background image

®

Wireless pH Sensor

PS-3204

8

013-14662B

Appendix B: Calibration

Prepare for Calibration

Calibration will need distilled water, two different pH buffer 
solutions, and containers for the water and buffer solutions. 
The sensor will need to be “paired” with a tablet or 
computer, and the data collection software should be 
running.

The 

pH Buffer Capsule Kit (SC-2321)

 from PASCO 

includes three vials, each with ten capsules of pH 4.0, 
pH 7.0, and pH 10.0, and a bottle of preservative 
solution which contains a pH indicator and will color 
each buffer solution for easy identification. 

Each capsule can make 100 milliliters (ml) of buffer 
solution. 

Containers such as 100 ml and 1000 ml beakers are 
available from PASCO. 

Make two buffer solutions with pH values that “bracket” 
the pH values to be measured. For this example, make 
one with pH 4 and the other with pH 7. The calibration 
buffer solutions should be at the same temperature as 
the solutions that will be measured.

Using SPARKvue Software for Calibration

NOTE: Check the on line Help System in SPARKvue for the 
most up-to-date information.

1.

Click (or press) the Experiment Tools button (

).

The 

Experiment Tools

 screen opens.

2.

Click 

Calibrate Sensor

.

The 

Calibrate Sensor: Select Measurement

 screen 

opens.

3.

Click the Sensor box and click the sensor to be 
calibrated.

4.

Click the 

Calibration Type

 box and click a calibration 

type. (For this example, click “

2-point

”.)

5.

Click 

Next

.

The 

Calibrate Sensor Enter Values

 screen opens.

6.

Rinse the end of the pH Probe in distilled water and 
then put the pH Probe into a pH 4 buffer solution.

7.

Under 

Calibration Point 1

, click the 

Standard Value

 

box and enter the known value (4.00 in this example).

8.

Under 

Calibration Point 1

, click 

Read From Sensor

.

The value measured by the sensor is transferred to the 

Sensor Value

 box.

9.

Take the probe out of the first buffer solution and rinse 
the end of the probe with distilled water. Dry the probe 
and then put the pH Probe into the pH 7 buffer solution.

10.

Under 

Calibration Point 2

, click the 

Standard Value

 

box and enter the pH buffer’s known value (7.00 in this 
example).

11.

Under 

Calibration Point 2

, click the 

Read From 

Sensor

 box.

The second value measured by the sensor is 
transferred to the 

Sensor Value

 box.

12.

Click OK.

Using PASCO Capstone for Calibration

NOTE: Check the on line Help System in PASCO Capstone 
for the most up-to-date information.

1.

Click 

Calibration

 (

) in the 

Tools

 Palette.

2.

Choose the probe you would like to calibrate now: 

pH 

Measurement

.

3.

Click 

Next

.

4.

Choose the type of calibration you would like to 
perform:

 Two Standards (2 point)

.

5.

Click 

Next

.

6.

Enter the first known pH buffer value (4.00 in this 
example) in the 

Standard Value

 text box.

7.

Rinse and dry the pH Probe

8.

Place the pH Probe in the pH 4 buffer solution and stir.

9.

When the 

Current Value

 stabilizes, click 

Set Current 

Value to Standard Value

.

10.

Click 

Next

.

11.

Enter the second known pH buffer value (7.00 in this 
example) in the 

Standard Value

 text box.

12.

Remove the probe from the first buffer solution and 
rinse and dry the end of the probe

13.

Place the probe in the second buffer solution and stir.

14.

When the 

Current Value

 stabilizes, click 

Set Current 

Value to Standard Value

.

Summary of Contents for PS-3204

Page 1: ...n one can be connected to a computer or tablet at the same time The pH Sensor can also be used with alternative probes such as Ion Selective Electrodes ISE and the Oxidation Reduction Potent ORP Probe...

Page 2: ...connector Twist the probe s BNC connector clockwise left to right about one quarter turn to lock it in place To remove the pH Probe from the Wireless pH Sensor reverse the process NOTE When removing t...

Page 3: ...e pH Probe in the solution to be measured The bulb shaped glass membrane should be entirely immersed 3 Press or click start or record to begin recording data 4 Wait for the reading to stabilize 5 Rins...

Page 4: ...or application If the problem remains press and hold the ON button for ten seconds and then release the button Turn on the sensor in the usual way Turn Bluetooth off and then turn it back on Retry pH...

Page 5: ...es training in the safe handling of flammable caustic and corrosive chemicals and a working knowledge of serial solution and calibration procedures Suggested Experiments Practically any experiment tha...

Page 6: ...ary by country and region It is your responsibility to recycle your electronic equipment per your local environmental laws and regulations to ensure that it will be recycled in a manner that protects...

Page 7: ...hing lower than 0x6 means an older version of Bluetooth Your device will need the PS 3500 USB Bluetooth 4 0 Adapter 1 The Mac Mini and MacBook Air were updated with Bluetooth SMART support in 2011 The...

Page 8: ...in distilled water and then put the pH Probe into a pH 4 buffer solution 7 Under Calibration Point 1 click the Standard Value box and enter the known value 4 00 in this example 8 Under Calibration Po...

Page 9: ...nt that the sensor sends to the GLX when it is in that standard In a one point calibration only one of the points is reset by the user Types of Calibration There are three types of calibration two poi...

Reviews: