background image

 

 

 

 
Manual 21354 Rev E 
ECN 50523 

 

 

Component Name 

Hazardous Substances 

Lead (Pb)  Mercury (Hg) 

Cadmium (Cd) 

Chromium VI 
Compounds 
(Cr(VI)) 

Polybrominated 
Biphenyls (PBB) 

Polybrominated 
Diphenyl Ethers 
(PBDE) 

Housing 

PCB Board 

Electrical Connectors 

Piezoelectric Crystals 

Epoxy 

Teflon 

Electronics 

Thick Film Substrate 

Wires 

Cables 

Plastic 

Solder 

Copper Alloy/Brass 

This table is prepared in accordance with the provisions of SJ/T 11364. 
O: Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit 
requirement of GB/T 26572. 

X: Indicates that said hazardous substance contained in at least one of the homogeneous materials for this part is above the limit 
requirement of GB/T 26572. 
Lead is present due to allowed exemption in Annex III or Annex IV of the European RoHS Directive 2011/65/EU. 

Summary of Contents for 357B11

Page 1: ...r charge output accel 3 pC g Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...F 482 C and extreme application up to 100 F 650 C for UHT 12 These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high tem...

Page 7: ...urface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the moun...

Page 8: ...or most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are que...

Page 9: ...o a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitiv...

Page 10: ...dure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes asso...

Page 11: ...Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the tes...

Page 12: ...s Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz...

Page 13: ...e specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERSIONS Optional versions have identical specifications and accessories as listed for the standard model except...

Page 14: ...ONS METRIC DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTING TORQUE 2 5 FOOT POUNDS 3 7 NEWTON METERS MOUNTING HOLE PREPARATION 218 5 54 X 300 7 62 MIN 1 4 28 UNF 2B X 200 5 08 MIN F M6 X 0 75 M6 X 1 00 M...

Page 15: ......

Reviews: