P
AGE
76
Performance Specifications
Measurement Error Limit
• Type J,K,T,E,N, & C thermocouples and RTD
+ or - 0.25% of reading plus 1 degree at 25
°
C
• Type R,S, & B thermo or - 0.25% of span at 25
°
C
• mVDC, mADC and VDC + or - 0.25% of scaled span plus
1 least significant digit at 25 degrees C
Ambient Temp. Error
0.01% of span per degree C deviation from 25 degrees C
Scan Rate
1 scan per second, 3 scan per second selectable
Display Resolution
0 to 3 decimal places (depending upon input type selected)
Auto Reset Windup Inhibit Auto reset is disabled when the process is outside of the
proportional band
Cold Junction
Self compensation for ambient temperature. All calibration
Compensation
values are stored in memory
Noise Rejection
Normal mode, 85dB minimum at 60 Hz or greater.
Common mode, 90dB m /- 24VAC maximum for RTD
input, 115 VAC maximum for other inputs
Line Voltage
115/230 VAC + /- 10% 50/60 Hz
Power Consumption
15VA maximum
Operating Temperature
0 to 55 degrees C
32 to 131 degrees F
Storage Temperature
-40 to 65 degrees C
-40 to 149 degrees F
Humidity
0 to 90% RH, noncondensing
Dimensions
1/4 DIN front panel (96mm X 96mm) 5.8 inches deep
Weight
3 pounds maximum
Vibration
0.5 to 100 Hz at 0.5g
Agency Approvals
UL and CSA
Warranty
3 years, details on the inside back cover.