background image

®

Model No.PI-8127

 Operation

3

Output:

 Connect a device such as a speaker, wave driver, or electrical circuit to the 

output jacks at the lower right corner of the front panel. 

Frequency Adjust

The Frequency Adjust knob serves two purposes. Turn the knob to change the fre-
quency. Press the end of the knob as if it were a button to toggle between the two fre-
quency ranges.

Range Selection:

 The default range is 0.001 kHz to 100.000 kHz or 150.000 kHz for 

the sine waveform (other waveforms are distorted above 100 kHz). The other range is 
0.001 Hz to 999.999 Hz. Press the end of the Frequency Adjust knob as if it were a 
button to switch between ranges. The default frequency for the lower range is 1.000 
Hz.

Frequency Adjust

: The default frequency for the default range is 1.000 kHz. Notice 

that the “1” is underlined in the display. This indicates that this is the digit that will 
change first when you turn the knob. For example, turn the knob one “click” to the 
right (clockwise), and the frequency increases to 2.000 kHz. Each “click” to the right 
(clockwise) or left (counterclockwise) will increment or decrement the 

digit that is 

underlined

 in the display. When you reach 9.000 kHz, the next “click” to the right 

will change the frequency to 10.000 kHz. Notice that the underline stays 
below the “ones” place. Turning the frequency knob one more “click” to the 
right changes the frequency to 11.000 kHz.

As a second example, if the frequency is 1.000 kHz, turning the Frequency Adjust 
knob one “click” to the left (counterclockwise) does not decrease the frequency to 
0.000 kHz. Instead, the frequency stays at 1.000 kHz (it is not possible to have zero 
frequency). To lower the frequency, press the right-hand Resolution Selection button 
to move the underline to the right (1.000 kHz) and then turn the Frequency Adjust 
knob counterclockwise one “click” to decrease the frequency to 0.900 kHz.

Resolution Selection:

 Use the Resolution Selection buttons beneath the Frequency 

Adjust knob to change the active (underlined) digit in the display. Pushing the left 
button moves the underline to the left and pushing the right button moves the under-
line to the right. 

Use both the Frequency Adjust knob and the Resolution Selection buttons to set the 
frequency. Remember that you can push the OUTPUT STANDBY button to disable 
the output while you adjust the frequency, and then push the button again to enable 
the output when you have the frequency set the way you want.

Clear Feature:

 While adjusting the frequency, you can clear (zero) all the digits to 

the right of the underline cursor by pressing the Voltage Adjust knob as if it were a 
button.

Voltage Adjust

The Voltage Adjust knob serves three purposes. 

Turn the knob to change the output voltage. 

When the LCD shows the waveform list or the menu, turn the Voltage Adjust 
knob to move the pointer up or down in the list or in the menu. 

Press the knob as if it were a button to select a choice from the waveform list or 
from the menu. 

Press the Frequency Adjust 

knob to change the range.

NOTE: You can conveniently go 
to the lowest frequency without 
overshooting by rapidly turning 
the Frequency Adjust knob. 

Resolution Selection buttons

Press the Voltage Adjust 

knob to select a waveform 

or menu item

NOTE

To set up the frequency, volt-
age, and other parameters 
without outputting a signal, 
press the OUTPUT 
STANDBY button. 

Make the adjustments, and 
press the OUTPUT 
STANDBY button again to 
begin the signal.

Summary of Contents for PI-8127

Page 1: ...ment See the PASCO Catalog for details Mechanical Wave Driver SF 9324 Open Speaker WA 9900 AC DC Electronics Laboratory EM 8656 RLC Circuit Board CI 6512 Output jacks 4 mm Resolution Selection buttons Output Standby button Power button Liquid Crystal Display LCD Frequency Adjust knob and Range Selection button Waveform Selection button Menu button Voltage Adjust knob and Menu Selection button ...

Page 2: ...Voltage Maximum The maximum current or maximum voltage can be set using a menu This is useful when the voltage applied to a circuit needs to be limited The LCD displays the output voltage and current Offset Voltage Any waveform can be offset up or down with a DC voltage ranging from 10 V to 10 V provided the peak voltage does not exceed 10 V Frequency Sweep Sweep between two frequencies at a selec...

Page 3: ...sible to have zero frequency To lower the frequency press the right hand Resolution Selection button to move the underline to the right 1 000 kHz and then turn the Frequency Adjust knob counterclockwise one click to decrease the frequency to 0 900 kHz Resolution Selection Use the Resolution Selection buttons beneath the Frequency Adjust knob to change the active underlined digit in the display Pus...

Page 4: ...part of the LCD turn the Frequency Adjust knob one click to the left or right Waveforms Press the WAVEFORM button to open the first waveform menu The menu shows five choices Sine Square Triangle Positive Ramp and Negative Ramp with the pointer at Sine Turn the Voltage Adjust knob one click to the left counterclock wise to move the pointer from Sine to Square Press the Voltage Adjust knob as if it ...

Page 5: ...n a modulated Mod waveform is selected the signal produced by the Function Generator will be multiplied by the signal from the external input Change the frequency of the Function Generator s signal with the Frequency Adjust knob Adjust the percent modulation with the Voltage Adjust knob The formula for the percent modulation is based on the amplitudes of the peak P and the trough T of the modulate...

Page 6: ... menu choice Press the Voltage Adjust knob as if it were a button to toggle switch to the other choice Low The Low setting is useful when the Function Generator is in a darkened environment Sweep Turn the Voltage Adjust knob so the pointer is next to Sweep and then press the knob as if it were a button to select the choice The default setting is Off The Sweep menu options offer several ways to tai...

Page 7: ...is 1 second and the maximum is 9999 seconds 2 h 46 min 39 s Press the Voltage Adjust knob as if it were a button to return to the menu of Sweep options Step The default step frequency is None This means that the default frequency sweep will be continuous from the initial to the final frequency As with the initial fre quency use the Voltage Adjust knob to select Step and press the knob as if it wer...

Page 8: ...ted for experiment in waves and acoustics Drive a speaker or several speakers with the Function Generator Accurately set the frequency to any level in the acoustic range or beyond This makes it particularly con venient for experiments such as measuring the speed of sound observing the interfer ence and diffraction of sound waves and investigating acoustic resonance in a cavity Use the Function Gen...

Page 9: ...ons that vary by country and region It is your responsibility to recycle your electronic equipment per your local environmental laws and regulations to ensure that it will be recycled in a manner that protects human health and the environment To find out where you can drop off your waste equipment for recycling please contact your local waste recycle disposal service or the place where you purchas...