background image

Theory of Operation

The Wireless Drop Counter uses an infrared light source and a
photodetector. Lenses spread the light into a "sheet" and refocus
it onto the photodetector. When a drop of fluid passes through
the Wireless Drop Counter, it partially blocks this sheet of light,
and the photodetector registers a momentary decrease in light
intensity.

The Wireless Drop Counter uses infrared light and ignores
visible light. On power up, it automatically adjusts the light level
for the best sensitivity. Use the Wireless Drop Counter away
from direct sunlight or other sources of infrared that may
interfere with its operation.

 Infrared light source

 Drop

 Detector

 Collimating lens

 Sheet of light

 Focusing lens

Troubleshooting the Sensor

• If the indicator light flashes twice for a single drop and the

Wireless Drop Counter registers more drops than actually
dispensed, fluid is most likely splashing back into the drop
window. Adjust the position of the beaker, Wireless Drop
Counter, or Drop Dispenser to eliminate splash back.

• If the indicator LED remains on constantly and the Wireless

Drop Counter does not register drops, the lenses in the
drop window are likely dirty or wet. Clean the lenses in the
drop window with water or a glass cleaner; dry with a
cotton swab or tissue.

• If the sensor loses Bluetooth connection and will not

reconnect, try cycling the ON button. Press and briefly hold
the ON button on the sensor for 10 seconds and then
release. Start the sensor in the usual way.

• If the sensor stops communicating with the computer

software or tablet application, try restarting the software or
application. If the problem remains, press and hold the ON
button on the sensor for 10 seconds and then release. Start
the sensor in the usual way.

- If this does not fix the problem, turn Bluetooth off and

then back on on the computing device. Retry.

• If the Wireless Drop Counter will not turn on, use the micro

USB cable to connect it to a USB port or charger.

Probe Care and Maintenance

The Wireless Drop Counter Sensor is splash-resistant but not
waterproof. Do not immerse the sensor in liquid. If the inside
surfaces of the drop window need to be cleaned, use a piece of
lens paper or a lint-free cloth and water or a glass cleaner.

Tips

• Except for opening and closing the stopcocks, do not touch

the syringe of the Drop Dispenser during the titration;
otherwise the drop size may change significantly.

• The drop rate must remain approximately constant (for

example, about 2 drops per second) in order for the drop
size to remain constant.

• Drop size depends on the exact position of the stopcock

valves. Therefore, the average drop size will be different for
every titration. For each different titration, re-calibrate the
sensor, since the new titrant may have differently sized
drops if it has a different density.

Software Help

The SPARKvue and PASCO Capstone Help provide additional
information on how to use this product with the software.
You can access the help within the software or online.

 SPARKvue

Software

Main Menu 

 > Help

Online

pasco.com/help/sparkvue

 PASCO Capstone

Software

Help > PASCO Capstone Help

Online

pasco.com/help/capstone

Technical Support

Need more help? Our knowledgeable and friendly Technical
Support staff is ready to answer your questions or walk you
through any issues.

 Chat

pasco.com

 Phone

1-800-772-8700 x1004 (USA)
+1 916 462 8384 (outside USA)

 Email

[email protected]

Regulatory Information

Limited Warranty

For a description of the product warranty, see the Warranty and Returns page at

www.pasco.com/legal

.

Copyright

This document is copyrighted with all rights reserved. Permission is granted to non-
profit educational institutions for reproduction of any part of this manual, providing
the reproductions are used only in their laboratories and classrooms, and are not
sold for profit. Reproduction under any other circumstances, without the written
consent of PASCO scientific, is prohibited.

Trademarks

PASCO and PASCO scientific are trademarks or registered trademarks of PASCO
scientific, in the United States and in other countries. All other brands, products, or
service names are or may be trademarks or service marks of, and are used to
identify, products or services of, their respective owners. For more information visit

www.pasco.com/legal

.

Wireless Drop Counter | PS-3214

6

Summary of Contents for PS-3214

Page 1: ...rt It includes a rechargeable battery and a Micro USB Cable It measures the number of drops of fluid that fall through the rectangular opening drop window of the sensor and converts the drops into a volume It is typically used in conjunction with a pH sensor and other equipment to perform a titration Data from the Wireless Drop Counter and other sensors are recorded and displayed using software su...

Page 2: ...utton until the lights turn on 2 Open PASCO Capstone 3 Click Hardware Setup 4 Select the sensor that matches its device ID A notification will appear if a firmware update is available Click Yes to update the firmware 5 Close Capstone ON OFF and Sleep Information Before turning the sensor on make sure that nothing is blocking the drop window To turn the sensor on press and hold the ON button When t...

Page 3: ... adjust the valve on the bottom stopcock so that the drops fall at a rate of about 1 drop per second Close the top valve after the rate is achieved Discard the titrant in the beaker as directed by the instructor 4 Mount the Wireless Drop Counter to the same support rod as the Drop Dispenser between the Dispenser and the base 5 Align the drop tip with the rectangular window in the Wireless Drop Cou...

Page 4: ... Sensor Data b Turn on the Wireless Drop Counter c From the list of available Bluetooth devices select the device matching the Device ID number XXX XXX printed on your sensor 8 Select a Template from the options on the right to enter the Experiment window 9 Select the Hardware Setup icon in the bottom right corner This window shows the name of all connected sensors and selected measurement choices...

Page 5: ...ic stirrer and a 150 mL beaker with 100 mL of 0 005 M HCl such that the end of the pH Probe and the Micro Stir Bar are submerged in the liquid in the beaker 5 Mount the Drop Dispenser above the Wireless Drop Counter and position it so that the drop tip is placed over the rectangular window of the Drop Counter Fill the syringe with 50 mL of 0 1 M NaOH Acid Base Titration 1 Set up the equipment as s...

Page 6: ...terproof Do not immerse the sensor in liquid If the inside surfaces of the drop window need to be cleaned use a piece of lens paper or a lint free cloth and water or a glass cleaner Tips Except for opening and closing the stopcocks do not touch the syringe of the Drop Dispenser during the titration otherwise the drop size may change significantly The drop rate must remain approximately constant fo...

Page 7: ...al requirements and other relevant provisions of the applicable EU Directives FCC statement This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation Battery disposal Batteries contain chemic...

Reviews: