page 8 of 18
5.2
Turn
off (OFF)
ON
OFF
Press once (briefly)
5.3 Turn on back light
ON
OFF
Press the key (2 s)
The back light illumination will be switched off after 30 s.
5.4 MIN/MAX Display
The additional display line can be switched to MIN/MAX or FS format.
The scroll function indicates MIN/MAX after the other.
To measure pressure peaks the MIN/MAX Function is in use. The MIN/MAX memory saves the
highest (MAX) and the lowest (MIN) reading. Switching off the instrument, the MIN/MAX memory will
be erased.
When running different pressure tests one after another, the MIN/MAX memory should be deleted
(
RESET
) after every test cycle.
MIN
MAX
FS
MIN/MAX and FS
value is indicated in the display
5.5 FS Full Scale Display
Displaying the upper limit of the scale (FS) is designed to increase readability of the bar-graph
function. The upper limit of the measurement range is indicated numerically.
FS is indicated in sequence after MIN and MAX.
MIN
MAX
FS
FS
value is indicated in the display
5.6 Erasing the MIN/MAX values
RESET
OK
Erases MIN/MAX values
5.7 OFL Display
This indicates that the applied pressure
is outside given full scale range.
If the message will remain displayed,
while the
ServiceJunior
is pressure
less, please consult a Parker Hannifin
Sales Office.