PCB Piezotronics 088A Installation And Operating Manual Download Page 11

 

6

 
The Modal Punch can be calibrated by hitting a freely-suspended mass instrumented with 
a quartz reference accelerometer. The mass, pendulously suspended or placed on a piece 
of foam rubber, will behave as a rigid body.  According to Newton

s second law of 

motion, at any instant in time, the force experienced by the mass is simply the mass 
multiplied by the measured acceleration.  
 
On a storage oscilloscope, dividing the peak output signal of the hammer (mV) by the 
mass (lb) times the peak acceleration (g

s) gives the hammer sensitivity directly in 

mV/lb.  Calibration on a FFT analyzer produces the same result as a function of 
frequency.  The transfer function of acceleration versus force produces a calibration 
constant (ideally 1/M) at each discrete frequency.  This calibration curve is linear over a 
specified frequency range.  
 
PCB

®

s Model 963A Gravimetric Calibrator is ideally suited for calibration of the Modal 

Punch.  The device calibrates accelerometers, force sensors, and PCB

®  

Impulse Force 

Test Hammers and tests the functional transfer behavior, sensitivity, and phase of sensor 
structures.   
 
 

7.0 Maintenance  

The rigid construction of the Modal Punch is designed to provide reliable, long-term 
service life.  Problems may arise when the instrument is exposed to temperatures 
exceeding the specific range, prolonged exposure to moisture, or extremely high 
mechanical shock.  There are no user-serviceable components in the Modal Punch, so 
users should not attempt to make repairs to the unit (and doing so will void the unit

warranty).  
 
Should service be required, the Modal Punch should be returned to the factory, along 
with a description of the problem, in order to expedite the repair process.  Before sending 
the Modal Punch back to the factory for service, PCB recommends that all cables and 
connections be checked to verify that they are not the cause of the problem. 
 
 

8.0 Cautions   

Although the Modal Punch is rugged in construction, personal injury or damage to the 
unit may result from misuse.  When observed, the following precautions can ensure long 
service life and safe, accurate data acquisition: 
 
1). Do not attempt to dismantle the sensor element from the Modal Punch structure.  All 
service should be performed at the factory.  
 
2). Never generate more than five times the rated impact force range for the Modal 
Punch.  In general, observe the maximum force rating for a 10V output.  100 volts will 
destroy the precision built-in micro-electronics and/or the sensing element.  
 

Summary of Contents for 088A

Page 1: ...Model 084A80 extender Model Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...ng the resultant motion with an accelerometer s Integration of the acceleration signal yields velocity or displacement data for use in computing classical transfer characteristics such as inertance co...

Page 7: ...41 supplied just below the impact surface An Extension Shaft model 084A80 is also supplied to adjust the Modal Punch to the desired length Additional shafts may also be purchased A foam grip is suppli...

Page 8: ...for higher frequency response B A softer tip is recommended for improved low frequency response 3 Connect the Modal Punch to an ICP power supply or equivalent with a coaxial cable not supplied The Mo...

Page 9: ...ling 716 684 0001 5 0 Testing To test the behavior of your structure and to tailor the frequency bandwidth of the impact force follow the below operation instructions 1 Position the Modal Punch on the...

Page 10: ...perfect this technique Distortion undershoot and oscillation of the analyzer display is caused by ringing of the analyzer s anti aliasing filters This behavior is normal To view the correct impulse wa...

Page 11: ...of the Modal Punch is designed to provide reliable long term service life Problems may arise when the instrument is exposed to temperatures exceeding the specific range prolonged exposure to moisture...

Page 12: ...to the user 9 0 CUSTOMER SERVICE PCB Piezotronics guarantees Total Customer Satisfaction If at any time for any reason you are not completely satisfied with any PCB product PCB will repair replace or...

Page 13: ...ment we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERSIONS Optional versions have identical specifications and accessories as li...

Page 14: ......

Reviews: