  and  the  input  voltage  must  not  be 
higher  than  30  volts.    If  it  is  desired  to  build  a 
power  unit,  rather  than  to  use  one  of  those 
available  from  PCB,  contact  the  factory  for 
approval of this change. 
 
The  Models  124A21,  124A22  and  124A24  are 
identical except for sensitivity. 
 
The  Model  124A21  has  a  sensitivity  of  20 
mV/psi which gives a full-scale range (for 

5% 

out) of 250 psi. 

The  Model  124A22  has  a  sensitivity  of  1.0 
mV/psi and a full-scale range of 5000 psi. 
 
The  Model  124A24  has  a  sensitivity  of  5.0 
mV/psi and a full-scale range of 1000 psi. 
 

5.0 

CALIBRATION

 

 
Because  of  the  shorter  time  constants  built  into 
these  models,  static  calibration  methods  are 
difficult to employ. 
 
Pressures  must  be  applied  rapidly  and  readout 
must  be  made  with  fast  recording  devices  such 
as storage oscilloscopes or similar types of data 
acquisition hardware. 
 
A  pneumatic  (compressed  air)  system  with 
quick  opening  valve  and  reference  gage  is  an 
excellent method of calibrating these sensors. 
 
To take full advantage of the time constant built 
into  the  sensor,  use  a  DC-coupled  power  unit 
such as the PCB Model 484B. 
 

6.0 

MAINTENANCE  

 
The  sealed  construction  and  miniature  size  of 
these sensors make field repair impractical. 
 
Contact  factory  for  assistance  should  problems 
arise. 

 
 
 
 
 

 

ICP is a registered trademark of PCB 

Piezotronics 

Summary of Contents for ICP 124A21

Page 1: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Page 2: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Page 3: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Page 4: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Page 5: ...eramic coated tip for protection against erosion by hot gasses when the sensor is flush mounted directly in combustion chambers NOTE The sensor probe in this series is factory installed at a precise d...

Page 6: ...ut voltage must not be higher than 30 volts If it is desired to build a power unit rather than to use one of those available from PCB contact the factory for approval of this change The Models 124A21...

Page 7: ...l Diaphragm Invar Invar Sealing Welded Hermetic Welded Hermetic Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 4 0 oz 112 gm Water Flow Rate at 50 psi 1 2 gal min 4 5 L min All spec...

Page 8: ...WATER COOLED PRESSURE SENSOR XXX 005 JDM 12 28 18 JDM 12 28 18 MJP 12 28 18 69242 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction the...

Page 9: ...SURE SENSOR XXX 005 JDM 12 28 18 JDM 12 28 18 MJP 12 28 18 SEAL SURFACE MUST BE FLAT AND FREE OF TOOL MARKS WITH AT LEAST A FINISH OF 63 1 6 RECOMMENDED MOUNTING TORQUE 5 10 in lb RECOMMENDED MOUNTING...

Reviews: