background image

 

Model Number

 

377B20

 

PRECISION CONDENSER MICROPHONE

 

Revision: K 

ECN #: 42896

 

 

Performance 

ENGLISH 

SI 

Nominal Microphone Diameter

 

1/2"

 

1/2"

 

 

Frequency Response Characteristic

 

Random Incidence

 

Random Incidence

 

 

Open Circuit Sensitivity

 

50 mV/Pa

 

50 mV/Pa

 

[3]

 

Open Circuit Sensitivity

(± 1.5 dB)

 

-26 dB re 1 V/Pa

 

-26 dB re 1 V/Pa

 

[3]

 

Frequency Range(± 1 dB)

 

5 to 6300 Hz

 

5 to 6300 Hz

 

 

Frequency Range

(± 2 dB)

 

3.15 to 12,500 Hz

 

3.15 to 12,500 Hz

 

 

Lower Limiting Frequency(-3 dB)

 

1.0 to 2.4 Hz

 

1.0 to 2.4 Hz

 

 

Resonant Frequency(90° Phase Shift)

 

12,000 Hz

 

12,000 Hz

 

[1]

 

Inherent Noise

 

<15 dB(A) re 20 

µPa

 

<15 dB(A) re 20 

µPa

 

 

Dynamic Range(3% Distortion Limit)

 

>144 dB re 20 µPa

 

>144 dB re 20 µPa

 

 

Standards Designation(IEC 651)

 

Type 1

 

Type 1

 

[4]

 

Environmental 

Temperature Range(Operating)

 

-40 to +248 °F

 

-40 to +120 °C

 

 

Temperature Coefficient of Sensi14 to +122°F (-10 to +50°C))

0.005 dB/°F

 

0.009 dB/°C

 

[1][3]

Static Pressure Coefficient

 

-0.01 dB/kPa

 

-0.01 dB/kPa

 

[1][3]

Humidity Coefficient of Sensitivity(0 to 100%, non-condensing)

 

± 0.001 dB/%RH

 

± 0.001 dB/%RH

 

[3]

 

Influence of Axial Vibration(0.1g (1 m/s²))

 

63 dB re 20 µPa

 

63 dB re 20 µPa

 

[1]

 

Electrical 

Capacitance(Polarized)

 

11 pF

 

11 pF

 

[1]

 

Polarization Voltage

 

0 V

 

0 V

 

[2]

 

Physical 

Housing Material

 

Stainless Alloy

 

Stainless Alloy

 

 

Venting

 

Rear

 

Rear

 

 

Mounting Thread(Preamplifier)

 

0.4606 - 60 UNS

 

11.7 mm - 60 UNS

 

 

Mounting Thread

(Grid)

 

0.5 - 60 UNS

 

12.7 mm - 60 UNS

 

 

Size (Diameter x Height)(with grid)

 

0.52 in x 0.64 in

 

13.2 mm x 16.2 

mm

 

 

Size (Diameter x Height)

(without grid)

 

0.5 in x 0.6 in

 

12.7 mm x 15.3 

mm

 

 

Weight

 

0.28 oz

 

7.8 gm

 

[1]

 

All specifications are at room temperature unless otherwise specified. 
In the interest of constant product improvement, we reserve the right to change specifications without notice. 

ICP

®

 is a registered trademark of PCB Group, Inc. 

 

OPTIONAL VERSIONS 

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

 

 
 
 
 

 

 

NOTES:

[1] Typical.
[2] Prepolarized
[3] re 250 Hz
[4] The microphone uses 50% to 70% of the frequency tolerances allowed by this standard.

 

 

 

SUPPLIED ACCESSORIES:  

Model ACS-20 Calibration of Precision Condensor Microphones (1) 

Entered: AP

 

Engineer: MJN

 

Sales: MV

 

Approved: MT

 

Spec Number:

 

Date: 5/7/2014

 

Date: 5/7/2014

 

Date: 5/7/2014

 

Date: 5/7/2014

 

31214 

 

3425 Walden Avenue, Depew, NY 14043

Phone: 716-684-0001 
Fax: 716-684-0987 
E-Mail: [email protected]

Summary of Contents for 378B20

Page 1: ...e 377B20 426E01 installation and operating manuals that comprise a Model 378B20 ICP Microphone System kit For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 8...

Page 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Page 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Page 4: ...he instrument will be used Whether for legal purposes or for quality assurance programs these standards will help determine the required quality accuracy and consistency of the instrument Standards th...

Page 5: ...ponse reactions of the human ear to the pressure fluctuations Some examples of typical sound pressure levels are shown in the table below dB Pressure Example 0 0 00002 Pa Threshold of Hearing 60 0 02...

Page 6: ...and noise cones the fre quency response of the microphone will be attenuated at higher frequencies Clamps and stands When holding a microphone in place it is important to minimize the influence of the...

Page 7: ...2LA manufacturers Our internal metrology laboratory is certified to ISO 9001 accredited by A2LA to AN SI IEC 17025 and ANSI NCSL Z540 1 complies with ISO 10012 1 and former MIL STD 45662A and uses equ...

Page 8: ...VDC Output Impedance 50 Ohm 50 Ohm Maximum Output Voltage 7 Vpk 7 Vpk Physical Housing Material Stainless Alloy Stainless Alloy Venting Rear Rear 6 Electrical Connector BNC Jack BNC Jack Mounting Thre...

Page 9: ......

Page 10: ...Microphone Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E m...

Page 11: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Page 12: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Page 13: ...he instrument will be used Whether for legal purposes or for quality assurance programs these standards will help determine the required quality accuracy and consistency of the instrument Standards th...

Page 14: ...ponse reactions of the human ear to the pressure fluctuations Some examples of typical sound pressure levels are shown in the table below dB Pressure Example 0 0 00002 Pa Threshold of Hearing 60 0 02...

Page 15: ...and noise cones the fre quency response of the microphone will be attenuated at higher frequencies Clamps and stands When holding a microphone in place it is important to minimize the influence of the...

Page 16: ...2LA manufacturers Our internal metrology laboratory is certified to ISO 9001 accredited by A2LA to AN SI IEC 17025 and ANSI NCSL Z540 1 complies with ISO 10012 1 and former MIL STD 45662A and uses equ...

Page 17: ...20 Pa 1 Electrical Capacitance Polarized 11 pF 11 pF 1 Polarization Voltage 0 V 0 V 2 Physical Housing Material Stainless Alloy Stainless Alloy Venting Rear Rear Mounting Thread Preamplifier 0 4606 60...

Page 18: ...ARIZED MICROPHONES DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 JDM 9 14 05 ECB 9 14 05 CCL 9 14 05 1 1 2 2 A A B B 60 15 3 64 16 4 36 9 2 50 12 7 4606 60 UNS 11 7 mm 60 UNS THREADS 52 13 2 47 12 0 4...

Page 19: ...plifier Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail...

Page 20: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Page 21: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Page 22: ...he instrument will be used Whether for legal purposes or for quality assurance programs these standards will help determine the required quality accuracy and consistency of the instrument Standards th...

Page 23: ...ponse reactions of the human ear to the pressure fluctuations Some examples of typical sound pressure levels are shown in the table below dB Pressure Example 0 0 00002 Pa Threshold of Hearing 60 0 02...

Page 24: ...and noise cones the fre quency response of the microphone will be attenuated at higher frequencies Clamps and stands When holding a microphone in place it is important to minimize the influence of the...

Page 25: ...2LA manufacturers Our internal metrology laboratory is certified to ISO 9001 accredited by A2LA to AN SI IEC 17025 and ANSI NCSL Z540 1 complies with ISO 10012 1 and former MIL STD 45662A and uses equ...

Page 26: ......

Page 27: ......

Reviews: