PCB Piezotronics 132A35 Installation And Operating Manual Download Page 7

 

Model Number

 

132A35

 

ICP® PRESSURE SENSOR

 

Revision: L 

ECN #: 41151

 

 

Performance 

ENGLISH 

SI 

Measurement Range

 

48 psi

 

330 kPa

 

 

Sensitivity(± 30 %)

 

140 mV/psi

 

20.3 mV/kPa

 

 

Maximum Pressure(Dynamic)

 

800 psi

 

5515 kPa

 

 

Resolution

 

1 mpsi

 

0.007 kPa

 

 

Resonant Frequency

 

 1000 kHz

 

 1000 kHz

 

 

Rise Time(Incident)

 

 3 µ sec

 

 3 µ sec

 

[1]

 

Rise Time

(Reflected)

 

 0.5 µ sec

 

 0.5 µ sec

 

 

Low Frequency Response(-5 %)

 

11 kHz

 

11 kHz

 

 

Environmental 

Temperature Range(Operating)

 

0 to +175 °F

 

-18 to +79 °C

 

 

Electrical 

Discharge Time Constant(at room temp)

 

 .000045 sec

 

 .000045 sec

 

 

Excitation Voltage

 

20 to 30 VDC

 

20 to 30 VDC

 

 

Constant Current Excitation

 

2 to 20 mA

 

2 to 20 mA

 

 

Output Impedance

 

 100 Ohm

 

 100 Ohm

 

 

Output Bias Voltage

 

8 to 14 VDC

 

8 to 14 VDC

 

 

Physical 

Sensing Element

 

Ceramic

 

Ceramic

 

 

Housing Material

 

Stainless Steel

 

Stainless Steel

 

 

Sealing

 

Epoxy

 

Epoxy

 

 

Electrical Connector

 

pigtail ends

 

pigtail ends

 

 

Weight(with cable)

 

0.018 oz

 

0.52 gm

 

 

All specifications are at room temperature unless otherwise specified. 
In the interest of constant product improvement, we reserve the right to change specifications without notice. 

ICP

®

 is a registered trademark of PCB Group, Inc. 

 

OPTIONAL VERSIONS 

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

 

 

 
M 

- Metric Mount

 

 
W 

- Water Resistant Cable

 

 

 

NOTES:

[1] Rise time in air at Mach 1

 

 

 

SUPPLIED ACCESSORIES:  

Model 060A28 Clamp Nut, 1/4-28 x 1/4" Hex (1) 
Model 070B09 Solder adaptor (1) 

Entered: AP

 

Engineer: MJK

 

Sales: RWM

 

Approved: APB

 

Spec Number:

 

Date: 3/27/2013

 

Date: 3/27/2013

 

Date: 3/27/2013

 

Date: 3/27/2013

 

132-1350-80 

 

3425 Walden Avenue, Depew, NY 14043

Phone: 716-684-0001 
Fax: 716-684-0987 
E-Mail: [email protected]

Summary of Contents for 132A35

Page 1: ...mp nut 1 4 28 mtg thd for high Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 F...

Page 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Page 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Page 4: ...rs is not attenuated nor is noise appreciably increased by the cable 2 0 DESCRIPTION 132 Series Microsensors feature small diameter pressure sensing surfaces incorporating very small 1 mm piezoelectri...

Page 5: ...n of indicator panel the IC amplifier is providing proper bias normally 11 VDC cable connections are normal and the system is ready to operate If the needle moves to the red area of the fault monitor...

Page 6: ...ook flat topped on the readout after about 10 volts The initial rise of the output will be accurate up to the point of clipping This can allow for an accurate time of arrival measurement even though t...

Page 7: ...terial Stainless Steel Stainless Steel Sealing Epoxy Epoxy Electrical Connector pigtail ends pigtail ends Weight with cable 0 018 oz 0 52 gm All specifications are at room temperature unless otherwise...

Page 8: ......

Reviews: