PCB Piezotronics ICP 105C02 Installation And Operating Manual Download Page 6

 

Drawing Number: 21106 
Revision: NR 

 

 

OPERATION MANUAL FOR 

MINIATURE ICP

 PRESSURE SENSOR 

105A AND 105C SERIES 

 

 1 

 
1.0 

INTRODUCTION

 

 
This  model  series  is  designed  to  satisfy  the  need  for 
measurement  of  dynamic  pressures  where  space  is 
limited, such as in model testing. 
 
The  series  consists  of  three  basic  mechanical 
configurations  or  styles,  each  featuring  a  .099

 

diameter  diaphragm  and  each  with  a  built-in 
impedance-converting 

amplifier 

for 

simple, 

convenient, two-wire, low-impedance operation. 
 

2.0 

DESCRIPTION

 

 
The Series 105A sensors utilize two quartz discs in a 
thickness  compression  mode  to  produce  a  high-
impedance level signal which is impressed across the 
input of a microminiature impedance 
converter built into the sensor housing. 
 
The signal is superimposed upon a +11 volt (nominal) 
bias  level  at  the  amplifier  output  and  is  normally 
decoupled from the bias voltage in the power unit. 
 
This  signal  can  be  fed  directly  into  most  common 
readout  instruments  such  as  oscilloscopes,  digital 
voltmeters,  AC  meters,  peak  meters,  recorders,  etc.  
The output impedance is less than 100 ohms. 
 
Consult  guide  G-0001B  "General  Guide  to  ICP

 

Instrumentation" for a more complete treatment of the 
ICP

 (Integrated Circuit Piezoelectric) principle. 

 

3.0 

INSTALLATION 

 
Because  of  the  miniature  size  of  these  units,  extra 
care  should  be  exercised  in  preparation  of  mounting 
ports and in installation of sensors into the ports. 
 
Individual  dimensional,  performance  and  installation 
drawings  are  included  in  the  manual  for  each 
particular model.  Adhere carefully to instructions on 
these drawings. 
 
It  is  not  necessary  to  use  low-noise  treated  coaxial 
cable  as  with  high-impedance  type  sensors.  PCB 
Model 002A Cable is recommended. 
 
 

 
 

 
3.1 

MODEL 105A  

 
This  is  the  smallest  of  the  three  styles  and  it  is 
supplied in only the 100 psi range.   
 
The  Model  105A  is  normally  mounted  using  Model 
060A04  Clamp  Nut  (supplied);  however,  if  lack  of 
space  precludes  the  use  of  the  clamp  nut,  the  sensor 
may be epoxied in place if necessary. 
 
Do not over torque clamp nut, as to do so may distort 
the body of the sensor and affect the sensitivity of the 
device. 
 
The  electrical  connection  to  this  model  is  made  by 
soldering  the  positive  (center  conductor)  lead  from 
the  power  unit  to  the  .040  diameter  solder  terminal 
provided. 
 
(IMPORTANT!    See Section 8.0, page 4) 
 
NOTE:    Do  not  overheat  this  terminal.    Use  a  small 
iron  with  only  enough  heat  to  obtain  good  solder 
joint. 
 
The  electrical  ground  return  is  made  to  the  sensor 
body.    This  connection  can  be  made  by  attaching 
electrically  to  the  structure  or  by  using  the  metal 
ribbon spotwelded to the sensor body. 
 

3.2 

MODELS 105C02, C12 & C22  

 
This style utilizes a 10-32 external thread to facilitate 
mounting.    It  is  supplied  in  three  full-scale  ranges: 
100, 1000 and 5000 psi. 
 
Prepare mounting port in accordance with installation 
drawing  provided  in  manual.  Do  not  over  torque.  
Use only enough torque to obtain a seal. 
 
The  electrical  connection  is  made  via  a  5-44  coaxial 
connector.  Use coaxial 002G10 cable, 5-44 to 10-32 
microdot. 
 
 

 
 
 
 
 

Summary of Contents for ICP 105C02

Page 1: ...Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...ndividual dimensional performance and installation drawings are included in the manual for each particular model Adhere carefully to instructions on these drawings It is not necessary to use low noise treated coaxial cable as with high impedance type sensors PCB Model 002A Cable is recommended 3 1 MODEL 105A This is the smallest of the three styles and it is supplied in only the 100 psi range The ...

Page 7: ...al Recessed Installation Because of its one piece machined diaphragm construction the 105A series is sensitive over its entire frontal area Use extreme care when preparing recessed installation port to ensure that no particles burrs or other residue are entrapped in the 010 gap ahead of the diaphragm Force on the diaphragm due to such entrapped particles may change the calibration of the sensor Re...

Page 8: ...pe XDCR Typical Circuit Connection For 105A Typical Circuit Connection For 105C02 C03 C12 C13 C22 C23 C33 C43 NOTE All PCB signal conditioners for ICP sensor contain current limiting circuitry to set the input current in the range of 2 to 20 mA Do not connect a power supply to these units without current limiting To do so will destroy the built in IC amplifier All PCB signal conditioners feature f...

Page 9: ...ls 4 Do not overtorque sensors when installing 5 Do not apply too much heat to the solder pin of style 1 8 0 TERMINAL LEAD AND GROUND To eliminate any possibility of overheating the solder terminal and destroying the microamplifier in the sensor a 2 0 inch long insulated wire has been provided on the solder terminal for soldering a connector lead from the power unit The ground return is provided b...