background image

12 

 

 

Quality Assurance Plan 

 

Under the U.S. Department of Transportation workplace testing program (see 49 CFR, 
Part  40),  transportation  employers  are  required  to  test  employees  working  in  certain 
safety  sensitive  positions  for  alcohol  under  certain  conditions.    The  DOT  workplace 
testing  program  requires  that  breath  test  instrument  manufacturers  provide  employers 
or Breath Alcohol Technicians with this Quality Assurance Plan, which together with the 
operation  instructions  provided  with  the 

Mark  V

,  will  assist  in  assuring  that  breath 

testers are calibrated to the required degree of accuracy. 
 
QAP: 

1.  Allowed  Calibration  Units:    Any  wet  bath  simulator  listed  on  the  NHTSA 

Conforming  Products  List  of  Calibration  Units  for  Breath  Alcohol  Tests.    When 
calibration  or  re-calibration  (not  an  accuracy  check)  is  needed  the  simulator 
should  be  used  with  a  certified  0.080%  BAC  solution,  following  the  operating 
manual provided by the wet bath simulator manufacturer.  Alternatively, a dry gas 
standard of 0.080% that has been approved by NHTSA may also be used. 

2.  External Calibration Check Interval: Calibration Checks (Accuracy Checks) should 

be performed monthly; after positive test results; or if the unit fails to air blank to 
0.000  after  2  attempts,  and  after  repair.   There  is  no  limitation  on the  number  of 
tests  that  may  be  conducted  between  calibration  checks,  providing  the  monthly 
checks are completed. 

3. External Calibration Check Tolerance:  +/- 0.005 or +/- 5% at or above 0.100. 
4. Intervals  for  Periodic  Inspection:    Self-diagnostics  and  visual  inspection  by  the 

operator  before  every  use.  Routine  maintenance  and  service  recommended 
every  2  years.  Calibration  should  be  performed  when  2  consecutive  calibration 
(accuracy) checks fail (out of tolerance). 

5. Events which require instrument be taken out of service:  Self-diagnostics failure 

or indication by error code.  The power supply is not providing necessary power 
to the Mark V.  The Mark V does not display “BLOW” within 2 minutes of power 
up. 

6.  For  other  information  regarding  quality  assurance  unique  to  this  instrument,  see 

the  Mark  V  Calibration  Software  Manual  as  set  forth  in  the  Mark  V  Operating 
Manual. 
 

Follow  the  operating  instructions  provided  by  the  manufacturer  for  proper 
procedure to use dry gas for calibration checks. 

 
Instruments indicating any of the above ERRORS should be returned to PAS Systems 
International, Inc. for repair. Please call PAS Systems International, Inc. Technical 
Service Team to describe the problem & for more information on sending the unit in for 
service. Please have the serial number of the unit.  
 

This QAP is subject to change and should neither be considered a final 
requirement nor a contractual term in any agreement to purchase the MARK V. 

Summary of Contents for MARK V Alcovisor

Page 1: ...K V Alcovisor Operating Manual PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com 800 660 SNIF 540 372 3431 Copyright 2011 by PAS Systems International Inc A...

Page 2: ......

Page 3: ...or Test Subject 5 Air Blank Feature 5 Measuring BrAC BAC 6 Automatic Sampling 6 Manual Sampling 7 Conducting a Passive Test No Mouthpiece 7 Accessing Test Records 8 Accuracy Checking and Calibration 8...

Page 4: ......

Page 5: ...cedure to be conducted in about 30 seconds or less This manual describes the operation maintenance calibration check and calibration adjustment of the MARK V This manual should be read completely and...

Page 6: ...r and displayed on screen The instrument is simple to operate and may be used as often as required provided that a suitable delay is allowed between successive tests This time delay allows the fuel ce...

Page 7: ...e cover When replacing the batteries always replace ALL four batteries Turning On Off ON Press and hold the Green ON OFF button for 2 3 seconds The display light comes on and a self test automatic Bla...

Page 8: ...rea and RIGHT green buttons to select the correct Date and Time b Select Update to confirm 2 Auto Off a Use the LEFT red and RIGHT green buttons to Select the time limit to turn off the device automat...

Page 9: ...e test The person must also avoid repeated deep breaths hyperventilation as this will temporarily cool down the breath and may lead to a false reading Should the person have diminished lung capacity a...

Page 10: ...click sounds Blowing can cease when the click is heard 7 After a few seconds the Test Result will be displayed on the LCD If the unit is left on it will power down after displaying the test result fo...

Page 11: ...RK V ON 2 When Please Blow appears hold the sample port mouthpiece opening about 4 inches from the subject s mouth 3 Have the subject moderately blow towards the mouthpiece holder opening for several...

Page 12: ...est is not within the acceptable range the unit must be re calibrated by PAS Systems International Inc or by persons who have been properly trained for alcohol detector calibration Due to the sensitiv...

Page 13: ...nnected to the simulator outlet should not be longer than 2 in order to prevent condensation 4 Plug Simulator in and turn switch to the ON position 5 Allow the solution to heat 6 After 15 20 minutes c...

Page 14: ...red value will be displayed If the measured value is within 0 005 BrAC of the altitude adjusted concentration value the MARK V is working accurately Note Suitable Dry Gas Standards and Regulators are...

Page 15: ...0 080 Dry Gas 0 1 0 080 500 0 981 0 078 1000 0 962 0 077 1500 0 943 0 075 2000 0 925 0 074 2500 0 907 0 073 3000 0 889 0 071 3500 0 872 0 070 4000 0 854 0 068 4500 0 837 0 067 5000 0 820 0 066 5500 0...

Page 16: ...There is no limitation on the number of tests that may be conducted between calibration checks providing the monthly checks are completed 3 External Calibration Check Tolerance 0 005 or 5 at or above...

Page 17: ...call PAS Systems International Inc technical service at 800 660 7643 PROBLEM CAUSE SOLUTION The display is only faintly lit or the unit will not turn on The power supply to the unit is completely exh...

Page 18: ...Temperature 23 to 131 F 5 to 55 C Storage Conditions 13 to 158 F 25 to 70 C not more than 95 relative humidity LCD Display Size 2 0 X 1 5 viewing area 5cm X 4 cm Automatic backlight for night time us...

Page 19: ...onal Inc or an authorized service technician Only original MARK V parts may be used To keep the instrument clean periodically use a mild disinfectant and a soft cloth on the outside of the case DO NOT...

Page 20: ...MARK V Alcovisor Law Enforcement PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 http www pasintl com Tel 540 372 3431 800 660 SNIF Printed 10 12 5 00...

Reviews: