OxyTechw² GAL Instruction Manual
7 Sensor Configuration
Before attempting to configure the sensor, please read the monitor user
manual that came with your monitor. The monitor manual will introduce
you to the basic set-up of the monitor, and will familiarise you with the
monitor menu structure and buttons.
The monitor leaves the factory with no sensors pre-installed.
Assuming the monitor has been physically connected to a sensor, the
next step is to register and configure the sensor before any measurements can be made. A single sensor
may provide one or more measurements. We advise only connecting one sensor at a time. Once the first
sensor has been registered, connect the second and register again. Repeat for any additional sensors.
All sensors must be registered to the monitor in this way, even if they are different types.
7.1 Sensor Config
From the MAIN MENU screen, select SENSOR CONFIG using the
arrow button, and press
.
7.2 Sensor Status
This option allows the user to review the current status of the 8 sensor
channels, these will all be set to disabled until a sensor is added.
Once a sensor has been installed the display will be updated to indicate the sensor type installed and
it's status.
7.3 Add Sensor
1. From the MAIN MENU screen, select SENSOR CONFIG by
pressing
, and press
.
2. The SENSOR MENU should be displayed. Press
to
highlight ADD SENSOR, and press
.
3. The Monitor will now search all possible addresses (0 to 240)
to find any attached sensors. During the search, any sensors
found will be displayed momentary before continuing with the
search.
4. Once the search is complete, the Monitor will display a list of sensors found. Each sensor will be
automatically allocated a new address from S:01 to S:08.
5. Repeat the above process to install a second, third or more sensors. A total of 8 sensors are
possible (expansion box may be required to add additional sensors).
6. Sensor addition is now complete.
7. If a single or multiple sensors have been found the
can be
pressed to escape from continuing the search.
7.4 S:0x OxyTechw² GAL
Once the sensor has been added and registered, the monitor will provide a
list of functions specific to the sensor. Press
or
to select the sensor
and press
. The CONFIG MENU will display a list of sensor functions.
Page 14 of 28
224083IM-Issue 04 20/03/2018
MAIN MENU
Monitor Config
Expansion Config
Sensor Config
Measurement Config
Alarm Config
Output Config
Information
SENSOR CONFIG
Sensor Status
Add Sensor
<No Sensors Installed>
SENSOR CONFIG
Sensor Status
Add Sensor
S:01 OxyTechw
2
GAL
M:01 %SAT
87.4
DISSOLVED OXYGEN
Monitor: OK 09:59:45 17/05/12