background image

MANUAL: 55308 
REV:  A 
ECO:  49582 

For optimum performance and measurement find a 
rigid  location  on  the  machine  casing  that  most 
accurately  represents  the  vibration  of  the  rotor, 
bearing, fan, etc. to be measured.).   
 
 

3.4  Mounting Sensor 

When  choosing  a  mounting  method,  consider  closely 
both the advantages and disadvantages of each technique.  
Characteristics  like  location,  ruggedness,  amplitude 
range,  accessibility,  temperature,  and  portability  are 
extremely  critical.    However,  the  most  important  and 
often overlooked consideration is the effect the mounting 
technique has on the high-frequency performance of the 
accelerometer. 
 
Shown  hereafter  are  six  possible  mounting  techniques 
and  their  effects  on  the  performance  of  a  typical 
piezoelectric  accelerometer.    (Note  that  not  all  of  the 
mounting methods may apply to your particular sensor.) 
The  mounting  configurations  and  corresponding  graph 
demonstrate  how  the  high-frequency  response  of  the 
accelerometer may be compromised as mass is added to 
the system and/or the mounting stiffness is reduced. 

 
Note:

  

The low-frequency response is unaffected by the 

mounting technique.  Its roll-off behavior is dependent 

 

upon the coupling type or time constant of the external 
charge  amplifying  device.    This  characteristic  may  be 
fixed or adjustable, depending on the  charge amplifier.  
Please  refer  to  the  appropriate  charge  amplifier 
specification  sheet  for  information  on  low-frequency 
characteristics. 
 

 

 

Figure 1.

  Assorted Mounting Configurations and Their 

Effects on High Frequency 
 

3.4.1 STUD MOUNT 

This  mounting  technique  requires  smooth,  flat  contact 
surfaces  for  proper  operation  and  is  recommended  for 
permanent and/or secure installations.  Stud mounting is 
also recommended when testing at high frequencies. 
 

Note:

  

Do NOT attempt mounting on curved, rough, or 

uneven  surfaces,  as  the  potential  for  misalignment  and 
limited  contact  surface  may  significantly  reduce  the 
sensor’s upper operating frequency range. 
 

 

 
Figure 2.

 Mounting Surface Preparation 

 
STEP 1:  

First, prepare a smooth, flat mounting surface, 

then  drill  and  tap  a  mounting  hole  in  the  center  of  this 
area  as  shown  in  Figure  2  and  in  accordance  with  the 

Installation Drawing 

for the specific sensor that is being 

mounted

 
A precision-machined mounting surface with a minimum 
finish of 63 

in (0.00016 mm) is recommended.  (If it is 

not  possible  to  properly  prepare  the  test  structure 
mounting  surface,  consider  adhesive  mounting  as  a 
possible alternative.) Inspect the area, checking that there 
are no burrs or other foreign particles interfering with the 
contact surface. 
 

STEP 2:  

Wipe clean the mounting surface and spread on 

a light film of grease, oil, or similar coupling fluid prior 
to installation. 
 

 

 

Figure 3.

 Mounting Surface Lubrication 

 
Adding 

coupling 

fluid 

improves 

vibration 

transmissibility  by  filling  small  voids  in  the  mounting 
surface and increasing the mounting stiffness.  For semi-

Summary of Contents for 357a100

Page 1: ...erometer Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...with PIEZOELECTRIC CHARGE MODE ACCELEROMETERS SPECIFICATION SHEET INSTALLATION DRAWING AND CALIBRATION INFORMATION ENCLOSED PCB ASSUMES NO RESPONSIBILITY FOR DAMAGE CAUSED TO THIS PRODUCT AS A RESULT...

Page 7: ...nector 4 2 5 In Line Differential Charge Amplifier 4 3 INSTALLATION OVERVIEW 4 3 1 Equipment Inspection 4 3 2 Polarity Test 4 3 3 Sensor Location 4 3 4 Mounting Sensor 5 3 5 Route Mineral Insulated Ha...

Page 8: ...re high temperatures preclude the use of accelerometers with built in microelectronics Enclosed is a Specification Sheet which lists the complete performance characteristics of the particular accelero...

Page 9: ...harge Amplifier The differential charge amplifier is purchased separately A conventional method for conditioning the high impedance signal generated by a charge output sensor is to use a differential...

Page 10: ...priate charge amplifier specification sheet for information on low frequency characteristics Figure 1 Assorted Mounting Configurations and Their Effects on High Frequency 3 4 1 STUD MOUNT This mountin...

Page 11: ...est object STEP 1 Prepare a smooth flat mounting surface A minimum surface finish of 63 in 0 00016 mm generally works best STEP 2 Stud mount the sensor to the appropriate adhesive mounting base accord...

Page 12: ...amp Cable To minimize triboelectric motion induced noise from the cable interfering with the sensors high impedance charge output cable clamps must be used Clamp the cable as close to the transducer a...

Page 13: ...ation Once each element is connected allow a few minutes for the system to thermally stabilize Place the switch on the charge amplifier in the OPERATE position and proceed with the measurement It is o...

Page 14: ...vices As demonstrated in Equation 1 the steady state DC output voltage Eo is Equation 1 i f i o R R e E 1 This equation shows that if the input leakage resistance at the charge amplifier is extremely...

Page 15: ...mentioned and one comprised of series capacitor Cs and total equivalent shunt resistance Ri This new cutoff frequency is Equation 3 s i o C R 16 f To avoid compromise of the low frequency response est...

Page 16: ...s checking against other equipment manufacturer recommendation and any risk associated with incorrect readings International standards such as ISO 10012 1 provide insight and suggested methods for det...

Page 17: ...ay be automated and vary depending on the type and number of accelerometers being calibrated Figure 2 Typical Back to Back Calibration System Because the acceleration is the same on both sensors the r...

Page 18: ...4 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the test sensor...

Page 19: ...rd Validate your calibration system with another accelerometer prior to each calibration session Check with the manufacturer for exact system specifications 7 5 CONCLUSIONS Without an adequate underst...

Page 20: ...5 in x 1 88 in 19 mm x 38 1 mm x 47 8 mm Weight 2 32 oz 65 8 gm 1 Electrical Connector 7 16 27 2 Pin 7 16 27 2 Pin Electrical Connection Position Side Side Mounting Through Holes 3 Through Holes 3 5 A...

Page 21: ...NG 63685 1 OF 1 2X ACCELEROMETER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 JDM 2 17 16 ECB 2 17 16 GCD 2 17 16 63685 PCB Piezotronics Inc claims proprietary rights in the information disclosed her...

Reviews: