Certificate:
70089987
Project:
70089987
Master Contract:
184981
Date Issued:
2016-08-09
DQD 507 Rev. 2016-02-18
Page 2
Entity Parameters
Temperature Code
Ui / Vmax = 30V
Ii / Imax = 100mA
Pi / Pmax = 0.7W
Ci = 3300nF
Li = 30µH
T6 (–196
O
C to +80
O
C)
T5 (–196
O
C to +95
O
C)
T4 (–196
O
C to 130°C)
T3 (–196
O
C to 195°C)
T2 (–196
O
C to 290°C)
T1 (–196
O
C to 440°C)
T710 (–196
O
C to 700°C)
Notes:
1.
For Canadian Installations, sensor case must be bonded to ground according to Section 18-182 of the CEC,
Part 1.
2.
For US Installations, sensor case must be bonded to ground according to Article 501.16 of the NEC.
APPLICABLE REQUIREMENTS
CSA-C22.2 No. 0-10 (R2015) - General requirements – Canadian Electrical Code, Part II
C22.2 No. 61010-1-12
- Safety requirements for electrical equipment for measurement, control, and
laboratory use – Part 1: General requirements
ANSI/UL 61010-1 3
rd
Ed.
- Electrical Equipment for Measurement, Control, and Laboratory Use; Part 1:
General Requirements
C22.2 No. 60079-0: 2015
- Explosive atmospheres – Part 0: Equipment – General requirements
ANSI/UL 60079-0 6
th
Ed.
- Explosive atmospheres – Part 0: Equipment – General requirements
C22.2 No. 60079-11: 2014
- Explosive atmospheres – Part 11: Equipment protection by intrinsic safety “i”
ANSI/UL 60079-11 6
th
Ed.
- Explosive atmospheres – Part 11: Equipment protection by intrinsic safety “i”
65225 REV. NR, ECO 45417