OPERATION MANUAL FOR MICROSENSORS 

132

 

Series 

 

 3 

4.1 

OPERATION BEYOND RANGE 
SPECIFICATION
 

 
Because  132  Series  Microsensors  can  be  installed 
using  RTV-like  potting  materials  to  hold  them  in 
place,  maximum  pressure  specifications  are  given  to 
prevent  the  sensors  from  being  extruded  from  the 
mounting 

holes. 

 

Static 

or 

slow 

dynamic 

overpressures are more apt to cause a problem than a 
short dynamic shock pulse.  Use of backing nuts will 
minimize the possibility of a problem. 
 
The  microsensors  are  capable  of  responding  to 
dynamic  overpressures  up  to  ten  times  the  specified 
range.  The output will be “clipped” (look flat-topped 
on the readout) after about 10 volts. 
 
The initial rise of the output will be accurate up to the 
point  of  clipping.    This  can  allow  for  an  accurate 
time-of-arrival  measurement,  even  though  the  peak 
pressure cannot be measured. 
 

 
5.0 

CALIBRATION 

 
A shock tube is used to calibrate the microsensors. 
 

 
6.0 

MAINTENANCE AND REPAIR 

 
Except  for  repair  of  wires,  the  microsensors  are  not 
field-repairable.    In  case  of  serious  malfunction, 
contact the PCB at (888) 684-0011. 
 
It  is  well  to  observe  the  following  precautions  in 
using the microsensors: 
 
1.

 

Do  not  exceed  specified  pressure  levels  without 
proper mounting. 

 
2.

 

Do  not  subject  microsensors  to  temperatures 
exceeding 174 

°

F (79 °C). 

 
3.

 

Do not apply voltage to sensors without current-
limiting diodes or other current protection. (PCB 
ICP

  power  supplies  feature  proper  voltage  and 

current protection.) 

 
4.

 

Do not apply more than 20 mA of current to the 
microsensors. 

 

Summary of Contents for 132A35

Page 1: ...mp nut 1 4 28 mtg thd for high Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 F...

Page 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Page 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Page 4: ...rs is not attenuated nor is noise appreciably increased by the cable 2 0 DESCRIPTION 132 Series Microsensors feature small diameter pressure sensing surfaces incorporating very small 1 mm piezoelectri...

Page 5: ...n of indicator panel the IC amplifier is providing proper bias normally 11 VDC cable connections are normal and the system is ready to operate If the needle moves to the red area of the fault monitor...

Page 6: ...ook flat topped on the readout after about 10 volts The initial rise of the output will be accurate up to the point of clipping This can allow for an accurate time of arrival measurement even though t...

Page 7: ...terial Stainless Steel Stainless Steel Sealing Epoxy Epoxy Electrical Connector pigtail ends pigtail ends Weight with cable 0 018 oz 0 52 gm All specifications are at room temperature unless otherwise...

Page 8: ......

Reviews: