background image

Model Number

1411-02ADB

FATIGUE RATED LOW PROFILE LOAD CELL, DUAL BRIDGE

Revision: D
ECN #: 43840

Performance

ENGLISH

SI 

Measurement Range(100 x 10^6 cycles)

100 klb

445 kN

[5][6]

Sensitivity(± 10 %)(Rated Output)

2.0 mV/V

2.0 mV/V

[7][1]

Non-Linearity

≤ 0.06 % FS

≤ 0.06 % FS

Hysteresis

≤ 0.06 % FS

≤ 0.06 % FS

Non-Repeatability

≤ 0.02 % FS

≤ 0.02 % FS

Eccentric Load Sensitivity

≤ 0.1 %/in

≤ 0.1 %/25.4mm

Resonant Frequency

18 kHz

18 kHz

Side Load Sensitivity

≤ 0.1 %/in

≤ 0.1 %/25.4mm

Creep(in 20 minutes)

0.025 %

0.025 %

Static Error Band

≤ 0.06 % FS

≤ 0.06 % FS

[5]

Environmental

Overload Limit

300 klb

1335 kN

Load Limit(Side Force, F

X

 or F

Y

)

100 klb

445 kN

[3]

Load Limit

(Bending Moment, M

X

 or M

Y

)

100k lbf-in

11.3k Nm

[3]

Load Limit

(Axial Torque, M

Z

)

100k lbf-in

11.3k Nm

[3]

Temperature Range(Operating)

-65 to +200 °F

-54 to +93 °C

Temperature Range

(Compensated)

+15 to +115 °F

-9 to +46 °C

Temperature Effect on Output(Maximum)

± 0.0008 %Reading/°F

± 0.0018 %Reading/°C

[4]

Temperature Effect on Zero Balance(Maximum)

± 0.0008 %FS/°F

± 0.0018 %FS/°C

[4]

Electrical

Bridge Resistance

350 Ohm

350 Ohm

[1]

Excitation Voltage(Recommended)

10 VDC

10 VDC

[2]

Insulation Resistance

≥ 5k GOhm

≥ 5k GOhm

Zero Balance

≤ 1 % FS

≤ 1 % FS

Output Polarity

+Tension

+Tension

Physical

Size (Diameter x Height)

11.00 in x 7.00 in

279.40 mm x 177.80 mm

[8]

Weight

135 lb

61.3 kg

Mounting Thread

2 3/4 - 8 UN-3B

No Metric Equivalent

Sensing Element

Strain Gage

Strain Gage

Mounting Torque

6000 lbf-in

678 Nm

Deflection at Full Scale Capacity

0.001 in

0.03 mm

Electrical Connector

PT02E-10-6P (2)

PT02E-10-6P (2)

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

NOTES:

[1]Nominal.
[2]Calibrated at 10 VDC, usable 5 to 20 VDC or VAC RMS.
[3]Singularly applied, i.e. no other extraneous loads.
[4]Over compensated operating temperature range.
[5]FS - Full Scale.
[6]Calibrated in tension and compression.
[7]RO - Rated Output.
[8]See Outline Drawing 51434 for Complete Dimensions

OPTIONAL ACCESSORIES: 

Model 084A40 Lifting Plug
Model 084A41 Lifting Plug
Model 084A58 Pre-Tension Stud
Model 084A59 Pre-Tension Stud
Model 181-012A PT06A-10-6S(SR)
Model 8311-01-10A CABLE

PCB Load & Torque, Inc.
24350 Indoplex Circle
Farmington Hills, MI 48335
UNITED STATES
Phone: 866-684-7107
Fax: 716-684-0987
E-Mail: [email protected]
Web site: 
http://www.pcbloadtorque.com

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard 

model except where noted below. More than one option may be used.

Entered: AP

Engineer: PE

Sales: RWM

Approved: JSD

Spec Number:

Date: 2/17/2015

Date: 2/17/2015

Date: 2/17/2015

Date: 2/17/2015

51822 

Summary of Contents for 1411-02ADB

Page 1: ...DUAL BRIDGE Installation and Operating Manual For assistance with the operation of this product contact PCB Load Torque Inc Toll free 866 684 7107 24 hour SensorLine 716 684 0001 Fax 248 888 8266 E ma...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...LECTRICAL INSTALLATION 5 5 1 Electrical Drawing Western Regional Std 5 5 2 Cable Grounding Considerations 5 6 0 CALIBRATION 5 6 1 Calibration Certificate Description 5 6 1 1 Measured Output 6 6 1 2 Hy...

Page 7: ...the parts under test have completed the planned fatigue test cycles 3 0 OVERVIEW Dual bridge fatigue rated load cells are specifically designed for repetitive testing Applications include material te...

Page 8: ...ote that the performance of the load cell may be affected if the factory installed mounting base is removed or modified 4 2 Mounting Load Cell to a Standard Base or Custom Fixture If the load cell is...

Page 9: ...ctor Protector 1 Remove the 4 40 x 1 4 button head cap screws from the connector using a 1 16 hex wrench as shown in Figure 7 2 Carefully pull the connector from the load cell Take care to not break o...

Page 10: ...wiring may result in unstable or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structur...

Page 11: ...s are adjusted such that the error due to eccentric loading is less than 0 1 per inch of offset Side loads moments and torque must be limited to avoid electrical or mechanical damage to the load cell...

Page 12: ...tion certificate and adjust the span or gain until the display reads the force value stated on the certificate 4 Repeat steps 1 3 to verify that a valid calibration setting has been obtained 5 If poss...

Page 13: ...ridge resistances A D excitation and B C the signal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB factory for assistance 11 2 1 Estimating Bridge B...

Page 14: ...factory for any reason For calibration services if possible a copy of the purchase order covering the requested services should be included with the returned load cell 13 0 WARRANTY Standard warranty...

Page 15: ...80 mm 8 Weight 135 lb 61 3 kg Mounting Thread 2 3 4 8 UN 3B No Metric Equivalent Sensing Element Strain Gage Strain Gage Mounting Torque 6000 lbf in 678 Nm Deflection at Full Scale Capacity 0 001 in 0...

Page 16: ...ECTOR WIRING CODE TYPICAL PIN A EXC PIN B SIG PIN C SIG PIN D EXC PIN E N C PIN F N C CONNECTOR B DUAL BRIDGE ONLY PROTECTIVE SHROUD THREAD x2 56 65 0 STANDARD LOAD CELL 2 3 4 8 UN 3B METRIC M LOAD CE...

Reviews: