M o d e l   N o .   P S - 2 1 6 9

 D i s s o l v e d  O x y g e n

7

DO

2

 Calibration

Perform a one-point calibration using 100% humidified air as the known standard, 
which is equivalent to 100% saturated (pure) water. Prepare 100% humidified air as 
follows:

1.

Place about 5 mL (to a height of about 1 cm) of deionized water into a clean elec-
trode storage bottle. 

2.

Slip the cap and O-ring of the soaker bottle over the end of the probe. Insert the 
probe into the soaker bottle and screw on the cap. Adjust the height of the end of 
the probe to about 2 cm above the water.

3.

Shake the probe and bottle vigorously for about 10 seconds. Shake off any large 
water drops from the membrane.

4.

Hold the probe vertically so that it is in the air 

above

 the water, not in the water 

itself.

5.

Perform a one-point calibration on either the 

Dissolved Oxygen (%)

 or 

Dis-

solved Oxygen (mg/L)

 measurement. If you are calibrating the percentage mea-

surement, enter “100” as the known value. If you are calibrating the mg/L 
measurement, look up the known concentration in Appendix D. For instructions 
on calibrating in DataStudio, see Appendix A. For instructions on calibrating on 
the Xplorer GLX, refer to the Xplorer GLX Users’ Guide.

Measuring DO

2

 

Follow the instructions above to fill the probe with electrolyte solution and calibrate 
the measurement. Start data collection. Remove the storage bottle from the probe. 
Immerse the end of the probe in the solution to be measured. Gently swirl the probe or 
stir the solution to prevent localized oxygen depletion at the probe end. Wait for the 
reading to stabilize. Rinse the electrode with deionized water before measuring a dif-
ferent solution.

DO

2

 Probe Maintenance and Storage

The probe comes with extra filling solution, a replacement membrane cartridge and 
O-ring, and a plunger for installing the cartridge. Order the Electrode Maintenance 
Kit (CI-6541) to replace these items.

Electrode Cleaning

To maintain optimal performance, periodically clean the silver electrode and replace 
the electrolyte filling solution as follows:

1.

Unscrew and remove the cartridge housing at the end of the probe. Discard the 
old filling solution from the cartridge housing.

Shake

vigorously

Electrode

storage bottle

Water level

Probe tip

Preparation of 100%

humidified air for

calibration

Syringe

Plunger

Replacement

O-ring

Replacement

membrane cartridge

Filling

Solution

Remove loose silver oxide

from cathode

Summary of Contents for PS-2169

Page 1: ...e a replacement membrane cartridge and O ring and a plunger See illustration on page 7 699 06320 Conductivity Electrode 10x 699 06621 Required Equipment PASPORT Interface2 2 PASPORT interfaces include the AirLink IW 4100 or PS 2005 Xplorer GLX PS 2002 Xplorer PS 2000 PowerLink PS 2001 and USB Link PS 2100 See PASCO catalog or www pasco com Optional Equipment Fast response Temperature Probe PS 2135...

Page 2: ...ns below to set up the multi sensor with DataStudio software if you are using a computer or the Xplorer GLX if you are using it standalone without a computer For detailed information about each of the multi sensor s component sensors see Temperature on page 3 Conductivity on page 3 pH ISE ORP on page 4 and Dissolved Oxygen on page 6 Note Only the tips and handles of the probes should be immersed i...

Page 3: ...ith an object The included stainless steel probe can be used in both dry conditions and in liquids such as water and other mild chemicals and solutions Tip For better chemical resistance use a Teflon cover CI 6549 Temperature Calibration The temperature measurement can be calibrated however for most applications cali bration is not necessary Use a two point calibration with two standards of known ...

Page 4: ...er for most applications cali bration is not necessary Use a one point calibration with a standard solution of known conductivity For instructions on preparing a standard solution see Appendix C For instructions on calibrating in DataStudio see Appendix A For instructions on calibrating on the Xplorer GLX refer to the Xplorer GLX Users Guide Conductivity Electrode Maintenance And Storage To ensure...

Page 5: ...rode Maintenance and Storage Cleaning If the pH electrode becomes contaminated use one of these methods to clean and restore it After any of these procedures soak the electrode in a pH 7 buffer solution for 30 minutes General Contamination Soak the electrode in 0 1 molar hydrochloric acid HCl for 15 minutes Protein Deposits Soak the electrode in a solution of 1 pepsin in 0 1 molar HCl Inorganic De...

Page 6: ...red by the sensor as an electric current Since the rate of O2 diffusion across the membrane is proportional DO2 concentration the rate of these reactions and the resulting electric current is also proportional to DO2 concentration The rate at which the probe processes oxygen also depends on temperature To correct for temperature variation the probe contains a temperature sensing thermistor which a...

Page 7: ...ructions on calibrating on the Xplorer GLX refer to the Xplorer GLX Users Guide Measuring DO2 Follow the instructions above to fill the probe with electrolyte solution and calibrate the measurement Start data collection Remove the storage bottle from the probe Immerse the end of the probe in the solution to be measured Gently swirl the probe or stir the solution to prevent localized oxygen depleti...

Page 8: ...nd of the housing as illustrated right 3 Examine the O ring in the housing If it is damaged remove it with fine tipped tweezers and insert the replacement 4 Use the plunger to push the replacement cartridge into the housing until it is seated at the tip of the housing 5 Follow the Adding Filling Solution instructions on page 6 to fill and reassem ble the probe Storage Short term up to two weeks Ad...

Page 9: ...Range three user selectable ranges 0 to 1000 µS cm 0 to 10000 µS cm 0 to 100000 µS cm with included 10x probe 0 to 100 µS cm 0 to 1000 µS cm 0 to 10000 µS cm with optional 1x probe Accuracy 10 of full scale without calibration Resolution better than 0 05 of full scale pH ISE ORP probe voltage Electrode Connector standard BNC Input Resistance 1012 Ω Voltage Range 2000 mV to 2000 mV Voltage Resoluti...

Page 10: ...ment field select Temperature or pH 3 Ensure that Calibrate all similar measurements simulta neously is not selected 4 From the Calibration Type menu select 2 Point Adjust Slope and Offset if it is not already selected 5 Place the temperature or pH probe in the first known standard 6 Enter the known temperature or pH of the standard in the Standard Value field for Point 1 7 Observe the Present Sen...

Page 11: ...resent Sensor Measurement and wait until it has stabilized For conductivity ensure that the present mea surement is within the selected sensor range If it is not press a button on the multi sensor to select a higher range 8 Click the Read From Sensor button for Point 2 9 Click OK to accept the new calibration Appendix B EZ Screen If you are using the Water Quality MultiMeasure Sensor with a PASPOR...

Page 12: ...nstruments calibrated in absolute units multiply the tabular values by 0 999505 Temp Atmospheric pressure in millimeters of mercury and inches of mercury C 795 31 30 790 31 10 785 30 91 780 30 71 775 30 51 770 30 31 765 30 12 760 29 92 755 29 72 750 29 53 745 29 33 740 29 13 735 28 94 730 28 74 725 28 54 720 28 35 715 28 15 710 27 95 705 27 76 700 27 56 0 0 15 3 15 2 15 1 15 0 14 9 14 8 14 7 14 6 ...

Page 13: ...0 3 10 3 10 2 10 1 10 1 10 0 9 9 9 9 9 8 9 7 9 7 9 6 9 5 9 5 9 4 9 3 9 3 15 5 10 4 10 4 10 3 10 2 10 2 10 1 10 0 10 0 9 9 9 8 9 8 9 7 9 6 9 6 9 5 9 4 9 4 9 3 9 2 9 2 16 0 10 3 10 2 10 2 10 1 10 0 10 0 9 9 9 8 9 8 9 7 9 7 9 6 9 5 9 5 9 4 9 3 9 3 9 2 9 1 9 1 16 5 10 2 10 1 10 1 10 0 9 9 9 9 9 8 9 7 9 7 9 6 9 5 9 5 9 4 9 4 9 3 9 2 9 2 9 1 9 0 9 0 17 0 10 1 10 0 10 0 9 9 9 8 9 8 9 7 9 6 9 6 9 5 9 4 9 ...

Page 14: ...1 6 1 38 0 6 9 6 9 6 8 6 8 6 7 6 7 6 7 6 6 6 6 6 5 6 5 6 4 6 4 6 3 6 3 6 2 6 2 6 1 6 1 6 0 38 5 6 9 6 8 6 8 6 7 6 7 6 6 6 6 6 6 6 5 6 5 6 4 6 4 6 3 6 3 6 2 6 2 6 1 6 1 6 0 6 0 39 0 6 8 6 8 6 7 6 7 6 6 6 6 6 5 6 5 6 5 6 4 6 4 6 3 6 3 6 2 6 2 6 1 6 1 6 0 6 0 6 0 39 5 6 8 6 7 6 7 6 6 6 6 6 5 6 5 6 5 6 4 6 4 6 3 6 3 6 2 6 2 6 1 6 1 6 0 6 0 6 0 5 9 40 0 6 7 6 7 6 6 6 6 6 5 6 5 6 4 6 4 6 4 6 3 6 3 6 2 6...

Page 15: ... 16 0 9 0 8 9 8 9 8 8 8 7 8 7 8 6 8 5 8 5 8 4 8 3 8 3 8 2 8 1 8 1 8 0 7 9 7 9 7 8 7 7 16 5 8 9 8 8 8 8 8 7 8 6 8 6 8 5 8 4 8 4 8 3 8 2 8 2 8 1 8 0 8 0 7 9 7 8 7 8 7 7 7 7 17 0 8 8 8 7 8 7 8 6 8 5 8 5 8 4 8 3 8 3 8 2 8 2 8 1 8 0 8 0 7 9 7 8 7 8 7 7 7 6 7 6 17 5 8 7 8 6 8 6 8 5 8 5 8 4 8 3 8 3 8 2 8 1 8 1 8 0 7 9 7 9 7 8 7 7 7 7 7 6 7 6 7 5 18 0 8 6 8 6 8 5 8 4 8 4 8 3 8 2 8 2 8 1 8 0 8 0 7 9 7 9 7 ...

Page 16: ...5 6 5 6 5 5 5 5 5 4 5 4 5 3 5 3 5 3 5 2 37 5 6 0 6 0 6 0 5 9 5 9 5 8 5 8 5 7 5 7 5 6 5 6 5 5 5 5 5 4 5 4 5 3 5 3 5 3 5 2 5 2 38 0 6 0 6 0 5 9 5 9 5 8 5 8 5 7 5 7 5 6 5 6 5 5 5 5 5 4 5 4 5 3 5 3 5 3 5 2 5 2 5 1 38 5 6 0 5 9 5 9 5 8 5 8 5 7 5 7 5 6 5 6 5 5 5 5 5 4 5 4 5 4 5 3 5 3 5 2 5 2 5 1 5 1 39 0 5 9 5 9 5 8 5 8 5 7 5 7 5 6 5 6 5 5 5 5 5 4 5 4 5 4 5 3 5 3 5 2 5 2 5 1 5 1 5 0 39 5 5 9 5 8 5 8 5 7...

Page 17: ... 0 952 0 948 28 0 1 000 0 997 0 994 0 991 0 987 0 984 0 981 0 978 0 975 0 972 0 968 0 965 0 962 0 959 0 955 0 952 0 949 29 0 1 000 0 997 0 994 0 991 0 988 0 984 0 981 0 978 0 975 0 972 0 969 0 965 0 962 0 959 0 956 0 952 0 949 30 0 1 000 0 997 0 994 0 991 0 988 0 985 0 981 0 978 0 975 0 972 0 969 0 966 0 962 0 959 0 956 0 953 0 950 31 0 1 000 0 997 0 994 0 991 0 988 0 985 0 982 0 978 0 975 0 972 0...

Page 18: ... 0 925 0 921 0 918 0 914 0 911 0 907 0 904 0 901 0 897 0 894 0 890 27 0 0 945 0 942 0 938 0 935 0 932 0 928 0 925 0 922 0 918 0 915 0 911 0 908 0 905 0 901 0 898 0 894 0 891 28 0 0 946 0 942 0 939 0 936 0 932 0 929 0 926 0 922 0 919 0 915 0 912 0 909 0 905 0 902 0 898 0 895 0 892 29 0 0 946 0 943 0 939 0 936 0 933 0 929 0 926 0 923 0 919 0 916 0 913 0 909 0 906 0 903 0 899 0 896 0 892 30 0 0 946 0...

Page 19: ...0 0 857 0 853 0 850 0 846 0 842 0 839 0 835 0 832 0 828 25 0 0 886 0 882 0 879 0 875 0 872 0 868 0 865 0 861 0 858 0 854 0 851 0 847 0 843 0 840 0 836 0 833 0 829 26 0 0 887 0 883 0 880 0 876 0 873 0 869 0 866 0 862 0 859 0 855 0 852 0 848 0 844 0 841 0 837 0 834 0 830 27 0 0 887 0 884 0 880 0 877 0 874 0 870 0 867 0 863 0 860 0 856 0 853 0 849 0 845 0 842 0 838 0 835 0 831 28 0 0 888 0 885 0 881 ...

Page 20: ... 816 0 812 0 809 0 805 0 801 0 797 0 794 0 790 0 786 0 783 0 779 0 775 0 771 0 768 0 764 0 760 21 0 0 821 0 817 0 814 0 810 0 806 0 802 0 799 0 795 0 791 0 788 0 784 0 780 0 777 0 773 0 769 0 766 0 762 22 0 0 822 0 818 0 815 0 811 0 807 0 804 0 800 0 796 0 793 0 789 0 785 0 782 0 778 0 774 0 771 0 767 0 763 23 0 0 823 0 820 0 816 0 812 0 809 0 805 0 801 0 798 0 794 0 790 0 787 0 783 0 779 0 776 0 ...

Page 21: ...ofit Reproduction under any other circumstances without the written consent of PASCO scientific is prohibited Trademarks PASCO PASCO scientific DataStudio MultiMeasure Sensors PASPORT and ScienceWorkshop are trademarks or registered trademarks of PASCO scientific in the United States and or in other countries All other brands products or service names are or may be trademarks or service marks of a...

Reviews: