background image

®

EcoZone System

ME-6668

2

The unique design of the EcoZones allows the system to 

remain closed, which allows for measurements to have a 

minimal impact on the system. Environmental science 

probes can be used to actively measure variables in the 

ecosystem model. A syringe and tubing are included to 

extract water for chemical testing, replenished water 

without opening the system, and inject pollutants for 

chamber manipulation.

Set-up and Experiment 

Considerations

Before setting up the EcoZone System, there are some 

things that should be taken into consideration.

Consider what you are interested in measuring. There 

are nine open spaces available for sensors in the lids of 

the chambers. If you intend to measure water quality 

(dissolved oxygen, conductivity, etc.), consider using the 

Advanced Water Quality Sensor, which supports several 

different probes, instead of individual sensors.

A cotton rope is included to allow for efficient wicking of 

water from the aquatic chamber to the other chambers. 

Be sure that this wick covers each chamber sufficiently. 

Lay the wick underneath the plants both under and over 

the plants in the terrestrial chamber, for example, to 

ensure that plants receive sufficient water. Use 

approximately 36 inches of rope.

Caution: This wick is very efficient at moving water 

through the chambers. Do not use more than one 

strand, or you will deplete your aquatic chamber and 

flood the other chambers. Also, it is not necessary to 

completely submerge the wick in the aquatic chamber. 

Only a small amount of rope at the surface is needed to 

move the water. Recommendation: Test the wick first to 

determine how much is needed for your situation.

Cotton rope is used for its excellent wicking ability. 

However, cotton in contact with soil may decompose 

over a period of weeks. You may need to check and 

replace it periodically.

Equipment Set-up

Important: When carrying the EcoZone System, do NOT pick the 

system up by the aquatic chamber (the heaviest chamber). Pick the 

entire system up by the tray.

1. Place the three EcoChambers into the custom tray.

2. Connect the EcoChambers using the rubber 

grommets as illustrated.

3. Fill each EcoChamber according to the instructions 

in the lab you are performing.

4. Select the sensors and interface that you intend to 

use for the experiment.

5. Place the lids onto the EcoChambers, lining up open 

arches at the bottom of each lid with the rubber 

connection grommets on the chambers.

6. Insert large stoppers with holes into the holes in the 

lids.

7. Insert probes into the holes in the stoppers.

8. Insert small solid stoppers into unused holes in the 

larger stoppers.

9. Connect the sensors to the interface.

Sensor Calibration

Calibrate sensors prior to using them in an experiment. 

Refer to the documentation provided with each sensor 

for calibration instructions.

Note: In most cases, sensors do not require calibration. 

If, for example, you wish to measure only relative 

dissolved oxygen changes, the sensor does not to be 

calibrated.

Data Analysis

Graphs showing two variables changing over time are 

most valuable in the EcoZones experiments because 

Chambers connected with grommets

Reviews: