SERIAL OUTPUT PARAMETERS
This menu allows you to set the serial ports COM1 and COM2 and the communication parameters.
The instrument has two independent serial ports connected to a single connector:
COM1 with RS232 or RS422 / RS485 interface
COM2 with PROFINET interface.
baUd r
BAUd RATE COM1
It defines the baudrate of the serial port RS232.
The value must be set to the same value of the PC / PLC or remote
display.
Value to be selected:
2400
9600
19200
38400
115200
Default: 9600
Prot-1
COM1 PROTOCOL
It defines how to use the serial port RS232
None:
Serial communication OFF
Contin:
Continuous transmission of the weight string. It can be used,
for example, to drive a weight repeater. See details in the relevant
paragraph.
deman:
When the Operator presses the front button or through Input
2, a string of weight is transmitted. The command is accepted if the
weight is stable. Between two consecutive transmissions the weight
must have a variation of at least the parameter “Weight delta”.
Auto:
It’s automatically transferred to a string of weight when the
weight stabilizes at a value higher than the minimum weight (20
divisions). Between two consecutive transmissions, the weight must
have a variation of at least the parameter “Weight delta”.
Slav:
ASCII protocol. See details in the relevant paragraph.
Mdbs:
MODBUS RTU (slave) protocol. used only if PROT-2 is confi-
gured equal NONE. Selectable communication patrameters:
n-8-1
n-8-2
E-8-1
o-8-1
Default: n-8-1