DUAL BRIDGE LOAD CELL OPERATION MANUAL 

 

3

 

3.2 Standard Components 

The following figure describes the standard components of the 
dual bridge fatigue-rated load cells. 
 
 
 
 
 
 
 
 
 
 
 
 
 

3.3 Optional Components 

The following figure describes the optional components of the 
dual bridge fatigue-rated load cells. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
NOTE:  Connector  protectors  are  standard  on  8”  and  11” 

models. 

 
Table 3 describes the available optional components 
 
 
 
 
 
 
 
 
 
 
 
 
 
NOTE:  Pre-tension  studs  include  the  threaded  stud  and  the 

jam nut. 

4.0 MECHANICAL INSTALLATION 

 
Standard  components  of  the  dual  bridge  fatigue-rated  load 
cells  have  all  been  installed  to  factory  specified  values  for 
optimum performance.  The following mechanical installation 
instructions should be observed if optional components are to 
be used. 
 

4.1 Mounting Bases 

All low-profile load cells are calibrated and supplied with a 
factory  installed  standard  mounting  base.  An  optional 
custom (customer supplied) mounting base can, with special 
arrangements,  be  installed  and  used  for  calibration. 
However,  there  will  be  an  extra  charge  for  removal  of  the 
standard  base,  and  installation  of  the  load  cell  on  the 
customer’s base / fixture.  
 

NOTE:  To insure catalog performance specifications are met 

the customer’s supplied fixture or base must be flat to 
within  0.0002  inch  and  provide  similar  stiffness  and 
hardness to a factory base. 

 
Unless otherwise specified/requested, the base and load cell 
are  calibrated  as  a  unit.  It  is  important  to  note  that  the 
performance of the load cell may be affected if the  factory-
installed mounting base is removed or modified. 

 

4.2 Mounting Load Cell to a Standard Base or 
Custom Fixture 

If the load cell is to be mounted to a custom base or fixture, it 
is necessary to consider the following: 
 

1.

 

The thickness of the new base or fixture must not be 
any  thinner  than  the  original  base  supplied  with  the 
load  cell.  Materials  used  must  have  the  same 
temperature  coefficient  of  expansion  (for  example, 
4140  Steel  or  2024-T4  Aluminum  for  250,500  and 
1,000 capacity aluminum cells). 

 
2.

 

The  mounting  surface  must  be  flat  (within  0.0002 
inches),  and  steel  fixtures  must  have  a  hardness  of 
Rockwell C 33 to 37.  

 
3.

 

The  threaded  holes  and  under-head  of  the  fasteners 
must  be  lightly  lubricated  with  30  weight  oil  or 
equivalent prior to tightening the mounting bolts. 

 

4.

 

Refer  to  Table  4  for  the  final  tightening  torque 
required for each bolt size. 

 

5.

 

Tighten the bolts in an incremental cross pattern with 
the  following  sequence:  start  with  all  bolts  finger 
tight followed by 25%, 50% and finally 100% of the 
required torque being applied. 

 
 

 

Load Cell 

Tension 

Base 

Connector A 

Connector B 

 

Connector 

Protector 

Pre-Tension 

Stud 

Pre-Tension 

Stud 

Jam Nut 

Jam Nut 

Figure 2

 

– Standard Components 

Figure 3

 

– Optional Components 

Item Number 

Item Description 

084A40 

Lifting Rings (side), 1411 Series 

084A41 

Lifting Plug (top), 1411 Series 

084A55 

Pre-Tension Stud (5/8”-18 Thread), 1403 Series 

084A56 

Pre-Tension Stud (1 ¼”-12 Thread), 1404 Series 

084A57 

Pre-Tension Stud (1 ¾”-12 Thread), 1408 Series 

084A58 

Pre-Tension Stud (2 ¾”-8 Thread), 1411 Series  

084A59 

Pre-Tension Stud (2 ¾”-8 to 1 ¾”-12 Thread) 1411 Series 

084A90 

Connector Protector, 1403 and 1404 Series 

 

Table 3

 

– Optional Components 

Summary of Contents for 1230-11ADB

Page 1: ...ELL DUAL BRIDGE Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ... repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment calibration cycles are typically established by the users own quality regimen When in doubt about a calibration cycle a good rule...

Page 3: ...ory customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and mounting hardware are not covered by warranty Mechanical damage...

Page 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Page 6: ...54 lb 25 5 kg Mounting Thread 1 3 4 12 UNF 3B No Metric Equivalent Sensing Element Strain Gage Strain Gage Mounting Torque 2600 in lb 294 Nm Deflection at Full Scale Capacity 0 001 in 0 025 mm Electrical Connector PT02E 10 6P 2 PT02E 10 6P 2 All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifica...

Page 7: ... 13 59229 PCB Load Torque Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Load and Torque Inc REVISIONS REV DESCRIPTION ECO NR RELEASED FOR PRODUCTION 06 18 14 PTE 43066 CONNECTOR B 2 62 66 5 1 00 25 4 3 12 79 2 4 85 123 2 1 41 35 8 CONNECTOR SHROUD n3 75 95 3 16X 1 2 20 X 3 25 ...

Page 8: ...ELECTRICAL INSTALLATION 5 5 1 Electrical Drawing Western Regional Std 5 5 2 Cable Grounding Considerations 5 6 0 CALIBRATION 5 6 1 Calibration Certificate Description 5 6 1 1 Measured Output 6 6 1 2 Hysteresis 6 6 1 3 Best Fit Output 6 6 1 4 Strain Gage Measurements 6 6 1 5 Shunt Calibration Standard Resistor 6 6 1 6 Static Error Band SEB 6 7 0 MOMENT COMPENSATION 6 8 0 FATIGUE OVERLOAD 6 9 0 SHUN...

Page 9: ...the parts under test have completed the planned fatigue test cycles 3 0 OVERVIEW Dual bridge fatigue rated load cells are specifically designed for repetitive testing Applications include material testing component life cycle testing and structural testing The load cells contain an extra bridge to be used either as a back up incase the primary bridge fails during a test or just for confirmation of...

Page 10: ...note that the performance of the load cell may be affected if the factory installed mounting base is removed or modified 4 2 Mounting Load Cell to a Standard Base or Custom Fixture If the load cell is to be mounted to a custom base or fixture it is necessary to consider the following 1 The thickness of the new base or fixture must not be any thinner than the original base supplied with the load ce...

Page 11: ...ector Protector 1 Remove the 4 40 x 1 4 button head cap screws from the connector using a 1 16 hex wrench as shown in Figure 7 2 Carefully pull the connector from the load cell Take care to not break or loosen any wire connections 3 Slide the connector protector 54823 02 behind the connector as shown in Figure 8 4 Apply a drop Loctite Threadlocker Blue 242 to the threads of the 4 40 x 3 8 socket h...

Page 12: ...m wiring may result in unstable or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded If the power cord ground and structure ground are not at the same potential it may be necessary to provide a secure structure ground perhaps by driving a copper rod and a...

Page 13: ...ls are adjusted such that the error due to eccentric loading is less than 0 1 per inch of offset Side loads moments and torque must be limited to avoid electrical or mechanical damage to the load cell The allowable extraneous loads are directly related to the rated capacity of the load cell Maximum side force allowed is 100 of the rated capacity for the load cell Maximum moment allowed is 100 of t...

Page 14: ...ation certificate and adjust the span or gain until the display reads the force value stated on the certificate 4 Repeat steps 1 3 to verify that a valid calibration setting has been obtained 5 If possible apply a known load to the measurement system to further verify that the calibration has been accurately set up 9 3 Estimating Shunt Resistor for a Given Load The following formula can be used to...

Page 15: ...bridge resistances A D excitation and B C the signal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB factory for assistance 11 2 1 Estimating Bridge Balance using an Ohm Meter A load cell that has been severely overloaded will exhibit a significant zero offset in the Wheatstone bridge It can be useful to estimate the bridge unbalance using a digit...

Page 16: ...e factory for any reason For calibration services if possible a copy of the purchase order covering the requested services should be included with the returned load cell 13 0 WARRANTY Standard warranty on dual bridge load cells covers parts and workmanship For full details refer to the Warranty Statement supplied with each load cell If the load cell is defective for reasons other than overloads re...