background image

Model 480M122

Battery-Powered Signal Conditioner

Installation and Operating Manual

For assistance with the operation of this product,

contact PCB Piezotronics, Inc.

Toll-free: 800-828-8840

24-hour SensorLine: 716-684-0001

Fax: 716-684-0987

E-mail: [email protected]

Web: www.pcb.com

Summary of Contents for 480C02

Page 1: ...l Conditioner Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E...

Page 2: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Page 3: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Page 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Page 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Page 6: ...ICP Signal Conditioners Operating Manual with Enclosed Warranty Information 3425 Walden Avenue Depew New York 14043 2495 Phone 716 684 0003 Fax 716 684 3823 Toll Free Line 1 800 828 8840 MANUAL NUMBE...

Page 7: ...tents Introduction page 3 Description page 4 Operation page 5 Performance Limits page 8 Battery Considerations page 8 Optional Power Connections page 9 Warranty page 10 Calibration Service page 10 Ret...

Page 8: ...They are powered internally by three replaceable batteries and provide constant current excitation to the charge amplifier within ICP sensors or to in line and adaptor style amplifiers and decouples...

Page 9: ...el meter indicates the voltage output when connected to a sensor and is color coded for various fault conditions shown in Figure 1 2 Optional connections include sockets for EXT PWR and BATT CHG A cen...

Page 10: ...age with full scale set at 27 volts 27 volts for new batteries see Figure 1 When an ICP sensor is connector to the input XDCR jack the meter will indicate approximately mid scale 11 volts nominal if t...

Page 11: ...pacitor is fully charged Such drifting is considered normal operation Coupling Time Constant TC AC Coupled The coupling discharge time constant DTC is the product of the capacitor 47 F and the paralle...

Page 12: ...e 5 the low frequency response of the coupling circuit is determined by the relationship in Equation 2 This requires use of a BNC Tee Figure 6 below provides a comparison of AC vs DC coupling the same...

Page 13: ...driven but it sacrifices some high frequency response PCB line powered signal conditioners 482 Series 483 Series should be considered for cable lengths beyond 100 feet 31 meters as they provide up to...

Page 14: ...Charging WARNING models 480C02 480E09 do not ship with rechargeable batteries and PCB external battery chargers are suitable for use with Nickel Cadmium Ni Cad rechargeable batteries only It should no...

Page 15: ...lt a factory PCB applications engineer concerning the situation as certain problems can often be corrected with simple on site procedures PCB will perform calibrations on model 480C02 on request Typic...

Page 16: ...funded PCB offers to all customers at no charge 24 hour phone support This service makes product or application support available to our customers day or night seven days a week When unforeseen proble...

Page 17: ...Internal Battery Internal Battery Internal Battery Type 9V 9V Battery Life Standard Alkaline 50 hours 50 hours Power Required Alternate DC power DC power DC Power 15 mA 15 mA 5 Internal Battery Quanti...

Page 18: ......

Reviews: