![Pavone Sistemi DAT 500 Technical Manual Download Page 27](http://html.mh-extra.com/html/pavone-sistemi/dat-500/dat-500_technical-manual_763995027.webp)
Page 25
CaL1br
SET
FUN
ParaN
SET
FUN
In-oUt
f1LtEr
Not1on
auto 0
0 trac
0 band
0
0
0
0
0
PRG
SET
FUN
SET
FUN
SET
FUN
SET
FUN
SET
FUN
PRG
Select Value
PRG
PRG
Select Value
PRG
PRG
Enter Value
PRG
PRG
Select Value
PRG
PRG
Enter Value
PRG
0 ALL
0
SET
FUN
PRG
Select Value
PRG
WEIGHTING PARAMETERS
The parameters in this menu allow you to adjust the timing of the acquisition and updating of the display
and the manual or automatic zeroing that the transmitter performs.
F1LtEr
WEIGHT FILTER
This parameter adjusts the refresh speed of the display and the serial
and analog output.
Low values of the filter speed up the display refresh.
High values of the filter slow down the display refresh.
Value Update
Response
0
50 Hz
25 Hz
1
50 Hz
16 Hz
2
25 Hz
8 Hz
3
25 Hz
5 Hz
4
25 Hz
2.5 Hz
5
10 Hz
1.5 Hz
6
10 Hz
1 Hz
7
10 Hz
0.7 Hz
8
5 Hz
0.4 Hz
9
5 Hz
0.2 Hz
Default: 5
Notion
WEIGHT STABILITY
This parameter defines the divisions number needed to deem the
weight stable.
A large number of divisions allows the transmitter to detect quickly
the weight stability, which is needed when executing tare and print
commands.
Value Change
0
Always stable weight
1
Stability determined quickly
2
Stability determined with medium parameters
3
Stability determined accurately
4
Stability determined with the highest accuracy
Default: 2
auto 0
AUTOZERO UPON SWITCHIN ON
This parameter defines the value of the maximum resettable weight
when the instrument is switched on.
This operation corresponds to a zero calibration of the system and
is executed only if the weight is stable and below the set value.
Value from 0 to the value of the CAPAC parameter.
Default: 0