7 - EN
2.2 Approvals and Compliance
2.2.1 Approvals
Directives
97/23/EC: Pressure Equipment Directive
2004/108/EC: Electromagnetic Compatibility Directive
2006/95/EC: Low Voltage Directive
Safety and Electromagnetic Compatibility Standards
This equipment has been tested and complies with the following European Standards:
EN 61326-1:2013
EMC - Electrical equipment for measurement, control and laboratory use. EMC requirements.
(Equipment tested to: Emissions - Light, Immunity - Heavy)
BS EN 61000-3-2:2006+A2:2009
Electromagnetic compatibility (EMC). Limits for harmonic current emissions (equipment input current = 16 A
per phase)
BS EN 61000-3-3:2013
Electromagnetic compatibility (EMC). Limits. Limitation of voltage changes, voltage fluctuations and flicker in public low-
voltage supply systems, for equipment with rated current = 16 A per phase and not subject to conditional connection.
BS EN 61010-1:2010
Safety requirements for electrical equipment for measurement, control, and laboratory use. General requirements
General
Designed generally in accordance with ASME VIII DIVISION 1: EDITION 2010 2011a Addenda
2.2.2 Compliance
This gas generator is compliant with FDA and European Pharmacopeia Regulations for use as a medical gas generator.
Inlet Parameters
Port Connections
Environmental Parameters
Air Quality
ISO 8573-1: 2010 Class 2.2.2
(2.2.1 with high oil vapour content)
Pressure
5 –13 bar g (72.5 - 188.5) psi g
Temperature
5 – 50
o
C (41 – 122
o
F)
Purity
20.948%
(wrt O2)
0.0314%
(wrt CO2)
Air Inlet
G1”
N
2
Outlet to Buffer
G1”
N
2
Inlet from Buffer
G1/2”
N
2
Outlet
G1/2”
Ambient Temperature
5 – 50
o
C
(41 – 122
o
F)
Humidity
50% @ 40
o
C (80% @
MAX
d
31
o
C)
IP Rating
IP20
/ NEMA 1
Pollution Degree
2
Installation Category
II
Altitude
< 2000 m
(6562 ft)
Noise
<80 dB (A)
Electrical Parameters
Packed Weights and Dimensions
Generator Supply
(1)
(1) The generator does not require adjustment when connecting to 115v and 230v
electrical supplies.
100 - 240 +/- 10% Vac 50/60Hz
Generator Power
(2)
(2) The power rating specified is for the generator alone and does not take in to
account any pre-treatment dryer connected to the dryer supply terminals of the
generator.
55 W
Fuse
(3)
(3) (Anti Surge (T), 250v, 5 x 20mm HBC, Breaking Capacity 1500A @ 250v, IEC 60127,
UL R/C Fuse).
3.15 A
Max Dryer Power
(4)
(4) The dryer is fed directly from the generator supply.
100W
Model
Height (H)
Width (W)
Depth (D)
Weight
mm
ins
mm
ins
mm
ins
Kg
lbs
N2-20P
725.5
28.6
1994
78.5
1090
42.9
398.4
878.3
N2-25P
1260
49.6
495.4 1092.1
N2-35P
1430
56.3
580.4 1279.6
N2-45P
1600
63.0
686.4 1513.3
N2-55P
825.5
32.5
1770
69.7
782.4 1724.9
N2-60P
1935
76.2
897.4 1978.4
N2-65P
828.5
32.6
2100
82.7
997.4 2198.9
N2-75P
831.5
32.7
2275
89.6 1093.4 2410.5
N2-80P
2445
96.3
1186.4 2615.6