PCB Piezotronics ICP 105C02 Installation And Operating Manual Download Page 8

 

Drawing Number: 21106 
Revision: NR 

 

 

OPERATION MANUAL FOR 

MINIATURE ICP

 PRESSURE SENSOR 

105A AND 105C SERIES 

 

 3 

The  natural  frequency  and  approximate  fastest 
pressure  step  rise  time  for  various  length  passages is 
shown in the following chart. (Medium, air at 25

C). 

 

PASSAGE 

LENGTH (in.) 

PASSAGE 

RESONANCE 

(kHz) 

APPROX. 
FASTEST 

PULSE RISE 

TIME (u Sec) 

.050 

66 

.100 

33 

10 

.200 

16.5 

20 

.50 

6.6 

50 

1.0 

3.3 

100 

 

Passage Resonance vs. Passage Length 

 
Actual resonant frequencies measured in practice may 
differ slightly from the chart values due to variations 
in  the  velocity  of  propagation  of  sound  caused  by 
changes in temperature and pressure of the medium in 
the passage. 
 
For best matching of passage to diaphragm, maintain 
the .010

 clearance ahead of the diaphragm as shown 

in Typical Recessed Installation figure.

 

 
4.0        OPERATION

 

 
Connect sensor to signal conditioner as shown below, 
depending upon sensor style being used. 
 
 

Model 480A

Signal Conditioner

To Readout

070A09 Solder
Connector Adaptor

105A

Solder Terminal

2-Wire Cable

Scope

XDCR

 

 
 

Typical Circuit Connection (For 105A) 

 
 
 

 

 
Typical  Circuit  Connection  (For  105C02,C03, 
C12, C13, C22, C23, C33, C43) 

 
NOTE:  All PCB signal conditioners for ICP

 sensor 

contain  current-limiting  circuitry  to  set  the  input 
current in the range of 2 to 20 mA. 
 
Do not connect a power supply to these units without 
current limiting.  To do so will destroy the built-in IC 
amplifier. 
 
All  PCB  signal  conditioners  feature  fault-monitoring 
voltmeters  on  the  front  panel.    During  normal 
operation,  the  meter  will  indicate  about  mid-scale 
(+11 volts).  If one of the connections to the sensor is 
open,  the  meter  will  read  full  scale  (yellow  area).  
Should a short occur on the sensor line, the meter will 
read 0 volts (red area). 
 

5.0 

CALIBRATION

 

 
Because  of  the  relatively  short  discharge  time 
constant of this series, static calibration means cannot 
be used.  Fast dynamic means such as step function or 
pulse generators must be used instead. 
 
Contact  the  factory  for  help  in  solving  specific 
calibration problems. 
 

6.0 

MAINTENANCE 

 
The  small  size  and  sealed  construction  of  these 
sensors  precludes  field  maintenance  and  repair.  
Should  a  problem  develop,  contact  the  factory  for 
assistance. 

 
 
 
 
 

002A Cable

XDCR Scope

To Readout

Model 480A
Signal

    Conditioner

XDCR

 

Summary of Contents for ICP 105C02

Page 1: ...Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...ndividual dimensional performance and installation drawings are included in the manual for each particular model Adhere carefully to instructions on these drawings It is not necessary to use low noise treated coaxial cable as with high impedance type sensors PCB Model 002A Cable is recommended 3 1 MODEL 105A This is the smallest of the three styles and it is supplied in only the 100 psi range The ...

Page 7: ...al Recessed Installation Because of its one piece machined diaphragm construction the 105A series is sensitive over its entire frontal area Use extreme care when preparing recessed installation port to ensure that no particles burrs or other residue are entrapped in the 010 gap ahead of the diaphragm Force on the diaphragm due to such entrapped particles may change the calibration of the sensor Re...

Page 8: ...pe XDCR Typical Circuit Connection For 105A Typical Circuit Connection For 105C02 C03 C12 C13 C22 C23 C33 C43 NOTE All PCB signal conditioners for ICP sensor contain current limiting circuitry to set the input current in the range of 2 to 20 mA Do not connect a power supply to these units without current limiting To do so will destroy the built in IC amplifier All PCB signal conditioners feature f...

Page 9: ...ls 4 Do not overtorque sensors when installing 5 Do not apply too much heat to the solder pin of style 1 8 0 TERMINAL LEAD AND GROUND To eliminate any possibility of overheating the solder terminal and destroying the microamplifier in the sensor a 2 0 inch long insulated wire has been provided on the solder terminal for soldering a connector lead from the power unit The ground return is provided b...

Page 10: ......

Page 11: ......

Page 12: ......

Reviews: