PCB Piezotronics 1203-14A Installation And Operating Manual Download Page 11

 

PCB Load & Torque, Inc.   

 Toll-Free in USA 866-684-7107   

 716-684-0001   

 www.pcbloadtorque.com

 

 

 

GENERAL PURPOSE LOW PROFILE LOAD CELL OPERATION MANUAL 

 

6

 

6.1.1 Measured Output 

 
The  applied  load  starting  at  zero  is  measured  in  five 
increments  to  full  scale.  Output  (mV/V)  is  measured  at  each 
increment.  The  straight-line  from  zero  to  the  full  scale 
measurement  is  compared  to  the  measured  readings  at  each 
increment  to  calculate  the  error  at  each  load  increment.    The 
deviations  (%  Full  Scale)  corresponding  to  non-linearity  at 
each measurement increment are then calculated. 

6.1.2 Hysteresis 

 
The  difference  between  the  ascending  and  descending 
measured readings at 40% of full scale is used to calculate the 
hysteresis value. 

6.1.3 Best Fit Output 

 
The  best  fit  calibration  second-order  equation  has  been 
calculated  from  the  calibration  data  by  the  method  of  least 
squares.  Deviation  between  measured  output  and  best-fit 
output  is  calculated  and  displayed  in  the  column  next  to  the 
best-fit  output  for  each  measurement  increment.  The 
deviations  (%  Full  Scale)  of  measured  outputs  from  the 
calculated best fit are tabulated for each measured reading. 

6.1.4 Strain Gage Measurements 

 
 
 
 
 
 
 
 
 
 
 

6.1.5 Shunt Calibration Standard Resistor 

 
All low profile load cell calibrations use a 120k

 (

±

0.1%) (for 

1200  series)  or  a  60k

  (

±

0.01%)  (for  1400  series)  precision 

resistor  shunt  calibration  value  that  is  supplied  into  the 
calibration report. 

6.1.6 Static Error Band (SEB) 

 
The  static  error  band  (SEB)  is  determined  by  the  maximum 
deviations of  the ascending and descending calibration points 
from  the  best  fit  straight  line  through  zero  output.  The  SEB 
includes the effects of nonlinearity, hysteresis, and non-return 
to minimum load. 
 
 

7.0 MOMENT COMPENSATION 

 
When  an  eccentric  axial  load  is 
applied to a load cell that is not 
coincident  exactly  with  the 
centerline  of  the  load  cell,  it 
introduces  bending  moments 
into  the  load  cell  structure.  The 
strain  gage  bridges  on  general 
purpose  low  profile  load  cells 
are  adjusted  such  that  the  error 
due  to  eccentric  loading  is  less 
than 

±

0.25%  per  inch  of  offset 

for  1200  series  load  cells  and 

±

0.1%  per  inch  of  offset  for 

fatigue-rated  1400  series  load 
cells. 
 
Side  loads,  moments  and  torque  must  be  limited  to  avoid 
electrical  or  mechanical  damage  to  the  load  cell.  The 
allowable  extraneous  loads  are  directly  related  to  the  rated 
capacity of the load cell. 
 

 

Maximum side force allowed is 100% of the rated 
capacity for the load cell. 

 

Maximum  moment  allowed  is  100%  of  the  rated 
capacity times 1 inch. 

 

Maximum  torque  allowed  about  the  measurement 
axis is rated capacity times 1 inch. 

 

8.0 FATIGUE & OVERLOAD 

 
Fatigue-rated low profile load cells (1400 series) are designed 
and  manufactured  to  perform  accurately  and  reliably  for  100 
million fully reversed full scale load cycles.  They have been 
specifically  designed  for  applications  involving  high  cycle 
fatigue  loading  on  products  and  structures.    The  designs  are 
based  on  nearly  three  decades  (since  1986)  of  experience  in 
the design and manufacture of precision low-height load cells. 

 

The load cell structure is machined from aircraft quality alloy 
steel, heat treated, and finished with attention to details critical 
for  maximum  fatigue  life.    For  maximum  stiffness  and  low 
stresses  throughout  the  load  cell  structure,  the  maximum 
stresses are concentrated precisely under the strain gages. 

 

The  load  cell  structural  safety  factor  against  fatigue  failure  is 
in excess of 2.5 for alloy steel and 2 for aluminum.  See Figure 
14. The maximum shear stress measured by the bonded strain 
gages is less than 40% of the endurance limit shear stress for 
alloy steel and 48% for aluminum.  
 
 
 
 

Bridge Resistance: 

350 Ohm Nominal 

Excitation: 

+P(A) to –P(D) Ohms 

Signals: 

+S(B) to –S(C) Ohms 

Leakage to Ground: 

> 5k GOhm 

Bridge Unbalance: 

±

1.0% Full Scale 

Output: 

4.0 mV/V Nominal (1200 Series) 

 

2.0 mV/V Nominal (1400 Series) 

Maximum Voltage: 

20 VDC 

 

Table 7

 

– Strain Gage Measurements 

 

M

X

 

M

Y

 

Primary 

Axis 

Figure 13

 

- Extraneous Loads 

Summary of Contents for 1203-14A

Page 1: ...AD CELL Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...4 5 0 ELECTRICAL INSTALLATION 5 5 1 Electrical Drawing Western Regional Std 5 5 2 Cable Grounding Considerations 5 6 0 CALIBRATION 5 6 1 Calibration Certificate Description 5 6 1 1 Measured Output 6...

Page 7: ...evaluated for potential failure modes of the fixtures It is important to use threaded fasteners properly when assembling test fixtures used with the load cells Fixtures with inadequately designed bol...

Page 8: ...modified 4 2 Mounting Load Cell to a Standard Base or Custom Fixture If the load cell is to be mounted to a custom base or fixture it is necessary to consider the following 1 The thickness of the new...

Page 9: ...f Connector Protector 1 Remove the 4 40 x 1 4 button head cap screws from the connector using a 1 16 hex wrench as shown in Figure 7 2 Carefully pull the connector from the load cell Take care to not...

Page 10: ...A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded If the power cord g...

Page 11: ...into the load cell structure The strain gage bridges on general purpose low profile load cells are adjusted such that the error due to eccentric loading is less than 0 25 per inch of offset for 1200 s...

Page 12: ...s on the load cell If possible remove all loads 2 Power up the host signal conditioner and connect it to the load cell via appropriate cable and allow for a 30 minute warm up 3 Set the load indicator...

Page 13: ...e load cell is mounted on 5 Check the signal conditioning electronics for proper setup 6 Check the insulation resistance of shielded conductors for short circuits 7 Check isolation resistance load cel...

Page 14: ...der Please request a return material authorization RMA before sending a load cell back to the factory for any reason For calibration services if possible a copy of the purchase order covering the requ...

Page 15: ...ng Element Strain Gage Strain Gage Mounting Torque 200 lbf in 23 Nm Deflection at Full Scale Capacity 0 002 in 0 05 mm Electrical Connector PT02E 10 6P PT02E 10 6P All specifications are at room tempe...

Page 16: ...LOAD CELLS 5 8 18UNF 3B METRIC M LOAD CELLS M16X2 0 4H SPECIAL LOAD CELL 1 14UNS 3B 2 50 63 5 030 76 45 THREAD x 88 22 3 STANDARD LOAD CELLS 5 8 18 UNF 3B METRIC M LOAD CELLS M16 X 2 0 4H SPECIAL LOA...

Reviews: