Pasport PS-2169 Instruction Manual Download Page 6

®

W a te r  Q u a l i t y   M u l ti M e a s u r e   S e n s o r

D i s s o l v e d  O x y g e n

6

Dissolved Oxygen

The DO

2

 component of the multi-sensor is designed for use in aqueous media at tem-

peratures ranging from 0 °C to 50 °C. For greatest accuracy, the following require-
ments should be met: 

Although the unit is temperature compensated, it should be calibrated at approxi-
mately the same temperature as the test solution.

The sensor must equilibrate for a short period when the temperature of the test 
solution changes.

The test solution should constantly flow past the membrane of the probe. This 
can be accomplished by gently swirling the electrode during measurement.

The DO

2

 probe consists of a platinum (Pt) cathode and a silver (Ag) anode sur-

rounded by potassium chloride (KCl

(aq)

) electrolyte filling solution with a silicon 

membrane separating the filling solution from the test solution. When the probe is 
placed in an aqueous medium that contains dissolved oxygen, O

2

 molecules diffuse 

across the membrane and react with water molecules (H

2

O) in the presence of elec-

trons from the cathode to form hydroxide ions (OH

-

). Hydroxide ions diffuse to the 

anode and react with silver atoms, forming silver oxide (Ag

2

O), H

2

O, and free elec-

trons. The net movement of negative charge (electrons) from the cathode to the anode 
is measured by the sensor as an electric current. Since the rate of O

2

 diffusion across 

the membrane is proportional DO

2

 concentration, the rate of these reactions and the 

resulting electric current is also proportional to DO

2

 concentration. The rate at which 

the probe processes oxygen also depends on temperature. To correct for temperature 
variation, the probe contains a temperature-sensing thermistor, which adjusts the gain 
of an amplifier in the multi-sensor.

Adding Filling Solution

Prior to its initial use, and after long-term storage or cleaning, prepare the probe by 
filling it with electrolyte solution as follows: 

1.

Remove the storage bottle from the end of the probe.

2.

Unscrew and remove the cartridge housing at the end of the probe, as illustrated 
(right). 

3.

Pull about 10 mL of new filling solution into the syringe (included with the 
probe). Be careful not to introduce air bubbles.

4.

Place the tip of the syringe in the cartridge housing, very close to—but not touch-
ing—the membrane. Slowly fill the housing to about 5 mm from the top. Agitate 
the housing while filling to avoid introducing air bubbles.

5.

Replace the housing over the electrode and screw it into place. Excess filling 
solution will spill from the housing as you do this.

6.

Rinse and dry the outside of the probe.

membrane

platinum

cathode

cartridge

housing

silver

anode

filling

solution

cathode

insulator

Detail of DO

2

 Probe Tip

Unscrew

cartridge

housing

Replace

cartridge

housing

Fill cartridge

housing

fill to here

Summary of Contents for PS-2169

Page 1: ...e a replacement membrane cartridge and O ring and a plunger See illustration on page 7 699 06320 Conductivity Electrode 10x 699 06621 Required Equipment PASPORT Interface2 2 PASPORT interfaces include the AirLink IW 4100 or PS 2005 Xplorer GLX PS 2002 Xplorer PS 2000 PowerLink PS 2001 and USB Link PS 2100 See PASCO catalog or www pasco com Optional Equipment Fast response Temperature Probe PS 2135...

Page 2: ...ns below to set up the multi sensor with DataStudio software if you are using a computer or the Xplorer GLX if you are using it standalone without a computer For detailed information about each of the multi sensor s component sensors see Temperature on page 3 Conductivity on page 3 pH ISE ORP on page 4 and Dissolved Oxygen on page 6 Note Only the tips and handles of the probes should be immersed i...

Page 3: ...ith an object The included stainless steel probe can be used in both dry conditions and in liquids such as water and other mild chemicals and solutions Tip For better chemical resistance use a Teflon cover CI 6549 Temperature Calibration The temperature measurement can be calibrated however for most applications cali bration is not necessary Use a two point calibration with two standards of known ...

Page 4: ...er for most applications cali bration is not necessary Use a one point calibration with a standard solution of known conductivity For instructions on preparing a standard solution see Appendix C For instructions on calibrating in DataStudio see Appendix A For instructions on calibrating on the Xplorer GLX refer to the Xplorer GLX Users Guide Conductivity Electrode Maintenance And Storage To ensure...

Page 5: ...rode Maintenance and Storage Cleaning If the pH electrode becomes contaminated use one of these methods to clean and restore it After any of these procedures soak the electrode in a pH 7 buffer solution for 30 minutes General Contamination Soak the electrode in 0 1 molar hydrochloric acid HCl for 15 minutes Protein Deposits Soak the electrode in a solution of 1 pepsin in 0 1 molar HCl Inorganic De...

Page 6: ...red by the sensor as an electric current Since the rate of O2 diffusion across the membrane is proportional DO2 concentration the rate of these reactions and the resulting electric current is also proportional to DO2 concentration The rate at which the probe processes oxygen also depends on temperature To correct for temperature variation the probe contains a temperature sensing thermistor which a...

Page 7: ...ructions on calibrating on the Xplorer GLX refer to the Xplorer GLX Users Guide Measuring DO2 Follow the instructions above to fill the probe with electrolyte solution and calibrate the measurement Start data collection Remove the storage bottle from the probe Immerse the end of the probe in the solution to be measured Gently swirl the probe or stir the solution to prevent localized oxygen depleti...

Page 8: ...nd of the housing as illustrated right 3 Examine the O ring in the housing If it is damaged remove it with fine tipped tweezers and insert the replacement 4 Use the plunger to push the replacement cartridge into the housing until it is seated at the tip of the housing 5 Follow the Adding Filling Solution instructions on page 6 to fill and reassem ble the probe Storage Short term up to two weeks Ad...

Page 9: ...Range three user selectable ranges 0 to 1000 µS cm 0 to 10000 µS cm 0 to 100000 µS cm with included 10x probe 0 to 100 µS cm 0 to 1000 µS cm 0 to 10000 µS cm with optional 1x probe Accuracy 10 of full scale without calibration Resolution better than 0 05 of full scale pH ISE ORP probe voltage Electrode Connector standard BNC Input Resistance 1012 Ω Voltage Range 2000 mV to 2000 mV Voltage Resoluti...

Page 10: ...ment field select Temperature or pH 3 Ensure that Calibrate all similar measurements simulta neously is not selected 4 From the Calibration Type menu select 2 Point Adjust Slope and Offset if it is not already selected 5 Place the temperature or pH probe in the first known standard 6 Enter the known temperature or pH of the standard in the Standard Value field for Point 1 7 Observe the Present Sen...

Page 11: ...resent Sensor Measurement and wait until it has stabilized For conductivity ensure that the present mea surement is within the selected sensor range If it is not press a button on the multi sensor to select a higher range 8 Click the Read From Sensor button for Point 2 9 Click OK to accept the new calibration Appendix B EZ Screen If you are using the Water Quality MultiMeasure Sensor with a PASPOR...

Page 12: ...nstruments calibrated in absolute units multiply the tabular values by 0 999505 Temp Atmospheric pressure in millimeters of mercury and inches of mercury C 795 31 30 790 31 10 785 30 91 780 30 71 775 30 51 770 30 31 765 30 12 760 29 92 755 29 72 750 29 53 745 29 33 740 29 13 735 28 94 730 28 74 725 28 54 720 28 35 715 28 15 710 27 95 705 27 76 700 27 56 0 0 15 3 15 2 15 1 15 0 14 9 14 8 14 7 14 6 ...

Page 13: ...0 3 10 3 10 2 10 1 10 1 10 0 9 9 9 9 9 8 9 7 9 7 9 6 9 5 9 5 9 4 9 3 9 3 15 5 10 4 10 4 10 3 10 2 10 2 10 1 10 0 10 0 9 9 9 8 9 8 9 7 9 6 9 6 9 5 9 4 9 4 9 3 9 2 9 2 16 0 10 3 10 2 10 2 10 1 10 0 10 0 9 9 9 8 9 8 9 7 9 7 9 6 9 5 9 5 9 4 9 3 9 3 9 2 9 1 9 1 16 5 10 2 10 1 10 1 10 0 9 9 9 9 9 8 9 7 9 7 9 6 9 5 9 5 9 4 9 4 9 3 9 2 9 2 9 1 9 0 9 0 17 0 10 1 10 0 10 0 9 9 9 8 9 8 9 7 9 6 9 6 9 5 9 4 9 ...

Page 14: ...1 6 1 38 0 6 9 6 9 6 8 6 8 6 7 6 7 6 7 6 6 6 6 6 5 6 5 6 4 6 4 6 3 6 3 6 2 6 2 6 1 6 1 6 0 38 5 6 9 6 8 6 8 6 7 6 7 6 6 6 6 6 6 6 5 6 5 6 4 6 4 6 3 6 3 6 2 6 2 6 1 6 1 6 0 6 0 39 0 6 8 6 8 6 7 6 7 6 6 6 6 6 5 6 5 6 5 6 4 6 4 6 3 6 3 6 2 6 2 6 1 6 1 6 0 6 0 6 0 39 5 6 8 6 7 6 7 6 6 6 6 6 5 6 5 6 5 6 4 6 4 6 3 6 3 6 2 6 2 6 1 6 1 6 0 6 0 6 0 5 9 40 0 6 7 6 7 6 6 6 6 6 5 6 5 6 4 6 4 6 4 6 3 6 3 6 2 6...

Page 15: ... 16 0 9 0 8 9 8 9 8 8 8 7 8 7 8 6 8 5 8 5 8 4 8 3 8 3 8 2 8 1 8 1 8 0 7 9 7 9 7 8 7 7 16 5 8 9 8 8 8 8 8 7 8 6 8 6 8 5 8 4 8 4 8 3 8 2 8 2 8 1 8 0 8 0 7 9 7 8 7 8 7 7 7 7 17 0 8 8 8 7 8 7 8 6 8 5 8 5 8 4 8 3 8 3 8 2 8 2 8 1 8 0 8 0 7 9 7 8 7 8 7 7 7 6 7 6 17 5 8 7 8 6 8 6 8 5 8 5 8 4 8 3 8 3 8 2 8 1 8 1 8 0 7 9 7 9 7 8 7 7 7 7 7 6 7 6 7 5 18 0 8 6 8 6 8 5 8 4 8 4 8 3 8 2 8 2 8 1 8 0 8 0 7 9 7 9 7 ...

Page 16: ...5 6 5 6 5 5 5 5 5 4 5 4 5 3 5 3 5 3 5 2 37 5 6 0 6 0 6 0 5 9 5 9 5 8 5 8 5 7 5 7 5 6 5 6 5 5 5 5 5 4 5 4 5 3 5 3 5 3 5 2 5 2 38 0 6 0 6 0 5 9 5 9 5 8 5 8 5 7 5 7 5 6 5 6 5 5 5 5 5 4 5 4 5 3 5 3 5 3 5 2 5 2 5 1 38 5 6 0 5 9 5 9 5 8 5 8 5 7 5 7 5 6 5 6 5 5 5 5 5 4 5 4 5 4 5 3 5 3 5 2 5 2 5 1 5 1 39 0 5 9 5 9 5 8 5 8 5 7 5 7 5 6 5 6 5 5 5 5 5 4 5 4 5 4 5 3 5 3 5 2 5 2 5 1 5 1 5 0 39 5 5 9 5 8 5 8 5 7...

Page 17: ... 0 952 0 948 28 0 1 000 0 997 0 994 0 991 0 987 0 984 0 981 0 978 0 975 0 972 0 968 0 965 0 962 0 959 0 955 0 952 0 949 29 0 1 000 0 997 0 994 0 991 0 988 0 984 0 981 0 978 0 975 0 972 0 969 0 965 0 962 0 959 0 956 0 952 0 949 30 0 1 000 0 997 0 994 0 991 0 988 0 985 0 981 0 978 0 975 0 972 0 969 0 966 0 962 0 959 0 956 0 953 0 950 31 0 1 000 0 997 0 994 0 991 0 988 0 985 0 982 0 978 0 975 0 972 0...

Page 18: ... 0 925 0 921 0 918 0 914 0 911 0 907 0 904 0 901 0 897 0 894 0 890 27 0 0 945 0 942 0 938 0 935 0 932 0 928 0 925 0 922 0 918 0 915 0 911 0 908 0 905 0 901 0 898 0 894 0 891 28 0 0 946 0 942 0 939 0 936 0 932 0 929 0 926 0 922 0 919 0 915 0 912 0 909 0 905 0 902 0 898 0 895 0 892 29 0 0 946 0 943 0 939 0 936 0 933 0 929 0 926 0 923 0 919 0 916 0 913 0 909 0 906 0 903 0 899 0 896 0 892 30 0 0 946 0...

Page 19: ...0 0 857 0 853 0 850 0 846 0 842 0 839 0 835 0 832 0 828 25 0 0 886 0 882 0 879 0 875 0 872 0 868 0 865 0 861 0 858 0 854 0 851 0 847 0 843 0 840 0 836 0 833 0 829 26 0 0 887 0 883 0 880 0 876 0 873 0 869 0 866 0 862 0 859 0 855 0 852 0 848 0 844 0 841 0 837 0 834 0 830 27 0 0 887 0 884 0 880 0 877 0 874 0 870 0 867 0 863 0 860 0 856 0 853 0 849 0 845 0 842 0 838 0 835 0 831 28 0 0 888 0 885 0 881 ...

Page 20: ... 816 0 812 0 809 0 805 0 801 0 797 0 794 0 790 0 786 0 783 0 779 0 775 0 771 0 768 0 764 0 760 21 0 0 821 0 817 0 814 0 810 0 806 0 802 0 799 0 795 0 791 0 788 0 784 0 780 0 777 0 773 0 769 0 766 0 762 22 0 0 822 0 818 0 815 0 811 0 807 0 804 0 800 0 796 0 793 0 789 0 785 0 782 0 778 0 774 0 771 0 767 0 763 23 0 0 823 0 820 0 816 0 812 0 809 0 805 0 801 0 798 0 794 0 790 0 787 0 783 0 779 0 776 0 ...

Page 21: ...ofit Reproduction under any other circumstances without the written consent of PASCO scientific is prohibited Trademarks PASCO PASCO scientific DataStudio MultiMeasure Sensors PASPORT and ScienceWorkshop are trademarks or registered trademarks of PASCO scientific in the United States and or in other countries All other brands products or service names are or may be trademarks or service marks of a...

Reviews: