Wireless pH Sensor

PS-3204

2

013-14662B

Data Collection Software

See the PASCO web site at www.pasco.com/software for help in 
selecting the right PASCO software and to check the latest versions.

 

Compatibility

See Appendix A for more information about the PS-3500 Adapter and 
Mac OS X models.

LED Information

The Bluetooth connection LED (light-emitting diode) and 
the battery status LED operate as follows:

*

Logging

: PASCO wireless sensors can either stream live 

data to a compatible device or log data (save it to the 
sensor’s memory). The data can then be uploaded to the 
device for display and analysis at a later time. Logging 
capability supports long-term or remote data collection 
while not connected to the device. 

Note: Versions of SPARKvue and PASCO Capstone 
available in 2016 will support logging. Check the PASCO 
Web page at:

www.pasco.com/software

for the latest software version.

Set-up

Connecting the pH Probe

Hold the Wireless pH Sensor in one hand and the pH 
Probe in the other. Align the tabs on the sensor’s BNC 
connector with the slots on the BNC connector at the end of 
the pH Probe. Push the probe’s BNC connector onto the 
sensor’s BNC connector. Twist the probe’s BNC connector 
clockwise (left-to-right) about one-quarter turn to lock it in 
place.

To remove the pH Probe from the Wireless pH Sensor, 
reverse the process. NOTE: When removing the probe, it 
may help to push the pH Probe onto the pH Sensor 
connector slightly before twisting the BNC connector 
counter-clockwise to release the probe. 

Removing the Storage 
Bottle

1.

Hold the pH Probe vertically so 
that the solution will not spill out 
of the bottle.

2.

Unscrew the plastic cap and 
remove the bottle. Keep the 
storage solution for later use.

3.

Push the bottle cap and the 
rubber washer up the pH Probe 
shaft to keep them out of the 
way.

Connecting the Sensor to a Wireless Device 

PASCO Capstone

SPARKvue

Mac OS X

Mac OS X

Windows

Windows

iOS 

Android

Chromebook

Platform

Bluetooth SMART Compatibility

iOS

iPad 3 and later
iPhone 4S and later
iPod touch 5 and later

SPARK Element

All models

Android

Android 4.3 and later

Chromebook

Chrome OS (requires PS-3500 Adapter*)

Mac OS X

Models introduced July 2011 or later*

Windows

Windows 7 and later (requires PS-3500 Adapter*)

Bluetooth LED

Status

Battery LED

Status

Red blink

Ready to pair

Red blink

Low power

Green blink

Connected

Yellow blink

Logging*

Sensor

pH Probe

slot

tab

1. Align tab and slot

2. Push probe onto BNC 

connector

3. Twist probe BNC 

connector clockwise

Cap 

and 

washer

Bottle

Summary of Contents for PS-3204

Page 1: ...n one can be connected to a computer or tablet at the same time The pH Sensor can also be used with alternative probes such as Ion Selective Electrodes ISE and the Oxidation Reduction Potent ORP Probe...

Page 2: ...connector Twist the probe s BNC connector clockwise left to right about one quarter turn to lock it in place To remove the pH Probe from the Wireless pH Sensor reverse the process NOTE When removing t...

Page 3: ...e pH Probe in the solution to be measured The bulb shaped glass membrane should be entirely immersed 3 Press or click start or record to begin recording data 4 Wait for the reading to stabilize 5 Rins...

Page 4: ...or application If the problem remains press and hold the ON button for ten seconds and then release the button Turn on the sensor in the usual way Turn Bluetooth off and then turn it back on Retry pH...

Page 5: ...es training in the safe handling of flammable caustic and corrosive chemicals and a working knowledge of serial solution and calibration procedures Suggested Experiments Practically any experiment tha...

Page 6: ...ary by country and region It is your responsibility to recycle your electronic equipment per your local environmental laws and regulations to ensure that it will be recycled in a manner that protects...

Page 7: ...hing lower than 0x6 means an older version of Bluetooth Your device will need the PS 3500 USB Bluetooth 4 0 Adapter 1 The Mac Mini and MacBook Air were updated with Bluetooth SMART support in 2011 The...

Page 8: ...in distilled water and then put the pH Probe into a pH 4 buffer solution 7 Under Calibration Point 1 click the Standard Value box and enter the known value 4 00 in this example 8 Under Calibration Po...

Page 9: ...nt that the sensor sends to the GLX when it is in that standard In a one point calibration only one of the points is reset by the user Types of Calibration There are three types of calibration two poi...