Installation
8
CAUTION:
Observe the proper input voltage and correct polarity is applied.
Step 1
Check supply voltage matches the specified voltage described on the nameplate
Step 2
Avoid use of common power supply to other electrical devices
Step 3
Check the voltage at the terminals is within the limits
Step 4
The inrush current of one converter is 1.0A/2ms
1
2
3
4
5
6
No.
Connection
1
AL2
2
AL1
3
COM: (For AL1/AL2)
4
FG: Grounding terminals
5
Power Supply: 0V
6
Power Supply: 24VDC
No.
Connection
7
(+): RS485
8
(+): P-OUT
9
(+): A-OUT
10
(-): RS-485
11
(-): P-OUT
12
(-): A-OUT
No. Spec. & Name
Description
1
ZERO ADJ switch Push to adjust ZERO point
2
ADRS address
switch
Set addresses and communication
parameters of RS-485. For setting address
“25”, fit the arrow of “x10 dial address
switch” to 2 and the arrow of “x1 dial
address switch” to 5.
3
BNC connector
Connect the exclusive cables. Check “IN”
and “OUT” polarities when connecting
cables.
4
ERROR lamp (red)
This lamp is “on” when receiving signal is
abnormal
5
AGC/ZERO lamp
(red)
This lamp blinks when ZERO adjustment is
being performed
6
ALARM lamp (red)
This lamp is “on” when high or low alarm is
actuated
!
Requirements for Installation Site of the Signal Converter:
Cont.
Names and Functions of Panel Component Parts
Wiring of Converter terminal block
•
Use the correct DIN rail type terminal block shown below.
• Terminals connections for DIN type
• Pin out for connections for PAFC-100 DIN type terminals