background image

Drop Counter

4

012-08470G

9.

Observe the data appearing on the pH versus drop count 
graph. After the equivalence point is reached, continue 
collecting data until the pH curve flattens.

10.

Close the stopcocks.

11.

Stop data recording.

12.

Read the final volume of fluid in the syringe and calcu-
late the net volume of titrant that was dispensed.

13.

Read the final drop count on the graph.

14.

In the calculator window of the software or datalogger, 
enter a calculation as shown below, but using your own 
values of total volume dispensed and final drop count. 
The calculation gives the volume of titrant dispensed 
over time based on the number of drops and the average 
volume per drop.

volume = [Drop Count (drops)] * 100/3050

In this example, “100” is the total volume dispensed (in 
mL) and “3050” is the final drop count.

15.

In the graph, change the horizontal axis to the calculated 
“volume”.

Tips

Except for opening and closing the stopcocks, do not 
touch the syringe during the titration; otherwise, the 
drop size may change significantly.

The drop rate must remain approximately constant 
(about 2 drops per second) in order for the drop size to 
remain constant.

Drop size depends on the exact position of the bottom 
stopcock. Therefore, the average drop size will be dif-
ferent for every titration. For each titration, re-enter the 
volume used and drop count values in the calculation.

Other Titrations

Titrations with different solutions can be performed using 
the same equipment and procedure. Other probes (such as 
conductivity or ORP) can be used in place of the pH probe.

Theory of Operation

The Drop Counter 
uses an infrared light 
source and a photo-
detector. Lenses 
spread the light into a 
“sheet” and refocus it 
onto the photodetec-
tor. When a drop of 
fluid passes through the Drop Counter, it partially blocks this 
sheet of light, and the photodetector registers a momentary 
decrease in light intensity.

The Drop Counter uses infrared light and ignores visible 
light. On power up, it automatically adjusts the light level for 
the best sensitivity. Use the Drop Counter away from direct 
sunlight or other sources of infrared that may interfere with 
it.

Troubleshooting

equivalence point

final drop count

Select volume calculation 

for horizontal axis

Problem

Cause

Solution

Indicator light 
flashes twice for a 
single drop. Drop 
Counter registers 
more drops than 
actually dispensed.

Fluid is splashing 
back into the 
drop window.

Adjust the position 
of the beaker, Drop 
Counter, or syringe 
to eliminate splash 
back

Indicator LED 
remains on 
constantly. Drop 
Counter does not 
register drops.

Lenses in drop 
window are dirty 
or wet.

Clean lenses with 
water and dry with a 
cotton swab or 
tissue.

detector

collimating

lens

focusing 
lens

sheet of light

drop

infrared light 

source

Reviews: