Drop Counter

2

012-08470G

Introduction

The Drop Counter measures the number of drops of fluid 
that fall through the rectangular opening of the sensor. It is 
typically used in conjunction with a pH sensor and other 
equipment to perform a titration. Data from the Drop 
Counter and other sensors are recorded and displayed in soft-
ware such as SPARKvue, PASCO Capstone, or on a datalog-
ging interface such as the SPARK SLS or Xplorer GLX.

The Drop Counter can be mounted on a support rod and the 
support rod can be clamped to a rod stand. The counter is 
equipped with three integrated probe holders. The included 
Micro Stir Bar fits onto the end of a pH probe or probe of the 
same diameter.

Assembling the Drop Dispenser

1.

Remove the plunger from the syringe.

2.

Connect the two stopcocks together and set the valves at 
right angles. Connect the stopcocks to the end of the 
syringe.

3.

Connect the drop tip to the bottom of the stopcocks.

 

Calibrating the Drop Counter

The Drop Dispenser has two stopcocks. The top stop cock is 
used to regulate flow rate and the bottom stopcock is used to 
turn the flow on and off. The bottom stopcock should either 
be in the completely open or completely closed position.

The following two methods illustrate the calibration proce-
dure.

Calibration (Generic Process)

1.

Attach the Drop Dispenser (syringe, two stopcocks and 
drop tip) to a support rod.

2.

Fill the syringe with titrant.

3.

Open both stopcock valves and adjust the valve on the 
top stopcock so that the drops fall at a rate of about 1 
drop per second. Close the bottom valve after the rate is 
achieved.

4.

Connect the Drop Counter to a PASPORT interface.

5.

Align the drop tip with the opening in the Drop Counter. 
Put a graduated 10 mL cylinder under the Drop Counter 
opening.

6.

Open the bottom stopcock valve and monitor the num-
ber of drops.

7.

After about 10 mL of titrant is collected in the graduated 
cylinder, close the bottom stopcock.

8.

Open the calibration screen in the data acquisition soft-
ware and input the number of drops and the volume.

Calibration Using SPARKvue or SPARK SLS

The drop counter is calibrated, using the SPARK SLS or 
SPARKvue Tools to correlate the number of drops counted 
to the volume of liquid dispensed from the Drop Dispenser.

1.

Attach the Drop Dispenser (syringe, two stopcocks and 
drop tip) to a support rod.

2.

Fill the syringe with titrant.

3.

Open both stopcock valves and adjust the valve on the 
top stopcock so that the drops fall at a rate of about 1 
drop per second. Close the bottom valve after the rate is 
achieved.

4.

Connect the Drop Counter and launch SPARK SLS or 
SPARKvue.

5.

Align the drop tip with the opening in the Drop Counter. 
Put a graduated 10 mL cylinder under the Drop Counter 
opening.

60 mL syringe 

with 50 mL of 

0.1 M NaOH

150 mL 

beaker with 

100 mL of 

0.005 M HCl

Three-finger 

Clamp

Multi 

Clamp

Stopcock

Stopcock

Drop Tip

pH Probe

Micro Stir 

Bar

Stir Plate